{"title":"利用声辐射力成像技术预测体内高强度聚焦超声病灶。","authors":"Xinwang Shi, Fenglong Zhao, Lian Feng, Yijing Liu, Xiaowei Zhou","doi":"10.1002/mp.17564","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Background</h3>\n \n <p>One big challenge in the noninvasive high-intensity focused ultrasound (HIFU) surgery is that the location and shape of its focus is unpredictable at the preoperative stage due to the complexity of sound wave propagation. The Acoustic Radiation Force Impulse (ARFI) imaging is a potential solution to this problem, but artifacts resulting from shear wave propagation remain to be solved.</p>\n </section>\n \n <section>\n \n <h3> Purpose</h3>\n \n <p>In this study, we proposed avoiding those artefacts by applying the ARFI technique at a high imaging frame rate within a very short time before the shear waves start to propagate.</p>\n </section>\n \n <section>\n \n <h3> Methods</h3>\n \n <p>Using single transmission with a convex imaging probe, two ultrafast imaging modalities (the diverging wave and the wide beam), were developed in the ARFI framework, and their reliabilities were validated on a nylon string phantom by the centroid tracking method borrowed from ultrasound localization microscopy (ULM). The proposed ARFI method was tested on a clinically equivalent HIFU system under different acoustic radiation intensities by in-vitro, ex-vivo and in-vivo experiments. In three experimental scenarios, we delivered short HIFU stimulation pulses at varying acoustic powers to induce tissue motion within the focal region. At each experimental site, both diverging wave and wide-beam imaging techniques were employed for motion estimation. Based on the focus prediction derived from the motion estimation, HIFU ablation treatment was performed. The treated samples were then incised to examine the damaged areas. Additionally, ultrasound B-mode images were acquired before and after the procedure and saved for analysis.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Quantitative analysis showed that the ARFI with wide beam imaging was able to predict the HIFU focus preoperatively, only with 1 to 3 mm of errors in focal central location, and less than 23% of percentage errors in focal area in most cases. However, the diverging wave imaging failed to predict the HIFU focus due to its low signal-to-noise ratio.</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>In conclusion, the inherent shear wave artefacts in ARFI for predicting the HIFU focus can be successfully avoided by carefully designing the imaging strategy and its working sequence. This ARFI technique was validated through a series of experiments on a clinically equivalent HIFU system, which demonstrated its capability in assisting surgical planning.</p>\n </section>\n </div>","PeriodicalId":18384,"journal":{"name":"Medical physics","volume":"52 3","pages":"1728-1745"},"PeriodicalIF":3.2000,"publicationDate":"2024-12-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Predicting the high intensity focused ultrasound focus in vivo using acoustic radiation force imaging\",\"authors\":\"Xinwang Shi, Fenglong Zhao, Lian Feng, Yijing Liu, Xiaowei Zhou\",\"doi\":\"10.1002/mp.17564\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Background</h3>\\n \\n <p>One big challenge in the noninvasive high-intensity focused ultrasound (HIFU) surgery is that the location and shape of its focus is unpredictable at the preoperative stage due to the complexity of sound wave propagation. The Acoustic Radiation Force Impulse (ARFI) imaging is a potential solution to this problem, but artifacts resulting from shear wave propagation remain to be solved.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Purpose</h3>\\n \\n <p>In this study, we proposed avoiding those artefacts by applying the ARFI technique at a high imaging frame rate within a very short time before the shear waves start to propagate.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Methods</h3>\\n \\n <p>Using single transmission with a convex imaging probe, two ultrafast imaging modalities (the diverging wave and the wide beam), were developed in the ARFI framework, and their reliabilities were validated on a nylon string phantom by the centroid tracking method borrowed from ultrasound localization microscopy (ULM). The proposed ARFI method was tested on a clinically equivalent HIFU system under different acoustic radiation intensities by in-vitro, ex-vivo and in-vivo experiments. In three experimental scenarios, we delivered short HIFU stimulation pulses at varying acoustic powers to induce tissue motion within the focal region. At each experimental site, both diverging wave and wide-beam imaging techniques were employed for motion estimation. Based on the focus prediction derived from the motion estimation, HIFU ablation treatment was performed. The treated samples were then incised to examine the damaged areas. Additionally, ultrasound B-mode images were acquired before and after the procedure and saved for analysis.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>Quantitative analysis showed that the ARFI with wide beam imaging was able to predict the HIFU focus preoperatively, only with 1 to 3 mm of errors in focal central location, and less than 23% of percentage errors in focal area in most cases. However, the diverging wave imaging failed to predict the HIFU focus due to its low signal-to-noise ratio.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusions</h3>\\n \\n <p>In conclusion, the inherent shear wave artefacts in ARFI for predicting the HIFU focus can be successfully avoided by carefully designing the imaging strategy and its working sequence. This ARFI technique was validated through a series of experiments on a clinically equivalent HIFU system, which demonstrated its capability in assisting surgical planning.</p>\\n </section>\\n </div>\",\"PeriodicalId\":18384,\"journal\":{\"name\":\"Medical physics\",\"volume\":\"52 3\",\"pages\":\"1728-1745\"},\"PeriodicalIF\":3.2000,\"publicationDate\":\"2024-12-11\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Medical physics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1002/mp.17564\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Medical physics","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/mp.17564","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Predicting the high intensity focused ultrasound focus in vivo using acoustic radiation force imaging
Background
One big challenge in the noninvasive high-intensity focused ultrasound (HIFU) surgery is that the location and shape of its focus is unpredictable at the preoperative stage due to the complexity of sound wave propagation. The Acoustic Radiation Force Impulse (ARFI) imaging is a potential solution to this problem, but artifacts resulting from shear wave propagation remain to be solved.
Purpose
In this study, we proposed avoiding those artefacts by applying the ARFI technique at a high imaging frame rate within a very short time before the shear waves start to propagate.
Methods
Using single transmission with a convex imaging probe, two ultrafast imaging modalities (the diverging wave and the wide beam), were developed in the ARFI framework, and their reliabilities were validated on a nylon string phantom by the centroid tracking method borrowed from ultrasound localization microscopy (ULM). The proposed ARFI method was tested on a clinically equivalent HIFU system under different acoustic radiation intensities by in-vitro, ex-vivo and in-vivo experiments. In three experimental scenarios, we delivered short HIFU stimulation pulses at varying acoustic powers to induce tissue motion within the focal region. At each experimental site, both diverging wave and wide-beam imaging techniques were employed for motion estimation. Based on the focus prediction derived from the motion estimation, HIFU ablation treatment was performed. The treated samples were then incised to examine the damaged areas. Additionally, ultrasound B-mode images were acquired before and after the procedure and saved for analysis.
Results
Quantitative analysis showed that the ARFI with wide beam imaging was able to predict the HIFU focus preoperatively, only with 1 to 3 mm of errors in focal central location, and less than 23% of percentage errors in focal area in most cases. However, the diverging wave imaging failed to predict the HIFU focus due to its low signal-to-noise ratio.
Conclusions
In conclusion, the inherent shear wave artefacts in ARFI for predicting the HIFU focus can be successfully avoided by carefully designing the imaging strategy and its working sequence. This ARFI technique was validated through a series of experiments on a clinically equivalent HIFU system, which demonstrated its capability in assisting surgical planning.
期刊介绍:
Medical Physics publishes original, high impact physics, imaging science, and engineering research that advances patient diagnosis and therapy through contributions in 1) Basic science developments with high potential for clinical translation 2) Clinical applications of cutting edge engineering and physics innovations 3) Broadly applicable and innovative clinical physics developments
Medical Physics is a journal of global scope and reach. By publishing in Medical Physics your research will reach an international, multidisciplinary audience including practicing medical physicists as well as physics- and engineering based translational scientists. We work closely with authors of promising articles to improve their quality.