Bérangère S Joly, Hélène Deniau, Chloé Doinel, Adeline Brouillard, Virginie Siguret, Paul Coppo, Agnès Veyradier
{"title":"使用everon®s100分析仪测量ADAMTS13活性的新型自动化方法的实际评估。","authors":"Bérangère S Joly, Hélène Deniau, Chloé Doinel, Adeline Brouillard, Virginie Siguret, Paul Coppo, Agnès Veyradier","doi":"10.1016/j.jtha.2025.02.038","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Thrombotic thrombocytopenic purpura (TTP) is a thrombotic microangiopathy characterized by a severe functional deficiency of ADAMTS-13. Measuring ADAMTS-13 activity is crucial for diagnosing TTP (<10 IU/dL), monitoring treatments, and detecting relapses (<20 IU/dL). The Technofluor assay (Technoclone) allows a rapid ADAMTS-13 activity measurement using the Ceveron s100 analyzer (Technoclone).</p><p><strong>Objectives: </strong>This study aimed to evaluate the analytical and clinical performance of this new test under real-world conditions.</p><p><strong>Methods: </strong>ADAMTS-13 activity was measured using 2 fluorometric methods: the Technofluor ADAMTS-13 activity on the Ceveron s100 (Technoclone) and our reference fluorescence resonance energy transfer-VWF73 method, in plasma samples collected under real-life conditions (January 12, 2024, to April 4, 2024) and retrospectively selected samples from our biobank. The analytical and clinical performance of the new test was assessed, focusing on the critical low levels (<30 IU/dL).</p><p><strong>Results: </strong>Four hundred samples were tested under real-world conditions and 100 others tested retrospectively. The Technofluor assay showed excellent analytical performance, with a detection limit of 0.1 IU/dL, repeatability coefficient of variation of <11%, and reproducibility coefficient of variation of 6.1% (high level, 90 IU/dL) and 7.5% (low level, 40 IU/dL). Clinical performances were strong at diagnosis (threshold, 10 IU/dL; sensitivity and positive predictive value, 1.0) and during the follow-up (threshold, 20 IU/dL; specificity, 1.0; 95% CI, 0.99-1.00; positive predictive value, 0.96; 95% CI, 0.90-1.01; negative predictive value, 0.98; 95% CI, 0.97-1.00). No analytical interference was observed.</p><p><strong>Conclusion: </strong>The Technofluor assay on the Ceveron s100 is a fast, reliable method for measuring ADAMTS-13 activity, proving effectiveness for diagnosis and monitoring of TTP patients. However, it remains crucial to interpret ADAMTS-13 activity measurement with clinical context to ensure accurate diagnosis and management.</p>","PeriodicalId":17326,"journal":{"name":"Journal of Thrombosis and Haemostasis","volume":" ","pages":""},"PeriodicalIF":5.5000,"publicationDate":"2025-03-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Real-world evaluation of a novel automated method for measuring ADAMTS-13 activity using the Ceveron s100 analyzer.\",\"authors\":\"Bérangère S Joly, Hélène Deniau, Chloé Doinel, Adeline Brouillard, Virginie Siguret, Paul Coppo, Agnès Veyradier\",\"doi\":\"10.1016/j.jtha.2025.02.038\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Thrombotic thrombocytopenic purpura (TTP) is a thrombotic microangiopathy characterized by a severe functional deficiency of ADAMTS-13. Measuring ADAMTS-13 activity is crucial for diagnosing TTP (<10 IU/dL), monitoring treatments, and detecting relapses (<20 IU/dL). The Technofluor assay (Technoclone) allows a rapid ADAMTS-13 activity measurement using the Ceveron s100 analyzer (Technoclone).</p><p><strong>Objectives: </strong>This study aimed to evaluate the analytical and clinical performance of this new test under real-world conditions.</p><p><strong>Methods: </strong>ADAMTS-13 activity was measured using 2 fluorometric methods: the Technofluor ADAMTS-13 activity on the Ceveron s100 (Technoclone) and our reference fluorescence resonance energy transfer-VWF73 method, in plasma samples collected under real-life conditions (January 12, 2024, to April 4, 2024) and retrospectively selected samples from our biobank. The analytical and clinical performance of the new test was assessed, focusing on the critical low levels (<30 IU/dL).</p><p><strong>Results: </strong>Four hundred samples were tested under real-world conditions and 100 others tested retrospectively. The Technofluor assay showed excellent analytical performance, with a detection limit of 0.1 IU/dL, repeatability coefficient of variation of <11%, and reproducibility coefficient of variation of 6.1% (high level, 90 IU/dL) and 7.5% (low level, 40 IU/dL). Clinical performances were strong at diagnosis (threshold, 10 IU/dL; sensitivity and positive predictive value, 1.0) and during the follow-up (threshold, 20 IU/dL; specificity, 1.0; 95% CI, 0.99-1.00; positive predictive value, 0.96; 95% CI, 0.90-1.01; negative predictive value, 0.98; 95% CI, 0.97-1.00). No analytical interference was observed.</p><p><strong>Conclusion: </strong>The Technofluor assay on the Ceveron s100 is a fast, reliable method for measuring ADAMTS-13 activity, proving effectiveness for diagnosis and monitoring of TTP patients. However, it remains crucial to interpret ADAMTS-13 activity measurement with clinical context to ensure accurate diagnosis and management.</p>\",\"PeriodicalId\":17326,\"journal\":{\"name\":\"Journal of Thrombosis and Haemostasis\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":5.5000,\"publicationDate\":\"2025-03-07\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Thrombosis and Haemostasis\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1016/j.jtha.2025.02.038\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"HEMATOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Thrombosis and Haemostasis","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1016/j.jtha.2025.02.038","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"HEMATOLOGY","Score":null,"Total":0}
Real-world evaluation of a novel automated method for measuring ADAMTS-13 activity using the Ceveron s100 analyzer.
Background: Thrombotic thrombocytopenic purpura (TTP) is a thrombotic microangiopathy characterized by a severe functional deficiency of ADAMTS-13. Measuring ADAMTS-13 activity is crucial for diagnosing TTP (<10 IU/dL), monitoring treatments, and detecting relapses (<20 IU/dL). The Technofluor assay (Technoclone) allows a rapid ADAMTS-13 activity measurement using the Ceveron s100 analyzer (Technoclone).
Objectives: This study aimed to evaluate the analytical and clinical performance of this new test under real-world conditions.
Methods: ADAMTS-13 activity was measured using 2 fluorometric methods: the Technofluor ADAMTS-13 activity on the Ceveron s100 (Technoclone) and our reference fluorescence resonance energy transfer-VWF73 method, in plasma samples collected under real-life conditions (January 12, 2024, to April 4, 2024) and retrospectively selected samples from our biobank. The analytical and clinical performance of the new test was assessed, focusing on the critical low levels (<30 IU/dL).
Results: Four hundred samples were tested under real-world conditions and 100 others tested retrospectively. The Technofluor assay showed excellent analytical performance, with a detection limit of 0.1 IU/dL, repeatability coefficient of variation of <11%, and reproducibility coefficient of variation of 6.1% (high level, 90 IU/dL) and 7.5% (low level, 40 IU/dL). Clinical performances were strong at diagnosis (threshold, 10 IU/dL; sensitivity and positive predictive value, 1.0) and during the follow-up (threshold, 20 IU/dL; specificity, 1.0; 95% CI, 0.99-1.00; positive predictive value, 0.96; 95% CI, 0.90-1.01; negative predictive value, 0.98; 95% CI, 0.97-1.00). No analytical interference was observed.
Conclusion: The Technofluor assay on the Ceveron s100 is a fast, reliable method for measuring ADAMTS-13 activity, proving effectiveness for diagnosis and monitoring of TTP patients. However, it remains crucial to interpret ADAMTS-13 activity measurement with clinical context to ensure accurate diagnosis and management.
期刊介绍:
The Journal of Thrombosis and Haemostasis (JTH) serves as the official journal of the International Society on Thrombosis and Haemostasis. It is dedicated to advancing science related to thrombosis, bleeding disorders, and vascular biology through the dissemination and exchange of information and ideas within the global research community.
Types of Publications:
The journal publishes a variety of content, including:
Original research reports
State-of-the-art reviews
Brief reports
Case reports
Invited commentaries on publications in the Journal
Forum articles
Correspondence
Announcements
Scope of Contributions:
Editors invite contributions from both fundamental and clinical domains. These include:
Basic manuscripts on blood coagulation and fibrinolysis
Studies on proteins and reactions related to thrombosis and haemostasis
Research on blood platelets and their interactions with other biological systems, such as the vessel wall, blood cells, and invading organisms
Clinical manuscripts covering various topics including venous thrombosis, arterial disease, hemophilia, bleeding disorders, and platelet diseases
Clinical manuscripts may encompass etiology, diagnostics, prognosis, prevention, and treatment strategies.