{"title":"螺纹插入式无铅起搏器植入过程中出现症状性室上性心动过速的外部直流电击1例报告。","authors":"Mattia Liccardo, Ersilia Cipolletta, Emma Arezzi, Monica Sicuranza, Maria Rea, Pietro Ricciardi, Giovanni Napolitano","doi":"10.1111/pace.15134","DOIUrl":null,"url":null,"abstract":"<p><p>We present the case of a 64-year-old man who, during the implantation of an active-fixation leadless pacemaker (LP, Aveir VR, Abbott, USA), underwent several external defibrillation shocks up to 240 Joules, due to symptomatic sustained supraventricular tachycardia at 160 bpm. The shocks, delivered both before and after the screwing of the device in the low interventricular septum, did not cause any technical damage to the device, and no complications were observed. The device was then deployed successfully. To our knowledge, this is the first documented case of external cardioversion (ECV) in a patient with an active-fixation LP. The results suggest that ECV in these patients appears to be safe and feasible.</p>","PeriodicalId":54653,"journal":{"name":"Pace-Pacing and Clinical Electrophysiology","volume":" ","pages":"315-319"},"PeriodicalIF":1.7000,"publicationDate":"2025-03-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"External DC Shocks of Symptomatic Supraventricular Tachycardia During Screw-In Leadless Pacemaker Implant: A Case Report.\",\"authors\":\"Mattia Liccardo, Ersilia Cipolletta, Emma Arezzi, Monica Sicuranza, Maria Rea, Pietro Ricciardi, Giovanni Napolitano\",\"doi\":\"10.1111/pace.15134\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>We present the case of a 64-year-old man who, during the implantation of an active-fixation leadless pacemaker (LP, Aveir VR, Abbott, USA), underwent several external defibrillation shocks up to 240 Joules, due to symptomatic sustained supraventricular tachycardia at 160 bpm. The shocks, delivered both before and after the screwing of the device in the low interventricular septum, did not cause any technical damage to the device, and no complications were observed. The device was then deployed successfully. To our knowledge, this is the first documented case of external cardioversion (ECV) in a patient with an active-fixation LP. The results suggest that ECV in these patients appears to be safe and feasible.</p>\",\"PeriodicalId\":54653,\"journal\":{\"name\":\"Pace-Pacing and Clinical Electrophysiology\",\"volume\":\" \",\"pages\":\"315-319\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2025-03-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Pace-Pacing and Clinical Electrophysiology\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1111/pace.15134\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2025/1/10 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pace-Pacing and Clinical Electrophysiology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.15134","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2025/1/10 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
External DC Shocks of Symptomatic Supraventricular Tachycardia During Screw-In Leadless Pacemaker Implant: A Case Report.
We present the case of a 64-year-old man who, during the implantation of an active-fixation leadless pacemaker (LP, Aveir VR, Abbott, USA), underwent several external defibrillation shocks up to 240 Joules, due to symptomatic sustained supraventricular tachycardia at 160 bpm. The shocks, delivered both before and after the screwing of the device in the low interventricular septum, did not cause any technical damage to the device, and no complications were observed. The device was then deployed successfully. To our knowledge, this is the first documented case of external cardioversion (ECV) in a patient with an active-fixation LP. The results suggest that ECV in these patients appears to be safe and feasible.
期刊介绍:
Pacing and Clinical Electrophysiology (PACE) is the foremost peer-reviewed journal in the field of pacing and implantable cardioversion defibrillation, publishing over 50% of all English language articles in its field, featuring original, review, and didactic papers, and case reports related to daily practice. Articles also include editorials, book reviews, Musings on humane topics relevant to medical practice, electrophysiology (EP) rounds, device rounds, and information concerning the quality of devices used in the practice of the specialty.