IF 2.5 2区 医学 Q1 ORTHOPEDICS
Lise-Lott Larsson, Johan Nordenadler, Gunilla Björling, Li Felländer-Tsai, Stergios Lazarinis, Bengt Ljungqvist, Janet Mattsson, Berit Reinmüller, Harald Brismar
{"title":"Correlation between a real-time bioparticle detection device and a traditional microbiological active air sampler monitoring air quality in an operating room during elective arthroplasty surgery: a prospective feasibility study.","authors":"Lise-Lott Larsson, Johan Nordenadler, Gunilla Björling, Li Felländer-Tsai, Stergios Lazarinis, Bengt Ljungqvist, Janet Mattsson, Berit Reinmüller, Harald Brismar","doi":"10.2340/17453674.2025.43002","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and purpose: </strong> The standard method for controlling operating room (OR) air quality is measuring bacteria-carrying particles per volume unit of air: colony forming units (CFU/m3). The result takes at least 2 days after sampling. Another method is real-time measurements of fluorescing bioparticles per unit volume of air (FBP/dm3). We aimed to compare simultaneous measurements of FBP/50 dm3 and CFU/m3 during ongoing arthroplasty surgery.</p><p><strong>Methods: </strong> 18 arthroplasties were performed in a modern OR with turbulent mixed airflow ventilation. The sampling heads of a BioAerosol Monitoring System (BAMS) and a microbiological active air sampler (Sartorius MD8 Air Sampler) were placed next to each other, and 6 parallel 10-minute registrations of FBP/50 dm3 and CFU/m3 were performed for each surgery. Parallel measurements were plotted against each other, Passing-Bablok nonparametric linear regression was performed, and the Spearman correlation coefficient (r) was calculated.</p><p><strong>Results: </strong> The r between FBP ≥ 3 μm/50 dm3 and CFU/m3 sampled for 96 x 10-minute intervals, was 0.70 (95% confidence interval [CI] 0.57-0.79). In the 25th percentile with the lowest 10-minute FBP ≥ 3μm/50 dm3, there were no CFU measurements with ≥ 10 and 4% with ≥ 5 CFU/m3. In the 75th percentile with the highest 10-minute FBP ≥ 3 μm/50 dm3, there were 58% CFU measurements with ≥ 10 and 88% with ≥ 5 CFU/m3. The r between FBP ≥ 3 μm/50 dm3 and CFU/m3 means sampled during 18 operations was 0.87 (CI 0.68-0.95).</p><p><strong>Conclusion: </strong> Low FBP ≥ 3 μm/50 dm3 measured by BAMS indicates low CFU/m3; conversely, high FBP ≥ 3 μm/50 dm3 indicates high CFU/m3. Real-time measurements of FBP ≥ 3 μm/50 dm3 can be used as a supplement to CFU/m3 monitoring OR air bacterial load.</p>","PeriodicalId":6916,"journal":{"name":"Acta Orthopaedica","volume":"96 ","pages":"176-181"},"PeriodicalIF":2.5000,"publicationDate":"2025-02-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11849163/pdf/","citationCount":"0","resultStr":null,"platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta Orthopaedica","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.2340/17453674.2025.43002","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

摘要

背景与目的:控制手术室空气质量的标准方法是测量每体积单位空气中携带细菌的颗粒:菌落形成单位(CFU/m3)。采样后至少需要2天才能得出结果。另一种方法是实时测量每单位空气体积(FBP/dm3)的荧光生物颗粒。我们的目的是比较正在进行的关节置换术中同时测量的FBP/50 dm3和CFU/m3。方法:18例关节置换术在现代手术室采用湍流混合气流通气。生物气溶胶监测系统(BAMS)和微生物活性空气采样器(Sartorius MD8空气采样器)的采样头相邻放置,每次手术进行6次平行的10分钟FBP/50 dm3和CFU/m3登记。平行测量结果相互对照,进行pass - bablok非参数线性回归,计算Spearman相关系数(r)。结果:96 × 10min采样FBP≥3 μm/50 dm3与CFU/m3之间的r为0.70(95%置信区间[CI] 0.57 ~ 0.79)。在最低10分钟FBP≥3μm/50 dm3的第25百分位数中,CFU≥10的无CFU测量值,≥5 CFU/m3的有4%。在最高10分钟FBP≥3 μm/50 dm3的第75百分位数中,58%的CFU≥10,88%的CFU≥5 /m3。18次手术中FBP≥3 μm/50 dm3与CFU/m3平均值之间的r为0.87 (CI 0.68 ~ 0.95)。结论:BAMS测量低FBP≥3 μm/50 dm3提示低CFU/m3;反之,FBP≥3 μm/50 dm3, CFU/m3高。实时测量FBP≥3 μm/50 dm3可作为CFU/m3监测OR空气细菌负荷的补充。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Correlation between a real-time bioparticle detection device and a traditional microbiological active air sampler monitoring air quality in an operating room during elective arthroplasty surgery: a prospective feasibility study.

Background and purpose:  The standard method for controlling operating room (OR) air quality is measuring bacteria-carrying particles per volume unit of air: colony forming units (CFU/m3). The result takes at least 2 days after sampling. Another method is real-time measurements of fluorescing bioparticles per unit volume of air (FBP/dm3). We aimed to compare simultaneous measurements of FBP/50 dm3 and CFU/m3 during ongoing arthroplasty surgery.

Methods:  18 arthroplasties were performed in a modern OR with turbulent mixed airflow ventilation. The sampling heads of a BioAerosol Monitoring System (BAMS) and a microbiological active air sampler (Sartorius MD8 Air Sampler) were placed next to each other, and 6 parallel 10-minute registrations of FBP/50 dm3 and CFU/m3 were performed for each surgery. Parallel measurements were plotted against each other, Passing-Bablok nonparametric linear regression was performed, and the Spearman correlation coefficient (r) was calculated.

Results:  The r between FBP ≥ 3 μm/50 dm3 and CFU/m3 sampled for 96 x 10-minute intervals, was 0.70 (95% confidence interval [CI] 0.57-0.79). In the 25th percentile with the lowest 10-minute FBP ≥ 3μm/50 dm3, there were no CFU measurements with ≥ 10 and 4% with ≥ 5 CFU/m3. In the 75th percentile with the highest 10-minute FBP ≥ 3 μm/50 dm3, there were 58% CFU measurements with ≥ 10 and 88% with ≥ 5 CFU/m3. The r between FBP ≥ 3 μm/50 dm3 and CFU/m3 means sampled during 18 operations was 0.87 (CI 0.68-0.95).

Conclusion:  Low FBP ≥ 3 μm/50 dm3 measured by BAMS indicates low CFU/m3; conversely, high FBP ≥ 3 μm/50 dm3 indicates high CFU/m3. Real-time measurements of FBP ≥ 3 μm/50 dm3 can be used as a supplement to CFU/m3 monitoring OR air bacterial load.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Acta Orthopaedica
Acta Orthopaedica 医学-整形外科
CiteScore
6.40
自引率
8.10%
发文量
105
审稿时长
4-8 weeks
期刊介绍: Acta Orthopaedica (previously Acta Orthopaedica Scandinavica) presents original articles of basic research interest, as well as clinical studies in the field of orthopedics and related sub disciplines. Ever since the journal was founded in 1930, by a group of Scandinavian orthopedic surgeons, the journal has been published for an international audience. Acta Orthopaedica is owned by the Nordic Orthopaedic Federation and is the official publication of this federation.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信