{"title":"一种用于弯曲髓内钉角度矫正的张力系统:体外分析。","authors":"Anderson Freitas, Vincenzo Giordano, Welington Ferreira Nunes, Kaio da Silva Santana, Fernanda Moura Viana, Robinson Esteves Pires, Rodrigo Sattamini Pires E Albuquerque, Pedro José Labronici","doi":"10.1007/s00590-024-04125-z","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Although several techniques have been described for bent intramedullary nail removal, there is no universally accepted strategy. We hypothesized that a device based on the action principle of a three-point bend fixture could facilitate extraction of bent intramedullary nails; this paper describes its design and experimental testing.</p><p><strong>Methods: </strong>Five large synthetic left femurs and five steel intramedullary nails were used. The nails were bent in a manual hydraulic press and implanted into the model femurs, which had been previously sectioned as necessary. A device to correct the angular deformity imparted to the nails was designed, and a prototype fabricated from AISI 1045 steel. Specimens were secured in a vise and the device attached; manual tension was then applied to gradually correct the nail deformity. Plain radiographs were obtained, and the nails were removed. The model femurs and intramedullary nails were visually inspected for proper contact of the device tip, coronal and sagittal alignment, and points of failure.</p><p><strong>Results: </strong>Before device application, the angles of the specimens averaged 40.2° (38.27°-41.32°) in the coronal plane and 7.22° (8.5°-10.2°) in the sagittal plane. After device application, these angles averaged 15.04° (2.5°-32.3°) and 4.16° (0°-9.8°), respectively. There were no implant fractures, but in two models, peri-implant fractures (one catastrophic) occurred close to the proximal and distal locking screws.</p><p><strong>Conclusion: </strong>The proposed device for bent intramedullary nail extraction was effective, but homogeneous behavior in terms of angle correction was not observed in the tested specimens. Further development is needed.</p>","PeriodicalId":50484,"journal":{"name":"European Journal of Orthopaedic Surgery and Traumatology","volume":"35 1","pages":"50"},"PeriodicalIF":1.4000,"publicationDate":"2024-12-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A tension system for angular correction of bent intramedullary nails: in vitro analysis.\",\"authors\":\"Anderson Freitas, Vincenzo Giordano, Welington Ferreira Nunes, Kaio da Silva Santana, Fernanda Moura Viana, Robinson Esteves Pires, Rodrigo Sattamini Pires E Albuquerque, Pedro José Labronici\",\"doi\":\"10.1007/s00590-024-04125-z\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>Although several techniques have been described for bent intramedullary nail removal, there is no universally accepted strategy. We hypothesized that a device based on the action principle of a three-point bend fixture could facilitate extraction of bent intramedullary nails; this paper describes its design and experimental testing.</p><p><strong>Methods: </strong>Five large synthetic left femurs and five steel intramedullary nails were used. The nails were bent in a manual hydraulic press and implanted into the model femurs, which had been previously sectioned as necessary. A device to correct the angular deformity imparted to the nails was designed, and a prototype fabricated from AISI 1045 steel. Specimens were secured in a vise and the device attached; manual tension was then applied to gradually correct the nail deformity. Plain radiographs were obtained, and the nails were removed. The model femurs and intramedullary nails were visually inspected for proper contact of the device tip, coronal and sagittal alignment, and points of failure.</p><p><strong>Results: </strong>Before device application, the angles of the specimens averaged 40.2° (38.27°-41.32°) in the coronal plane and 7.22° (8.5°-10.2°) in the sagittal plane. After device application, these angles averaged 15.04° (2.5°-32.3°) and 4.16° (0°-9.8°), respectively. There were no implant fractures, but in two models, peri-implant fractures (one catastrophic) occurred close to the proximal and distal locking screws.</p><p><strong>Conclusion: </strong>The proposed device for bent intramedullary nail extraction was effective, but homogeneous behavior in terms of angle correction was not observed in the tested specimens. Further development is needed.</p>\",\"PeriodicalId\":50484,\"journal\":{\"name\":\"European Journal of Orthopaedic Surgery and Traumatology\",\"volume\":\"35 1\",\"pages\":\"50\"},\"PeriodicalIF\":1.4000,\"publicationDate\":\"2024-12-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"European Journal of Orthopaedic Surgery and Traumatology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/s00590-024-04125-z\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"ORTHOPEDICS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"European Journal of Orthopaedic Surgery and Traumatology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s00590-024-04125-z","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

摘要

目的:虽然有几种技术已经描述了弯曲髓内钉去除,没有普遍接受的策略。我们假设一种基于三点弯曲夹具作用原理的装置可以方便地拔出弯曲的髓内钉;本文介绍了该系统的设计和实验测试。方法:采用5根大型合成左股骨和5根钢髓内钉。在手动液压机中将钉子弯曲,并将其植入先前已根据需要进行切片的模型股骨中。设计了一种矫正钉角畸形的装置,并用AISI 1045钢制作了一个原型。将标本固定在虎钳中并连接装置;然后用手拉紧逐渐矫正指甲畸形。行x线平片检查,取下钉。目视检查模型股骨和髓内钉是否正确接触装置尖端,冠状和矢状对齐以及失效点。结果:应用器械前,标本冠状面平均角度为40.2°(38.27°~ 41.32°),矢状面平均角度为7.22°(8.5°~ 10.2°)。设备应用后,这些角度平均分别为15.04°(2.5°-32.3°)和4.16°(0°-9.8°)。没有种植体骨折,但在两个模型中,种植体周围骨折(一个灾难性的)发生在近端和远端锁定螺钉附近。结论:所提出的弯曲髓内钉拔除装置是有效的,但在角度矫正方面,在测试样本中没有观察到均匀的行为。需要进一步发展。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A tension system for angular correction of bent intramedullary nails: in vitro analysis.

Purpose: Although several techniques have been described for bent intramedullary nail removal, there is no universally accepted strategy. We hypothesized that a device based on the action principle of a three-point bend fixture could facilitate extraction of bent intramedullary nails; this paper describes its design and experimental testing.

Methods: Five large synthetic left femurs and five steel intramedullary nails were used. The nails were bent in a manual hydraulic press and implanted into the model femurs, which had been previously sectioned as necessary. A device to correct the angular deformity imparted to the nails was designed, and a prototype fabricated from AISI 1045 steel. Specimens were secured in a vise and the device attached; manual tension was then applied to gradually correct the nail deformity. Plain radiographs were obtained, and the nails were removed. The model femurs and intramedullary nails were visually inspected for proper contact of the device tip, coronal and sagittal alignment, and points of failure.

Results: Before device application, the angles of the specimens averaged 40.2° (38.27°-41.32°) in the coronal plane and 7.22° (8.5°-10.2°) in the sagittal plane. After device application, these angles averaged 15.04° (2.5°-32.3°) and 4.16° (0°-9.8°), respectively. There were no implant fractures, but in two models, peri-implant fractures (one catastrophic) occurred close to the proximal and distal locking screws.

Conclusion: The proposed device for bent intramedullary nail extraction was effective, but homogeneous behavior in terms of angle correction was not observed in the tested specimens. Further development is needed.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
3.00
自引率
5.90%
发文量
265
审稿时长
3-8 weeks
期刊介绍: The European Journal of Orthopaedic Surgery and Traumatology (EJOST) aims to publish high quality Orthopedic scientific work. The objective of our journal is to disseminate meaningful, impactful, clinically relevant work from each and every region of the world, that has the potential to change and or inform clinical practice.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信