利用环形线技术释放楔形对接按钮,用标准取物导管取出Aveir无铅起搏器。

IF 2.3 3区 医学 Q2 CARDIAC & CARDIOVASCULAR SYSTEMS
Kensuke Yano, Masaya Shinohara, Hideto Sasaki, Takuto Suzuki, Hosei Kikushima, Ryo Wada, Shintaro Yao, Hideki Koike, Tadashi Fujino, Takanori Ikeda
{"title":"利用环形线技术释放楔形对接按钮,用标准取物导管取出Aveir无铅起搏器。","authors":"Kensuke Yano, Masaya Shinohara, Hideto Sasaki, Takuto Suzuki, Hosei Kikushima, Ryo Wada, Shintaro Yao, Hideki Koike, Tadashi Fujino, Takanori Ikeda","doi":"10.1111/jce.16556","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>The docking button of the Aveir leadless pacemaker (LP) is occasionally difficult to access due to its inappropriate position in the right ventricle (RV).</p><p><strong>Methods and results: </strong>We report a case where the Aveir LP was successfully retrieved by releasing the wedged docking button in the inferior wall of the RV using a loop wire technique.</p><p><strong>Conclusion: </strong>The loop wire technique may be useful to change the position of the Aveir LP. This may be helpful to retrieve the Aveir LP even in a case where the docking button is wedged in the inferior wall of the RV.</p>","PeriodicalId":15178,"journal":{"name":"Journal of Cardiovascular Electrophysiology","volume":" ","pages":""},"PeriodicalIF":2.3000,"publicationDate":"2025-01-07","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Retrieval of an Aveir Leadless Pacemaker With a Standard Retrieval Catheter by Releasing the Wedged Docking Button Using a Loop Wire Technique.\",\"authors\":\"Kensuke Yano, Masaya Shinohara, Hideto Sasaki, Takuto Suzuki, Hosei Kikushima, Ryo Wada, Shintaro Yao, Hideki Koike, Tadashi Fujino, Takanori Ikeda\",\"doi\":\"10.1111/jce.16556\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>The docking button of the Aveir leadless pacemaker (LP) is occasionally difficult to access due to its inappropriate position in the right ventricle (RV).</p><p><strong>Methods and results: </strong>We report a case where the Aveir LP was successfully retrieved by releasing the wedged docking button in the inferior wall of the RV using a loop wire technique.</p><p><strong>Conclusion: </strong>The loop wire technique may be useful to change the position of the Aveir LP. This may be helpful to retrieve the Aveir LP even in a case where the docking button is wedged in the inferior wall of the RV.</p>\",\"PeriodicalId\":15178,\"journal\":{\"name\":\"Journal of Cardiovascular Electrophysiology\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":2.3000,\"publicationDate\":\"2025-01-07\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Cardiovascular Electrophysiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1111/jce.16556\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/jce.16556","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

摘要

Aveir无导线起搏器(LP)的对接按钮由于其在右心室(RV)中的位置不合适,有时难以接近。方法和结果:我们报告了一例使用环丝技术通过释放右心室下壁的楔形对接按钮成功取出Aveir LP的病例。结论:环形钢丝技术可用于改变Aveir LP的位置。这可能有助于检索Aveir LP,即使在对接按钮楔入RV下壁的情况下。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Retrieval of an Aveir Leadless Pacemaker With a Standard Retrieval Catheter by Releasing the Wedged Docking Button Using a Loop Wire Technique.

Introduction: The docking button of the Aveir leadless pacemaker (LP) is occasionally difficult to access due to its inappropriate position in the right ventricle (RV).

Methods and results: We report a case where the Aveir LP was successfully retrieved by releasing the wedged docking button in the inferior wall of the RV using a loop wire technique.

Conclusion: The loop wire technique may be useful to change the position of the Aveir LP. This may be helpful to retrieve the Aveir LP even in a case where the docking button is wedged in the inferior wall of the RV.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
5.20
自引率
14.80%
发文量
433
审稿时长
3-6 weeks
期刊介绍: Journal of Cardiovascular Electrophysiology (JCE) keeps its readership well informed of the latest developments in the study and management of arrhythmic disorders. Edited by Bradley P. Knight, M.D., and a distinguished international editorial board, JCE is the leading journal devoted to the study of the electrophysiology of the heart.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信