{"title":"在酶和钒酸盐氧化试验中,依曲巴格诱导的胆红素浓度错误升高:血清依曲巴格高浓度的介入。","authors":"Shun Oda, Kosuke Doki, Naoshi Obara, Yoshiharu Suzuki, Shigeru Chiba, Masato Homma","doi":"10.1097/FTD.0000000000001307","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The administration of eltrombopag, used to restore low blood count, demonstrates a positive interference of blood bilirubin levels when analyzed through the diazo assay. However, research on bilirubin measurements using other methods is limited. Therefore, using an enzymatic assay, this study aimed to investigate the effect of serum eltrombopag on bilirubin measurements in patients with aplastic anemia. It further assessed the concentration-dependent effect of eltrombopag on bilirubin measurements using enzymatic and vanadate oxidation assays.</p><p><strong>Methods: </strong>Total and conjugated bilirubin concentrations measured using an enzymatic assay and serum eltrombopag concentrations were examined in 227 serum samples collected from 30 patients with aplastic anemia receiving eltrombopag. Eltrombopag-spiked samples were analyzed using the enzymatic and vanadate oxidation assays for total and conjugated bilirubin to determine its concentration-dependent effects.</p><p><strong>Results: </strong>A strong positive correlation was observed between total bilirubin and serum eltrombopag concentrations in patients receiving eltrombopag (r = 0.820). However, the correlation between conjugated bilirubin and serum eltrombopag concentrations was weaker (r = 0.413). In eltrombopag-spiked serum samples, the enzymatic assay showed significant false elevation of total bilirubin concentrations at ≥6.0 mcg/mL; no interference with conjugated bilirubin measurements was observed. The vanadate oxidation assay showed mild positive biases of 0.2 and 0.1 mg/dL for total and conjugated bilirubin concentrations, respectively, at a high eltrombopag concentration (50 mcg/mL).</p><p><strong>Conclusions: </strong>Eltrombopag causes clinically significant concentration-dependent interference in total blood bilirubin, but not in conjugated bilirubin measurements through the enzymatic assay in patients with aplastic anemia. The vanadate oxidation assay may be used as an alternative to measure total blood bilirubin when the eltrombopag concentration is below 50 mcg/mL.</p>","PeriodicalId":23052,"journal":{"name":"Therapeutic Drug Monitoring","volume":" ","pages":""},"PeriodicalIF":2.8000,"publicationDate":"2025-02-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Eltrombopag-Induced Falsely Elevated Bilirubin Concentrations in Enzymatic and Vanadate Oxidation Assays: Involvement of High Serum Eltrombopag Concentrations.\",\"authors\":\"Shun Oda, Kosuke Doki, Naoshi Obara, Yoshiharu Suzuki, Shigeru Chiba, Masato Homma\",\"doi\":\"10.1097/FTD.0000000000001307\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The administration of eltrombopag, used to restore low blood count, demonstrates a positive interference of blood bilirubin levels when analyzed through the diazo assay. However, research on bilirubin measurements using other methods is limited. Therefore, using an enzymatic assay, this study aimed to investigate the effect of serum eltrombopag on bilirubin measurements in patients with aplastic anemia. It further assessed the concentration-dependent effect of eltrombopag on bilirubin measurements using enzymatic and vanadate oxidation assays.</p><p><strong>Methods: </strong>Total and conjugated bilirubin concentrations measured using an enzymatic assay and serum eltrombopag concentrations were examined in 227 serum samples collected from 30 patients with aplastic anemia receiving eltrombopag. Eltrombopag-spiked samples were analyzed using the enzymatic and vanadate oxidation assays for total and conjugated bilirubin to determine its concentration-dependent effects.</p><p><strong>Results: </strong>A strong positive correlation was observed between total bilirubin and serum eltrombopag concentrations in patients receiving eltrombopag (r = 0.820). However, the correlation between conjugated bilirubin and serum eltrombopag concentrations was weaker (r = 0.413). In eltrombopag-spiked serum samples, the enzymatic assay showed significant false elevation of total bilirubin concentrations at ≥6.0 mcg/mL; no interference with conjugated bilirubin measurements was observed. The vanadate oxidation assay showed mild positive biases of 0.2 and 0.1 mg/dL for total and conjugated bilirubin concentrations, respectively, at a high eltrombopag concentration (50 mcg/mL).</p><p><strong>Conclusions: </strong>Eltrombopag causes clinically significant concentration-dependent interference in total blood bilirubin, but not in conjugated bilirubin measurements through the enzymatic assay in patients with aplastic anemia. The vanadate oxidation assay may be used as an alternative to measure total blood bilirubin when the eltrombopag concentration is below 50 mcg/mL.</p>\",\"PeriodicalId\":23052,\"journal\":{\"name\":\"Therapeutic Drug Monitoring\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":2.8000,\"publicationDate\":\"2025-02-05\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Therapeutic Drug Monitoring\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1097/FTD.0000000000001307\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Therapeutic Drug Monitoring","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1097/FTD.0000000000001307","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

背景:通过重氮试验分析,用于恢复低血细胞计数的伊曲波帕的管理显示出对血液胆红素水平的积极干扰。然而,使用其他方法测量胆红素的研究是有限的。因此,本研究采用酶促测定法,旨在探讨血清eltrombopag对再生障碍性贫血患者胆红素测定的影响。它进一步评估了使用酶和钒酸盐氧化法测定的电曲巴格对胆红素测量的浓度依赖效应。方法:对30例接受依曲波帕治疗的再生障碍性贫血患者采集的227份血清样本,采用酶法测定总胆红素和结合胆红素浓度,并检测依曲波帕的血清浓度。用酶和钒酸盐氧化法对总胆红素和共轭胆红素进行分析,以确定其浓度依赖性。结果:接受依曲巴格治疗的患者血清总胆红素与依曲巴格浓度呈正相关(r = 0.820)。然而,结合胆红素与血清eltrombopag浓度的相关性较弱(r = 0.413)。在添加了电曲巴格的血清样品中,酶促分析显示总胆红素浓度在≥6.0 mcg/mL时显着虚假升高;未观察到对共轭胆红素测量的干扰。钒酸盐氧化试验显示,在高浓度(50 mcg/mL)下,总胆红素浓度和偶联胆红素浓度分别为0.2和0.1 mg/dL。结论:依曲巴格对再生障碍性贫血患者的总血胆红素产生临床显著的浓度依赖性干扰,但对酶法测定的结合胆红素没有影响。钒酸盐氧化试验可作为替代方法,用于测量总血胆红素,当电曲巴格浓度低于50微克/毫升。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Eltrombopag-Induced Falsely Elevated Bilirubin Concentrations in Enzymatic and Vanadate Oxidation Assays: Involvement of High Serum Eltrombopag Concentrations.

Background: The administration of eltrombopag, used to restore low blood count, demonstrates a positive interference of blood bilirubin levels when analyzed through the diazo assay. However, research on bilirubin measurements using other methods is limited. Therefore, using an enzymatic assay, this study aimed to investigate the effect of serum eltrombopag on bilirubin measurements in patients with aplastic anemia. It further assessed the concentration-dependent effect of eltrombopag on bilirubin measurements using enzymatic and vanadate oxidation assays.

Methods: Total and conjugated bilirubin concentrations measured using an enzymatic assay and serum eltrombopag concentrations were examined in 227 serum samples collected from 30 patients with aplastic anemia receiving eltrombopag. Eltrombopag-spiked samples were analyzed using the enzymatic and vanadate oxidation assays for total and conjugated bilirubin to determine its concentration-dependent effects.

Results: A strong positive correlation was observed between total bilirubin and serum eltrombopag concentrations in patients receiving eltrombopag (r = 0.820). However, the correlation between conjugated bilirubin and serum eltrombopag concentrations was weaker (r = 0.413). In eltrombopag-spiked serum samples, the enzymatic assay showed significant false elevation of total bilirubin concentrations at ≥6.0 mcg/mL; no interference with conjugated bilirubin measurements was observed. The vanadate oxidation assay showed mild positive biases of 0.2 and 0.1 mg/dL for total and conjugated bilirubin concentrations, respectively, at a high eltrombopag concentration (50 mcg/mL).

Conclusions: Eltrombopag causes clinically significant concentration-dependent interference in total blood bilirubin, but not in conjugated bilirubin measurements through the enzymatic assay in patients with aplastic anemia. The vanadate oxidation assay may be used as an alternative to measure total blood bilirubin when the eltrombopag concentration is below 50 mcg/mL.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Therapeutic Drug Monitoring
Therapeutic Drug Monitoring 医学-毒理学
CiteScore
5.00
自引率
8.00%
发文量
213
审稿时长
4-8 weeks
期刊介绍: Therapeutic Drug Monitoring is a peer-reviewed, multidisciplinary journal directed to an audience of pharmacologists, clinical chemists, laboratorians, pharmacists, drug researchers and toxicologists. It fosters the exchange of knowledge among the various disciplines–clinical pharmacology, pathology, toxicology, analytical chemistry–that share a common interest in Therapeutic Drug Monitoring. The journal presents studies detailing the various factors that affect the rate and extent drugs are absorbed, metabolized, and excreted. Regular features include review articles on specific classes of drugs, original articles, case reports, technical notes, and continuing education articles.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信