Karanjeet Chauhan, Alistair Royse, Ian Goh, Ian Crozier, Gareth Wynn
{"title":"新型胸骨下血管外除颤器的最古老人体成功提取经验。","authors":"Karanjeet Chauhan, Alistair Royse, Ian Goh, Ian Crozier, Gareth Wynn","doi":"10.1111/jce.16511","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Substernal extravascular defibrillators (EV ICDs) have been shown to be effective and safe for patients at risk of sudden cardiac death, however, there is little evidence around the safety of extracting chronic devices.</p><p><strong>Methods and results: </strong>We present a 50-year-old patient in whom a Medtronic EV ICD system was successfully removed without specialist extraction tools, 186 weeks after implantation, by an operator experienced in transvenous lead extraction but without formal training in EVICD implantation.</p><p><strong>Conclusion: </strong>The successful extraction of an EV ICD system is possible without specialised tools at least 3.6 years post-implant.</p>","PeriodicalId":15178,"journal":{"name":"Journal of Cardiovascular Electrophysiology","volume":" ","pages":"295-297"},"PeriodicalIF":2.3000,"publicationDate":"2025-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Oldest-in-Human Successful Extraction Experience of a Novel Substernal Extravascular Defibrillator.\",\"authors\":\"Karanjeet Chauhan, Alistair Royse, Ian Goh, Ian Crozier, Gareth Wynn\",\"doi\":\"10.1111/jce.16511\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>Substernal extravascular defibrillators (EV ICDs) have been shown to be effective and safe for patients at risk of sudden cardiac death, however, there is little evidence around the safety of extracting chronic devices.</p><p><strong>Methods and results: </strong>We present a 50-year-old patient in whom a Medtronic EV ICD system was successfully removed without specialist extraction tools, 186 weeks after implantation, by an operator experienced in transvenous lead extraction but without formal training in EVICD implantation.</p><p><strong>Conclusion: </strong>The successful extraction of an EV ICD system is possible without specialised tools at least 3.6 years post-implant.</p>\",\"PeriodicalId\":15178,\"journal\":{\"name\":\"Journal of Cardiovascular Electrophysiology\",\"volume\":\" \",\"pages\":\"295-297\"},\"PeriodicalIF\":2.3000,\"publicationDate\":\"2025-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Cardiovascular Electrophysiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1111/jce.16511\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/11/22 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/jce.16511","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/11/22 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
摘要
简介:对于有心脏性猝死风险的患者而言,经证实胸骨下血管外除颤器(EV ICD)是有效和安全的,然而,关于提取慢性装置的安全性却鲜有证据:我们介绍了一位 50 岁的患者,在植入美敦力 EV ICD 186 周后,由一位在经静脉导联取出方面经验丰富但未接受过 EVICD 植入正规培训的操作者在没有专业取出工具的情况下成功取出了该系统:结论:植入 EV ICD 系统至少 3.6 年后,无需专业工具即可成功取出。
Oldest-in-Human Successful Extraction Experience of a Novel Substernal Extravascular Defibrillator.
Introduction: Substernal extravascular defibrillators (EV ICDs) have been shown to be effective and safe for patients at risk of sudden cardiac death, however, there is little evidence around the safety of extracting chronic devices.
Methods and results: We present a 50-year-old patient in whom a Medtronic EV ICD system was successfully removed without specialist extraction tools, 186 weeks after implantation, by an operator experienced in transvenous lead extraction but without formal training in EVICD implantation.
Conclusion: The successful extraction of an EV ICD system is possible without specialised tools at least 3.6 years post-implant.
期刊介绍:
Journal of Cardiovascular Electrophysiology (JCE) keeps its readership well informed of the latest developments in the study and management of arrhythmic disorders. Edited by Bradley P. Knight, M.D., and a distinguished international editorial board, JCE is the leading journal devoted to the study of the electrophysiology of the heart.