Mathilde Levardon, Damien Autret, Thomas Le Dorze, Camille Guillerminet, Stéphane Dufreneix
{"title":"临床实践中的纯脑磁共振工作流程:质量保证和患者定位生成器之间的比较。","authors":"Mathilde Levardon, Damien Autret, Thomas Le Dorze, Camille Guillerminet, Stéphane Dufreneix","doi":"10.1002/acm2.14583","DOIUrl":null,"url":null,"abstract":"<p><strong>Background and purpose: </strong>Routine quality control procedures are still required for sCT based on artificial intelligence (AI) to verify the performance of the generators. The aim of this study was to evaluate three generators based on AI or bulk density (BD) assignment for the patient-specific quality assurance (PSQA) of another AI-based generator in clinical routine. A patient positioning study based on 2D/2D kV-image comparing the performances of four sCT generators was also performed.</p><p><strong>Materials and methods: </strong>On the four generators available commercially at our institution, one was chosen as the clinical one, and the three others were used for PSQA. Several dose metrics were calculated like the mean error, dose-volume histogram metrics, and 1%/1 mm gamma analysis. A comparison against CT was considered as a reference. Translations and rotations found during patient positioning based on sCT were compared to those based on CT.</p><p><strong>Results: </strong>Some of the metrics calculated against CT revealed patients outside the tolerances chosen (1% for point metrics; 90% for gamma pass rate). None of the generators was able to identify these outliers for all metrics studied. Performing a PSQA with other sCT generators introduced several false positives and false negatives. None of the generators was able to clearly identify, for all metrics studied, a true sCT failure caused by a metal implant. The smallest positioning deviations were found for the BD assignment sCT, the largest for the only AI generator not based on a T1 Dixon MR sequence.</p><p><strong>Conclusions: </strong>PSQA of a sCT generator with another sCT generator should be performed with great care. Patient positioning is an important aspect to consider when evaluating a sCT generator. The results of this study should help medical physicists willing to set up a MR-only workflow for the brain based on a 2D/2D kV-image patient positioning.</p>","PeriodicalId":14989,"journal":{"name":"Journal of Applied Clinical Medical Physics","volume":" ","pages":"e14583"},"PeriodicalIF":2.0000,"publicationDate":"2024-11-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Brain MR-only workflow in clinical practice: A comparison among generators for quality assurance and patient positioning.\",\"authors\":\"Mathilde Levardon, Damien Autret, Thomas Le Dorze, Camille Guillerminet, Stéphane Dufreneix\",\"doi\":\"10.1002/acm2.14583\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background and purpose: </strong>Routine quality control procedures are still required for sCT based on artificial intelligence (AI) to verify the performance of the generators. The aim of this study was to evaluate three generators based on AI or bulk density (BD) assignment for the patient-specific quality assurance (PSQA) of another AI-based generator in clinical routine. A patient positioning study based on 2D/2D kV-image comparing the performances of four sCT generators was also performed.</p><p><strong>Materials and methods: </strong>On the four generators available commercially at our institution, one was chosen as the clinical one, and the three others were used for PSQA. Several dose metrics were calculated like the mean error, dose-volume histogram metrics, and 1%/1 mm gamma analysis. A comparison against CT was considered as a reference. Translations and rotations found during patient positioning based on sCT were compared to those based on CT.</p><p><strong>Results: </strong>Some of the metrics calculated against CT revealed patients outside the tolerances chosen (1% for point metrics; 90% for gamma pass rate). None of the generators was able to identify these outliers for all metrics studied. Performing a PSQA with other sCT generators introduced several false positives and false negatives. None of the generators was able to clearly identify, for all metrics studied, a true sCT failure caused by a metal implant. The smallest positioning deviations were found for the BD assignment sCT, the largest for the only AI generator not based on a T1 Dixon MR sequence.</p><p><strong>Conclusions: </strong>PSQA of a sCT generator with another sCT generator should be performed with great care. Patient positioning is an important aspect to consider when evaluating a sCT generator. The results of this study should help medical physicists willing to set up a MR-only workflow for the brain based on a 2D/2D kV-image patient positioning.</p>\",\"PeriodicalId\":14989,\"journal\":{\"name\":\"Journal of Applied Clinical Medical Physics\",\"volume\":\" \",\"pages\":\"e14583\"},\"PeriodicalIF\":2.0000,\"publicationDate\":\"2024-11-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Applied Clinical Medical Physics\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1002/acm2.14583\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Applied Clinical Medical Physics","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1002/acm2.14583","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Brain MR-only workflow in clinical practice: A comparison among generators for quality assurance and patient positioning.
Background and purpose: Routine quality control procedures are still required for sCT based on artificial intelligence (AI) to verify the performance of the generators. The aim of this study was to evaluate three generators based on AI or bulk density (BD) assignment for the patient-specific quality assurance (PSQA) of another AI-based generator in clinical routine. A patient positioning study based on 2D/2D kV-image comparing the performances of four sCT generators was also performed.
Materials and methods: On the four generators available commercially at our institution, one was chosen as the clinical one, and the three others were used for PSQA. Several dose metrics were calculated like the mean error, dose-volume histogram metrics, and 1%/1 mm gamma analysis. A comparison against CT was considered as a reference. Translations and rotations found during patient positioning based on sCT were compared to those based on CT.
Results: Some of the metrics calculated against CT revealed patients outside the tolerances chosen (1% for point metrics; 90% for gamma pass rate). None of the generators was able to identify these outliers for all metrics studied. Performing a PSQA with other sCT generators introduced several false positives and false negatives. None of the generators was able to clearly identify, for all metrics studied, a true sCT failure caused by a metal implant. The smallest positioning deviations were found for the BD assignment sCT, the largest for the only AI generator not based on a T1 Dixon MR sequence.
Conclusions: PSQA of a sCT generator with another sCT generator should be performed with great care. Patient positioning is an important aspect to consider when evaluating a sCT generator. The results of this study should help medical physicists willing to set up a MR-only workflow for the brain based on a 2D/2D kV-image patient positioning.
期刊介绍:
Journal of Applied Clinical Medical Physics is an international Open Access publication dedicated to clinical medical physics. JACMP welcomes original contributions dealing with all aspects of medical physics from scientists working in the clinical medical physics around the world. JACMP accepts only online submission.
JACMP will publish:
-Original Contributions: Peer-reviewed, investigations that represent new and significant contributions to the field. Recommended word count: up to 7500.
-Review Articles: Reviews of major areas or sub-areas in the field of clinical medical physics. These articles may be of any length and are peer reviewed.
-Technical Notes: These should be no longer than 3000 words, including key references.
-Letters to the Editor: Comments on papers published in JACMP or on any other matters of interest to clinical medical physics. These should not be more than 1250 (including the literature) and their publication is only based on the decision of the editor, who occasionally asks experts on the merit of the contents.
-Book Reviews: The editorial office solicits Book Reviews.
-Announcements of Forthcoming Meetings: The Editor may provide notice of forthcoming meetings, course offerings, and other events relevant to clinical medical physics.
-Parallel Opposed Editorial: We welcome topics relevant to clinical practice and medical physics profession. The contents can be controversial debate or opposed aspects of an issue. One author argues for the position and the other against. Each side of the debate contains an opening statement up to 800 words, followed by a rebuttal up to 500 words. Readers interested in participating in this series should contact the moderator with a proposed title and a short description of the topic