新型Auto + A3阈值函数在维持房室同步中的应用

IF 1.3 4区 医学 Q3 CARDIAC & CARDIOVASCULAR SYSTEMS
Pace-Pacing and Clinical Electrophysiology Pub Date : 2025-01-01 Epub Date: 2024-12-10 DOI:10.1111/pace.15119
Ryo Ohinata, Naoya Inoue, Shuji Morikawa
{"title":"新型Auto + A3阈值函数在维持房室同步中的应用","authors":"Ryo Ohinata, Naoya Inoue, Shuji Morikawa","doi":"10.1111/pace.15119","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>The VDD-leadless pacemaker aims to maintain high atrioventricular (AV) synchrony, but proper sensing setup is crucial.</p><p><strong>Methods and results: </strong>This case was an elderly woman with an AV block who received a dual-chamber pacemaker. Due to ventricular pacing failure, a Micra AV2 was implanted. However, postimplantation, high sinus rate, and frequent premature ventricular contractions were noted. Therefore, the auto + A3 threshold function was used in the atrial-sensing setup. A few days later, the MAM test confirmed good AV synchrony.</p><p><strong>Conclusion: </strong>The auto + A3 threshold function equipped in the Micra AV2 may offer to maintain high AV synchrony.</p>","PeriodicalId":54653,"journal":{"name":"Pace-Pacing and Clinical Electrophysiology","volume":" ","pages":"79-82"},"PeriodicalIF":1.3000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Utility of the Novel Auto + A3 Threshold Function in Maintaining Atrioventricular Synchrony.\",\"authors\":\"Ryo Ohinata, Naoya Inoue, Shuji Morikawa\",\"doi\":\"10.1111/pace.15119\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>The VDD-leadless pacemaker aims to maintain high atrioventricular (AV) synchrony, but proper sensing setup is crucial.</p><p><strong>Methods and results: </strong>This case was an elderly woman with an AV block who received a dual-chamber pacemaker. Due to ventricular pacing failure, a Micra AV2 was implanted. However, postimplantation, high sinus rate, and frequent premature ventricular contractions were noted. Therefore, the auto + A3 threshold function was used in the atrial-sensing setup. A few days later, the MAM test confirmed good AV synchrony.</p><p><strong>Conclusion: </strong>The auto + A3 threshold function equipped in the Micra AV2 may offer to maintain high AV synchrony.</p>\",\"PeriodicalId\":54653,\"journal\":{\"name\":\"Pace-Pacing and Clinical Electrophysiology\",\"volume\":\" \",\"pages\":\"79-82\"},\"PeriodicalIF\":1.3000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Pace-Pacing and Clinical Electrophysiology\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1111/pace.15119\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/12/10 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q3\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pace-Pacing and Clinical Electrophysiology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.15119","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/12/10 0:00:00","PubModel":"Epub","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

摘要

vdd -无导线起搏器旨在保持房室(AV)高度同步,但适当的传感设置至关重要。方法和结果:该病例是一名老年妇女,房室传导阻滞,接受双室起搏器。由于心室起搏衰竭,植入了一个Micra AV2。然而,植入后,窦性窦率高,室性早搏频繁。因此,在心房传感设置中使用auto + A3阈值函数。几天后,MAM测试证实了良好的AV同步。结论:Micra AV2配备auto + A3阈值功能可能有助于维持高AV同步性。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Utility of the Novel Auto + A3 Threshold Function in Maintaining Atrioventricular Synchrony.

Introduction: The VDD-leadless pacemaker aims to maintain high atrioventricular (AV) synchrony, but proper sensing setup is crucial.

Methods and results: This case was an elderly woman with an AV block who received a dual-chamber pacemaker. Due to ventricular pacing failure, a Micra AV2 was implanted. However, postimplantation, high sinus rate, and frequent premature ventricular contractions were noted. Therefore, the auto + A3 threshold function was used in the atrial-sensing setup. A few days later, the MAM test confirmed good AV synchrony.

Conclusion: The auto + A3 threshold function equipped in the Micra AV2 may offer to maintain high AV synchrony.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Pace-Pacing and Clinical Electrophysiology
Pace-Pacing and Clinical Electrophysiology 医学-工程:生物医学
CiteScore
2.70
自引率
5.60%
发文量
209
审稿时长
2-4 weeks
期刊介绍: Pacing and Clinical Electrophysiology (PACE) is the foremost peer-reviewed journal in the field of pacing and implantable cardioversion defibrillation, publishing over 50% of all English language articles in its field, featuring original, review, and didactic papers, and case reports related to daily practice. Articles also include editorials, book reviews, Musings on humane topics relevant to medical practice, electrophysiology (EP) rounds, device rounds, and information concerning the quality of devices used in the practice of the specialty.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信