{"title":"使用相衬3p-Dixon方法同时获取水、脂肪和速度图像。","authors":"Esteban Denecken ,&nbsp;Cristóbal Arrieta ,&nbsp;Diego Hernando ,&nbsp;Julio Sotelo ,&nbsp;Hernán Mella ,&nbsp;Sergio Uribe","doi":"10.1016/j.mri.2025.110341","DOIUrl":null,"url":null,"abstract":"<div><h3>Purpose</h3><div>Phase-contrast MRI (2D PC-MRI) and Dixon techniques share the characteristic that the difference in frequency between water and fat, as well as the velocity, are encoded in the phase of the MR signal. We propose to take advantage of this characteristic to obtain both sets of images simultaneously. Such an acquisition will improve efficiency by obtaining both types of images in the same scan and will provide co-registered images of water-fat species and velocity images. This, in turn, will correct fat artifacts due to chemical shift in PC-MRI based measurements.</div></div><div><h3>Methods</h3><div>This study presents a novel PC multi-echo (PCME-MRI) sequence jointly with a 3-point (3p-) Dixon pipeline that enables reconstruction of water, fat, and velocity images simultaneously. The proposed 3p-Dixon approach preserves the phase information of water-fat images, while velocity images are obtained from the resulting water components.</div></div><div><h3>Results</h3><div>Numerical phantom tests and 2D MR axial images of the neck acquired in 12 healthy volunteers demonstrated the feasibility of the PC 3p-Dixon method, showing comparable performance to standard techniques. In volunteers the median and range MAE comparing PC 3p-Dixon, and standard 3p-Dixon fat fraction were 0.06 and [0.03, 0.09]. The median and range of velocity for PC 3p-Dixon were 6.15 ml and [3.86, 7.21]ml, compared to 6.43 ml and [4.62, 8.27]ml obtained by 2D PC-MRI.</div></div><div><h3>Conclusion</h3><div>Numerical phantom experiments and acquisitions from healthy volunteers showed promising results in fat fraction and velocity estimation of PC 3p-Dixon compared with standard 3p-Dixon and 2D PC-MRI, obtaining both data sets in similar times as standard 3p Dixon.</div></div>","PeriodicalId":18165,"journal":{"name":"Magnetic resonance imaging","volume":"118 ","pages":"Article 110341"},"PeriodicalIF":2.1000,"publicationDate":"2025-01-31","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Simultaneous acquisition of water, fat and velocity images using a phase-contrast 3p-Dixon method\",\"authors\":\"Esteban Denecken ,&nbsp;Cristóbal Arrieta ,&nbsp;Diego Hernando ,&nbsp;Julio Sotelo ,&nbsp;Hernán Mella ,&nbsp;Sergio Uribe\",\"doi\":\"10.1016/j.mri.2025.110341\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Purpose</h3><div>Phase-contrast MRI (2D PC-MRI) and Dixon techniques share the characteristic that the difference in frequency between water and fat, as well as the velocity, are encoded in the phase of the MR signal. We propose to take advantage of this characteristic to obtain both sets of images simultaneously. Such an acquisition will improve efficiency by obtaining both types of images in the same scan and will provide co-registered images of water-fat species and velocity images. This, in turn, will correct fat artifacts due to chemical shift in PC-MRI based measurements.</div></div><div><h3>Methods</h3><div>This study presents a novel PC multi-echo (PCME-MRI) sequence jointly with a 3-point (3p-) Dixon pipeline that enables reconstruction of water, fat, and velocity images simultaneously. The proposed 3p-Dixon approach preserves the phase information of water-fat images, while velocity images are obtained from the resulting water components.</div></div><div><h3>Results</h3><div>Numerical phantom tests and 2D MR axial images of the neck acquired in 12 healthy volunteers demonstrated the feasibility of the PC 3p-Dixon method, showing comparable performance to standard techniques. In volunteers the median and range MAE comparing PC 3p-Dixon, and standard 3p-Dixon fat fraction were 0.06 and [0.03, 0.09]. The median and range of velocity for PC 3p-Dixon were 6.15 ml and [3.86, 7.21]ml, compared to 6.43 ml and [4.62, 8.27]ml obtained by 2D PC-MRI.</div></div><div><h3>Conclusion</h3><div>Numerical phantom experiments and acquisitions from healthy volunteers showed promising results in fat fraction and velocity estimation of PC 3p-Dixon compared with standard 3p-Dixon and 2D PC-MRI, obtaining both data sets in similar times as standard 3p Dixon.</div></div>\",\"PeriodicalId\":18165,\"journal\":{\"name\":\"Magnetic resonance imaging\",\"volume\":\"118 \",\"pages\":\"Article 110341\"},\"PeriodicalIF\":2.1000,\"publicationDate\":\"2025-01-31\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Magnetic resonance imaging\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0730725X25000232\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Magnetic resonance imaging","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0730725X25000232","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

目的:相位对比MRI (2D PC-MRI)和Dixon技术的共同特点是,水和脂肪之间的频率差异以及速度都被编码在MR信号的相位中。我们建议利用这一特性同时获得两组图像。通过在同一次扫描中获得两种类型的图像,这种采集将提高效率,并将提供水-脂肪种类和速度图像的共同配准图像。反过来,这将纠正基于PC-MRI测量的化学位移造成的脂肪伪影。方法:本研究提出了一种新的PC多回声(PCME-MRI)序列,结合3点(3p-) Dixon管道,可以同时重建水、脂肪和速度图像。提出的3p-Dixon方法保留了水-脂肪图像的相位信息,同时从所得的水成分中获得速度图像。结果:12名健康志愿者的数值幻像测试和颈部二维MR轴向图像证明了PC 3p-Dixon方法的可行性,表现出与标准技术相当的性能。在志愿者中,比较PC 3p-Dixon脂肪分数和标准3p-Dixon脂肪分数的MAE中位数和范围分别为0.06和[0.03,0.09]。pcp3 - dixon的速度中位数和范围分别为6.15 ml和[3.86,7.21]ml,而2D PC- mri的速度中位数和范围分别为6.43 ml和[4.62,8.27]ml。结论:与标准3p-Dixon和2D PC- mri相比,从健康志愿者的数值模拟实验和获取的PC 3p-Dixon的脂肪含量和速度估计结果令人满意,获得的数据集与标准3p-Dixon的时间相似。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Simultaneous acquisition of water, fat and velocity images using a phase-contrast 3p-Dixon method

Purpose

Phase-contrast MRI (2D PC-MRI) and Dixon techniques share the characteristic that the difference in frequency between water and fat, as well as the velocity, are encoded in the phase of the MR signal. We propose to take advantage of this characteristic to obtain both sets of images simultaneously. Such an acquisition will improve efficiency by obtaining both types of images in the same scan and will provide co-registered images of water-fat species and velocity images. This, in turn, will correct fat artifacts due to chemical shift in PC-MRI based measurements.

Methods

This study presents a novel PC multi-echo (PCME-MRI) sequence jointly with a 3-point (3p-) Dixon pipeline that enables reconstruction of water, fat, and velocity images simultaneously. The proposed 3p-Dixon approach preserves the phase information of water-fat images, while velocity images are obtained from the resulting water components.

Results

Numerical phantom tests and 2D MR axial images of the neck acquired in 12 healthy volunteers demonstrated the feasibility of the PC 3p-Dixon method, showing comparable performance to standard techniques. In volunteers the median and range MAE comparing PC 3p-Dixon, and standard 3p-Dixon fat fraction were 0.06 and [0.03, 0.09]. The median and range of velocity for PC 3p-Dixon were 6.15 ml and [3.86, 7.21]ml, compared to 6.43 ml and [4.62, 8.27]ml obtained by 2D PC-MRI.

Conclusion

Numerical phantom experiments and acquisitions from healthy volunteers showed promising results in fat fraction and velocity estimation of PC 3p-Dixon compared with standard 3p-Dixon and 2D PC-MRI, obtaining both data sets in similar times as standard 3p Dixon.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Magnetic resonance imaging
Magnetic resonance imaging 医学-核医学
CiteScore
4.70
自引率
4.00%
发文量
194
审稿时长
83 days
期刊介绍: Magnetic Resonance Imaging (MRI) is the first international multidisciplinary journal encompassing physical, life, and clinical science investigations as they relate to the development and use of magnetic resonance imaging. MRI is dedicated to both basic research, technological innovation and applications, providing a single forum for communication among radiologists, physicists, chemists, biochemists, biologists, engineers, internists, pathologists, physiologists, computer scientists, and mathematicians.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信