人工乳房植入物碳离子放射治疗停止功率比的评价。

IF 1.7 Q3 RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING
M Muranaka, M Sakai, T Ogano, Y Hoshino, M Nakao, K Yusa, T Suto, T Ohno
{"title":"人工乳房植入物碳离子放射治疗停止功率比的评价。","authors":"M Muranaka, M Sakai, T Ogano, Y Hoshino, M Nakao, K Yusa, T Suto, T Ohno","doi":"10.1007/s12194-024-00860-4","DOIUrl":null,"url":null,"abstract":"<p><p>The number of patients requiring breast reconstruction with artificial implants has been increasing, and so is the use of carbon-ion radiotherapy (CIRT). Consequently, a growing number of patients with artificial breast implants are expected to undergo CIRT. Because artificial breasts are composed of a silicone polymer gel with a silicon-oxygen backbone, which differs significantly from human tissues, the stopping power ratio for carbon beams cannot be accurately converted from CT values using standard CT-to-stopping power ratio tables (CT-SP tables). Incorrect stopping power ratios can lead to significant problems in CIRT, including erroneous calculations of carbon beam range. To address this, we measured the CT values and stopping power ratios of three commercial artificial breasts using a 380 MeV/u carbon beam. Our results revealed significant deviations from the CT-SP table values. For instance, calculations for treating lung cancer with incorrect stopping power ratios resulted in errors of approximately 5 mm in range calculations, adversely affecting dose distribution to the target. Although further studies with various products are needed, it is crucial to conduct thorough patient consultations and develop treatment plans using accurate stopping power ratios.</p>","PeriodicalId":46252,"journal":{"name":"Radiological Physics and Technology","volume":" ","pages":""},"PeriodicalIF":1.7000,"publicationDate":"2024-12-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Evaluation of stopping power ratio of artificial breast implants for carbon-ion radiotherapy.\",\"authors\":\"M Muranaka, M Sakai, T Ogano, Y Hoshino, M Nakao, K Yusa, T Suto, T Ohno\",\"doi\":\"10.1007/s12194-024-00860-4\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The number of patients requiring breast reconstruction with artificial implants has been increasing, and so is the use of carbon-ion radiotherapy (CIRT). Consequently, a growing number of patients with artificial breast implants are expected to undergo CIRT. Because artificial breasts are composed of a silicone polymer gel with a silicon-oxygen backbone, which differs significantly from human tissues, the stopping power ratio for carbon beams cannot be accurately converted from CT values using standard CT-to-stopping power ratio tables (CT-SP tables). Incorrect stopping power ratios can lead to significant problems in CIRT, including erroneous calculations of carbon beam range. To address this, we measured the CT values and stopping power ratios of three commercial artificial breasts using a 380 MeV/u carbon beam. Our results revealed significant deviations from the CT-SP table values. For instance, calculations for treating lung cancer with incorrect stopping power ratios resulted in errors of approximately 5 mm in range calculations, adversely affecting dose distribution to the target. Although further studies with various products are needed, it is crucial to conduct thorough patient consultations and develop treatment plans using accurate stopping power ratios.</p>\",\"PeriodicalId\":46252,\"journal\":{\"name\":\"Radiological Physics and Technology\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-12-04\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Radiological Physics and Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/s12194-024-00860-4\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Radiological Physics and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12194-024-00860-4","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
引用次数: 0

摘要

需要人工乳房重建的患者数量一直在增加,碳离子放射治疗(CIRT)的使用也在增加。因此,越来越多的人工乳房植入患者预计将接受CIRT。由于人造乳房是由硅氧骨架的有机硅聚合物凝胶组成,这与人体组织有很大的不同,因此使用标准的CT-停止功率比表(CT- sp表)无法准确地从CT值转换碳束的停止功率比。不正确的停止功率比会导致CIRT中的重大问题,包括碳束范围的错误计算。为了解决这个问题,我们使用380 MeV/u的碳束测量了三个商用人工乳房的CT值和停止功率比。我们的结果显示了与CT-SP表值的显著偏差。例如,使用不正确的停止功率比进行治疗肺癌的计算,导致范围计算误差约为5mm,对靶的剂量分布产生不利影响。虽然需要对各种产品进行进一步的研究,但至关重要的是进行彻底的患者咨询并制定使用准确停止功率比的治疗计划。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Evaluation of stopping power ratio of artificial breast implants for carbon-ion radiotherapy.

The number of patients requiring breast reconstruction with artificial implants has been increasing, and so is the use of carbon-ion radiotherapy (CIRT). Consequently, a growing number of patients with artificial breast implants are expected to undergo CIRT. Because artificial breasts are composed of a silicone polymer gel with a silicon-oxygen backbone, which differs significantly from human tissues, the stopping power ratio for carbon beams cannot be accurately converted from CT values using standard CT-to-stopping power ratio tables (CT-SP tables). Incorrect stopping power ratios can lead to significant problems in CIRT, including erroneous calculations of carbon beam range. To address this, we measured the CT values and stopping power ratios of three commercial artificial breasts using a 380 MeV/u carbon beam. Our results revealed significant deviations from the CT-SP table values. For instance, calculations for treating lung cancer with incorrect stopping power ratios resulted in errors of approximately 5 mm in range calculations, adversely affecting dose distribution to the target. Although further studies with various products are needed, it is crucial to conduct thorough patient consultations and develop treatment plans using accurate stopping power ratios.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Radiological Physics and Technology
Radiological Physics and Technology RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING-
CiteScore
3.00
自引率
12.50%
发文量
40
期刊介绍: The purpose of the journal Radiological Physics and Technology is to provide a forum for sharing new knowledge related to research and development in radiological science and technology, including medical physics and radiological technology in diagnostic radiology, nuclear medicine, and radiation therapy among many other radiological disciplines, as well as to contribute to progress and improvement in medical practice and patient health care.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信