{"title":"3特斯拉全胰腺积水T1成像。","authors":"Elizabeth Huaroc Moquillaza, Kilian Weiss, Lisa Steinhelfer, Jonathan Stelter, Marcus R Makowski, Rickmer Braren, Mariya Doneva, Dimitrios C Karampinos","doi":"10.1007/s10334-025-01224-8","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>A fast T<sub>1</sub> mapping method of the whole pancreas remains a challenge, due to the complex anatomy of the organ. In addition, a technique for pancreas water T<sub>1</sub> mapping is needed, since the T<sub>1</sub> is biased in the presence of fat. The purpose of this work is to accelerate the acquisition of water selective T<sub>1</sub> (wT<sub>1</sub>) mapping for the whole pancreas at 3 T.</p><p><strong>Methods: </strong>The proposed method combines a continuous inversion-recovery Look-Locker acquisition with a single-shot gradient echo spiral readout, water-fat separation and dictionary matching for wT<sub>1</sub> mapping of the whole pancreas at 3 T. The bias of T<sub>1</sub> in the presence of fat was evaluated in a phantom by comparing the modified Look-Locker inversion-recovery (MOLLI) and the proposed method to MRS measurements. The present method was validated in 11 volunteers by evaluating its pancreas coverage and repeatability and by comparing it to MOLLI. Four pancreatitis patients were evaluated using the proposed method and clinical scans.</p><p><strong>Results: </strong>The phantom wT<sub>1</sub> results are in better agreement to MRS ( <math> <mrow><msub><mtext>wT</mtext> <mn>1</mn></msub> <mo>=</mo> <mn>1.02</mn> <msub><mtext>*wT</mtext> <mrow><mn>1</mn> <mtext>MRS</mtext></mrow> </msub> <mo>-</mo> <mn>47.81</mn> <mo>,</mo> <mrow></mrow> <msup><mi>R</mi> <mn>2</mn></msup> <mrow><mo>=</mo> <mn>0.99</mn> <mo>)</mo> <mrow></mrow></mrow> </mrow> </math> than MOLLI ( <math> <mrow><msub><mtext>T</mtext> <mrow><mn>1</mn> <mtext>MOLLI</mtext></mrow> </msub> <mo>=</mo> <mn>1.13</mn> <mrow></mrow> <mo>∗</mo> <msub><mtext>wT</mtext> <mrow><mn>1</mn> <mtext>MRS</mtext></mrow> </msub> <mo>-</mo> <mn>74.65</mn> <mo>,</mo> <msup><mi>R</mi> <mn>2</mn></msup> <mrow><mo>=</mo> <mn>0.98</mn> <mo>)</mo></mrow> </mrow> </math> . The volunteer wT<sub>1</sub> results demonstrate the whole pancreas coverage capability for different fat fractions, good repeatability ( <math> <mrow><mmultiscripts><mtext>wT</mtext> <mrow><mn>1</mn> <mo>,</mo> <mrow></mrow> <msup><mn>2</mn> <mo>∘</mo></msup> </mrow> <mrow></mrow></mmultiscripts> <mo>=</mo> <mn>0.98</mn> <mrow></mrow> <mo>∗</mo> <mmultiscripts><mtext>wT</mtext> <mrow><mn>1</mn> <mo>,</mo> <mrow></mrow> <msup><mn>1</mn> <mo>∘</mo></msup> </mrow> <mrow></mrow></mmultiscripts> <mo>+</mo> <mn>17.40</mn> <mo>,</mo> <msup><mi>R</mi> <mn>2</mn></msup> <mrow><mo>=</mo> <mn>0.69</mn> <mo>)</mo></mrow> </mrow> </math> and lower T<sub>1</sub> values than MOLLI ( <math> <mrow><msub><mtext>wT</mtext> <mn>1</mn></msub> <mo>=</mo> <mn>0.34</mn> <msub><mtext>*T</mtext> <mrow><mn>1</mn> <mtext>MOLLI</mtext></mrow> </msub> <mo>+</mo> <mn>383.65</mn> <mo>,</mo> <msup><mrow><mspace></mspace> <mtext>R</mtext></mrow> <mn>2</mn></msup> <mrow><mo>=</mo> <mn>0.26</mn> <mo>)</mo></mrow> </mrow> </math> . The wT<sub>1</sub> maps in patients captured diverse pancreatitis regions with higher values <math><mrow><mo>(</mo> <msub><mtext>wT</mtext> <mrow><mn>1</mn> <mtext>Patients</mtext></mrow> </msub> <mo>=</mo> <mfenced><mrow><mn>831</mn> <mo>-</mo> <mn>1696</mn></mrow> </mfenced> <mspace></mspace> <mtext>ms</mtext> <mo>)</mo></mrow> </math> than in the volunteers <math> <mfenced> <mrow><msub><mtext>wT</mtext> <mrow><mn>1</mn> <mtext>Volunteers</mtext></mrow> </msub> <mo>=</mo> <mfenced><mrow><mn>605</mn> <mo>-</mo> <mn>799</mn></mrow> </mfenced> <mspace></mspace> <mtext>ms</mtext></mrow> </mfenced> </math> , thus showing their potential clinical feasibility.</p><p><strong>Conclusion: </strong>The present work proposes a wT<sub>1</sub> mapping methodology of the whole pancreas at 3 T, where 24 slices ( <math><mrow><mn>2</mn> <mo>×</mo> <mn>2</mn> <mo>×</mo> <mn>5</mn> <mspace></mspace> <msup><mrow><mtext>mm</mtext></mrow> <mn>3</mn></msup> </mrow> </math> ) were acquired in three short breath-holds of 12 s each.</p>","PeriodicalId":18067,"journal":{"name":"Magnetic Resonance Materials in Physics, Biology and Medicine","volume":" ","pages":"271-283"},"PeriodicalIF":2.0000,"publicationDate":"2025-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11913948/pdf/","citationCount":"0","resultStr":"{\"title\":\"Whole pancreas water T<sub>1</sub> mapping at 3 Tesla.\",\"authors\":\"Elizabeth Huaroc Moquillaza, Kilian Weiss, Lisa Steinhelfer, Jonathan Stelter, Marcus R Makowski, Rickmer Braren, Mariya Doneva, Dimitrios C Karampinos\",\"doi\":\"10.1007/s10334-025-01224-8\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>A fast T<sub>1</sub> mapping method of the whole pancreas remains a challenge, due to the complex anatomy of the organ. In addition, a technique for pancreas water T<sub>1</sub> mapping is needed, since the T<sub>1</sub> is biased in the presence of fat. The purpose of this work is to accelerate the acquisition of water selective T<sub>1</sub> (wT<sub>1</sub>) mapping for the whole pancreas at 3 T.</p><p><strong>Methods: </strong>The proposed method combines a continuous inversion-recovery Look-Locker acquisition with a single-shot gradient echo spiral readout, water-fat separation and dictionary matching for wT<sub>1</sub> mapping of the whole pancreas at 3 T. The bias of T<sub>1</sub> in the presence of fat was evaluated in a phantom by comparing the modified Look-Locker inversion-recovery (MOLLI) and the proposed method to MRS measurements. The present method was validated in 11 volunteers by evaluating its pancreas coverage and repeatability and by comparing it to MOLLI. Four pancreatitis patients were evaluated using the proposed method and clinical scans.</p><p><strong>Results: </strong>The phantom wT<sub>1</sub> results are in better agreement to MRS ( <math> <mrow><msub><mtext>wT</mtext> <mn>1</mn></msub> <mo>=</mo> <mn>1.02</mn> <msub><mtext>*wT</mtext> <mrow><mn>1</mn> <mtext>MRS</mtext></mrow> </msub> <mo>-</mo> <mn>47.81</mn> <mo>,</mo> <mrow></mrow> <msup><mi>R</mi> <mn>2</mn></msup> <mrow><mo>=</mo> <mn>0.99</mn> <mo>)</mo> <mrow></mrow></mrow> </mrow> </math> than MOLLI ( <math> <mrow><msub><mtext>T</mtext> <mrow><mn>1</mn> <mtext>MOLLI</mtext></mrow> </msub> <mo>=</mo> <mn>1.13</mn> <mrow></mrow> <mo>∗</mo> <msub><mtext>wT</mtext> <mrow><mn>1</mn> <mtext>MRS</mtext></mrow> </msub> <mo>-</mo> <mn>74.65</mn> <mo>,</mo> <msup><mi>R</mi> <mn>2</mn></msup> <mrow><mo>=</mo> <mn>0.98</mn> <mo>)</mo></mrow> </mrow> </math> . The volunteer wT<sub>1</sub> results demonstrate the whole pancreas coverage capability for different fat fractions, good repeatability ( <math> <mrow><mmultiscripts><mtext>wT</mtext> <mrow><mn>1</mn> <mo>,</mo> <mrow></mrow> <msup><mn>2</mn> <mo>∘</mo></msup> </mrow> <mrow></mrow></mmultiscripts> <mo>=</mo> <mn>0.98</mn> <mrow></mrow> <mo>∗</mo> <mmultiscripts><mtext>wT</mtext> <mrow><mn>1</mn> <mo>,</mo> <mrow></mrow> <msup><mn>1</mn> <mo>∘</mo></msup> </mrow> <mrow></mrow></mmultiscripts> <mo>+</mo> <mn>17.40</mn> <mo>,</mo> <msup><mi>R</mi> <mn>2</mn></msup> <mrow><mo>=</mo> <mn>0.69</mn> <mo>)</mo></mrow> </mrow> </math> and lower T<sub>1</sub> values than MOLLI ( <math> <mrow><msub><mtext>wT</mtext> <mn>1</mn></msub> <mo>=</mo> <mn>0.34</mn> <msub><mtext>*T</mtext> <mrow><mn>1</mn> <mtext>MOLLI</mtext></mrow> </msub> <mo>+</mo> <mn>383.65</mn> <mo>,</mo> <msup><mrow><mspace></mspace> <mtext>R</mtext></mrow> <mn>2</mn></msup> <mrow><mo>=</mo> <mn>0.26</mn> <mo>)</mo></mrow> </mrow> </math> . The wT<sub>1</sub> maps in patients captured diverse pancreatitis regions with higher values <math><mrow><mo>(</mo> <msub><mtext>wT</mtext> <mrow><mn>1</mn> <mtext>Patients</mtext></mrow> </msub> <mo>=</mo> <mfenced><mrow><mn>831</mn> <mo>-</mo> <mn>1696</mn></mrow> </mfenced> <mspace></mspace> <mtext>ms</mtext> <mo>)</mo></mrow> </math> than in the volunteers <math> <mfenced> <mrow><msub><mtext>wT</mtext> <mrow><mn>1</mn> <mtext>Volunteers</mtext></mrow> </msub> <mo>=</mo> <mfenced><mrow><mn>605</mn> <mo>-</mo> <mn>799</mn></mrow> </mfenced> <mspace></mspace> <mtext>ms</mtext></mrow> </mfenced> </math> , thus showing their potential clinical feasibility.</p><p><strong>Conclusion: </strong>The present work proposes a wT<sub>1</sub> mapping methodology of the whole pancreas at 3 T, where 24 slices ( <math><mrow><mn>2</mn> <mo>×</mo> <mn>2</mn> <mo>×</mo> <mn>5</mn> <mspace></mspace> <msup><mrow><mtext>mm</mtext></mrow> <mn>3</mn></msup> </mrow> </math> ) were acquired in three short breath-holds of 12 s each.</p>\",\"PeriodicalId\":18067,\"journal\":{\"name\":\"Magnetic Resonance Materials in Physics, Biology and Medicine\",\"volume\":\" \",\"pages\":\"271-283\"},\"PeriodicalIF\":2.0000,\"publicationDate\":\"2025-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11913948/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Magnetic Resonance Materials in Physics, Biology and Medicine\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1007/s10334-025-01224-8\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2025/3/6 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic Resonance Materials in Physics, Biology and Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10334-025-01224-8","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/3/6 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

目的:由于胰腺解剖结构复杂,整个胰腺的快速T1定位方法仍然是一个挑战。此外,胰脏水T1成像技术是必要的,因为T1在脂肪的存在下是有偏差的。这项工作的目的是加速获得全胰腺在3 t时的水选择性T1 (wT1)图谱。所提出的方法结合了连续的反转恢复Look-Locker采集和单次梯度回波螺旋读数,水-脂肪分离和字典匹配,以在3 t时对整个胰腺进行wT1映射。通过将改进的Look-Locker反转恢复(MOLLI)和所提出的方法与MRS测量结果进行比较,在幻影中评估脂肪存在时T1的偏差。通过评估其胰腺覆盖范围和可重复性,并将其与MOLLI进行比较,在11名志愿者中验证了该方法。使用所提出的方法和临床扫描对4例胰腺炎患者进行了评估。结果:幻影wT1结果与MRS (wT1 = 1.02 *wT 1 MRS - 47.81, r2 = 0.99)的一致性优于MOLLI (t1 MOLLI = 1.13 *wT 1 MRS - 74.65, r2 = 0.98)。志愿者的wT1结果显示了不同脂肪组分的全胰腺覆盖能力,重复性好(wT 1,2°= 0.98∗wT 1,2°+ 17.40,r2 = 0.69), T1值低于MOLLI (wT1 = 0.34 * T1 MOLLI + 383.65, r2 = 0.26)。患者的wT1图谱捕获不同胰腺炎区域的值(wT1 patients = 831 - 1696 ms)高于志愿者(wT1 volunteers = 605 - 799 ms),显示了其潜在的临床可行性。结论:本工作提出了一种3t全胰腺的wT1制图方法,其中在3次短屏气(每次12 s)中获得24片(2 × 2 × 5 mm 3)。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Whole pancreas water T1 mapping at 3 Tesla.

Purpose: A fast T1 mapping method of the whole pancreas remains a challenge, due to the complex anatomy of the organ. In addition, a technique for pancreas water T1 mapping is needed, since the T1 is biased in the presence of fat. The purpose of this work is to accelerate the acquisition of water selective T1 (wT1) mapping for the whole pancreas at 3 T.

Methods: The proposed method combines a continuous inversion-recovery Look-Locker acquisition with a single-shot gradient echo spiral readout, water-fat separation and dictionary matching for wT1 mapping of the whole pancreas at 3 T. The bias of T1 in the presence of fat was evaluated in a phantom by comparing the modified Look-Locker inversion-recovery (MOLLI) and the proposed method to MRS measurements. The present method was validated in 11 volunteers by evaluating its pancreas coverage and repeatability and by comparing it to MOLLI. Four pancreatitis patients were evaluated using the proposed method and clinical scans.

Results: The phantom wT1 results are in better agreement to MRS ( wT 1 = 1.02 *wT 1 MRS - 47.81 , R 2 = 0.99 ) than MOLLI ( T 1 MOLLI = 1.13 wT 1 MRS - 74.65 , R 2 = 0.98 ) . The volunteer wT1 results demonstrate the whole pancreas coverage capability for different fat fractions, good repeatability ( wT 1 , 2 = 0.98 wT 1 , 1 + 17.40 , R 2 = 0.69 ) and lower T1 values than MOLLI ( wT 1 = 0.34 *T 1 MOLLI + 383.65 , R 2 = 0.26 ) . The wT1 maps in patients captured diverse pancreatitis regions with higher values ( wT 1 Patients = 831 - 1696 ms ) than in the volunteers wT 1 Volunteers = 605 - 799 ms , thus showing their potential clinical feasibility.

Conclusion: The present work proposes a wT1 mapping methodology of the whole pancreas at 3 T, where 24 slices ( 2 × 2 × 5 mm 3 ) were acquired in three short breath-holds of 12 s each.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
4.60
自引率
0.00%
发文量
58
审稿时长
>12 weeks
期刊介绍: MAGMA is a multidisciplinary international journal devoted to the publication of articles on all aspects of magnetic resonance techniques and their applications in medicine and biology. MAGMA currently publishes research papers, reviews, letters to the editor, and commentaries, six times a year. The subject areas covered by MAGMA include: advances in materials, hardware and software in magnetic resonance technology, new developments and results in research and practical applications of magnetic resonance imaging and spectroscopy related to biology and medicine, study of animal models and intact cells using magnetic resonance, reports of clinical trials on humans and clinical validation of magnetic resonance protocols.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信