{"title":"EPID三维剂量重建在乳腺癌调强放疗体内剂量验证中的初步应用。","authors":"Jie Dong ,&nbsp;Zhenghuan Li ,&nbsp;Wentao Huang ,&nbsp;Fantu Kong ,&nbsp;Luxi Chen ,&nbsp;Meifang Zhang ,&nbsp;Shen Huang ,&nbsp;Huamei Yan ,&nbsp;Xiangying Xu","doi":"10.1016/j.ejmp.2024.104884","DOIUrl":null,"url":null,"abstract":"<div><div>A preliminary study was conducted using electronic portal imaging device (EPID) based dose verification in pre-treatment and in vivo dose reconstruction modes for breast cancer intensity-modulated radiation therapy (IMRT) technique with known repositioning set-up errors. For 43 IMRT plans, the set-up errors were determined from 43 sets of EPID images and 258 sets of cone beam computed tomography images. In-house developed Edose software was used to reconstruct the dose distribution using the pre-treatment and on-treatment (in vivo) EPID acquired fluence maps. The maximum setup error was &lt; 3.5 mm. For 43 pre-treatment cases, the γ pass rate (3 %/3 mm) is 98.49 % ± 1.15 %. The chest wall target Δ<em>V</em><sub>98%P</sub>, Δ<em>V</em><sub>95%P</sub>, andΔ<em>V</em><sub>90%P</sub> are all &lt; 5 %, while the majority of the ipsilateral lung Δ<em>V</em><sub>5Gy</sub>, Δ<em>V</em><sub>20Gy</sub>, and Δ<em>V</em><sub>30Gy</sub> are also &lt; 5 %. For 258 in vivo cases, the γ pass rate is 90.98 % ± 6.53 %, with the chest wall target Δ<em>V</em><sub>90%P</sub> and ipsilateral lung Δ<em>V</em><sub>30Gy</sub> both &lt; 5 %, while the other volume differences all exceed 5 %. The γ pass rate for in vivo verification is significantly lower than pre-treatment values. Although the in vivo γ verification satisfies the medical physics requirements, the reconstructed coverage of the chest wall target is far below the clinical dosimetry requirements. In vivo 3D dose reconstruction directly predicts changes in the planning target volume to aid clinicians better understand the actual dose received by patients with intra-fractional motion and anatomical changes.</div></div>","PeriodicalId":56092,"journal":{"name":"Physica Medica-European Journal of Medical Physics","volume":"129 ","pages":"Article 104884"},"PeriodicalIF":3.3000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Preliminary application of EPID three-dimensional dose reconstruction in in vivo dose verification of breast cancer intensity-modulated radiation therapy\",\"authors\":\"Jie Dong ,&nbsp;Zhenghuan Li ,&nbsp;Wentao Huang ,&nbsp;Fantu Kong ,&nbsp;Luxi Chen ,&nbsp;Meifang Zhang ,&nbsp;Shen Huang ,&nbsp;Huamei Yan ,&nbsp;Xiangying Xu\",\"doi\":\"10.1016/j.ejmp.2024.104884\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><div>A preliminary study was conducted using electronic portal imaging device (EPID) based dose verification in pre-treatment and in vivo dose reconstruction modes for breast cancer intensity-modulated radiation therapy (IMRT) technique with known repositioning set-up errors. For 43 IMRT plans, the set-up errors were determined from 43 sets of EPID images and 258 sets of cone beam computed tomography images. In-house developed Edose software was used to reconstruct the dose distribution using the pre-treatment and on-treatment (in vivo) EPID acquired fluence maps. The maximum setup error was &lt; 3.5 mm. For 43 pre-treatment cases, the γ pass rate (3 %/3 mm) is 98.49 % ± 1.15 %. The chest wall target Δ<em>V</em><sub>98%P</sub>, Δ<em>V</em><sub>95%P</sub>, andΔ<em>V</em><sub>90%P</sub> are all &lt; 5 %, while the majority of the ipsilateral lung Δ<em>V</em><sub>5Gy</sub>, Δ<em>V</em><sub>20Gy</sub>, and Δ<em>V</em><sub>30Gy</sub> are also &lt; 5 %. For 258 in vivo cases, the γ pass rate is 90.98 % ± 6.53 %, with the chest wall target Δ<em>V</em><sub>90%P</sub> and ipsilateral lung Δ<em>V</em><sub>30Gy</sub> both &lt; 5 %, while the other volume differences all exceed 5 %. The γ pass rate for in vivo verification is significantly lower than pre-treatment values. Although the in vivo γ verification satisfies the medical physics requirements, the reconstructed coverage of the chest wall target is far below the clinical dosimetry requirements. In vivo 3D dose reconstruction directly predicts changes in the planning target volume to aid clinicians better understand the actual dose received by patients with intra-fractional motion and anatomical changes.</div></div>\",\"PeriodicalId\":56092,\"journal\":{\"name\":\"Physica Medica-European Journal of Medical Physics\",\"volume\":\"129 \",\"pages\":\"Article 104884\"},\"PeriodicalIF\":3.3000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Physica Medica-European Journal of Medical Physics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S1120179724013528\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physica Medica-European Journal of Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S1120179724013528","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

一项基于电子门静脉成像装置(EPID)的乳腺癌调强放疗(IMRT)技术治疗前和体内剂量重建模式剂量验证的初步研究,已知重新定位设置错误。对于43个IMRT计划,从43组EPID图像和258组锥束计算机断层图像中确定了设置误差。使用内部开发的Edose软件,利用处理前和处理中(体内)EPID获取的通量图重建剂量分布。最大设置误差为98%P, ΔV95%P、andΔV90%P均为5Gy, ΔV20Gy、ΔV30Gy均为90%P,同侧肺ΔV30Gy均为5Gy
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Preliminary application of EPID three-dimensional dose reconstruction in in vivo dose verification of breast cancer intensity-modulated radiation therapy
A preliminary study was conducted using electronic portal imaging device (EPID) based dose verification in pre-treatment and in vivo dose reconstruction modes for breast cancer intensity-modulated radiation therapy (IMRT) technique with known repositioning set-up errors. For 43 IMRT plans, the set-up errors were determined from 43 sets of EPID images and 258 sets of cone beam computed tomography images. In-house developed Edose software was used to reconstruct the dose distribution using the pre-treatment and on-treatment (in vivo) EPID acquired fluence maps. The maximum setup error was < 3.5 mm. For 43 pre-treatment cases, the γ pass rate (3 %/3 mm) is 98.49 % ± 1.15 %. The chest wall target ΔV98%P, ΔV95%P, andΔV90%P are all < 5 %, while the majority of the ipsilateral lung ΔV5Gy, ΔV20Gy, and ΔV30Gy are also < 5 %. For 258 in vivo cases, the γ pass rate is 90.98 % ± 6.53 %, with the chest wall target ΔV90%P and ipsilateral lung ΔV30Gy both < 5 %, while the other volume differences all exceed 5 %. The γ pass rate for in vivo verification is significantly lower than pre-treatment values. Although the in vivo γ verification satisfies the medical physics requirements, the reconstructed coverage of the chest wall target is far below the clinical dosimetry requirements. In vivo 3D dose reconstruction directly predicts changes in the planning target volume to aid clinicians better understand the actual dose received by patients with intra-fractional motion and anatomical changes.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
6.80
自引率
14.70%
发文量
493
审稿时长
78 days
期刊介绍: Physica Medica, European Journal of Medical Physics, publishing with Elsevier from 2007, provides an international forum for research and reviews on the following main topics: Medical Imaging Radiation Therapy Radiation Protection Measuring Systems and Signal Processing Education and training in Medical Physics Professional issues in Medical Physics.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信