模拟傅立叶变换四极杆离子捕获器 (FT-QIT) 中的空间电荷效应。

IF 3.1 2区 化学 Q2 BIOCHEMICAL RESEARCH METHODS
Walter Wißdorf, Marco Thinius, Thorsten Benter
{"title":"模拟傅立叶变换四极杆离子捕获器 (FT-QIT) 中的空间电荷效应。","authors":"Walter Wißdorf, Marco Thinius, Thorsten Benter","doi":"10.1021/jasms.4c00296","DOIUrl":null,"url":null,"abstract":"<p><p>We present ion dynamics simulations regarding the effect of space charge on the performance of Fourier transform quadrupole ion traps (FT-QITs) with special attention to signal stability, mass resolving power, and sensitivity. Ion trajectory simulations within an idealized QIT geometry are performed by applying a dedicated application (QITSim) using an in-house developed open simulation framework (IDSimF). Image current detection transients are generated by the application and are subsequently transformed into frequency spectra of ion secular motion. Such frequency spectra are the basis for the calculation of mass spectra in FT-QIT instruments. The simulation results are used to assess the extent of space charge induced effects regarding the analytical performance of FT-QIT systems. The simulation results for two Cl<sup>+</sup> and seven Xe<sup>+</sup> isotope ions exhibit diverse space charge induced phenomena. Most prominently, complete isotope signal fusion, quantitative individual signal suppression, and severe signal distortions are observed even at low absolute numbers of trapped ions. Furthermore, significant shifts of the secular oscillation frequencies occur, even when the signal shape appears to be mostly unaffected and when the frequency separation for trapped ions with different masses is large. This significantly limits the applicability of FT-QIT systems for high resolution mass spectrometry. An increase of the RF amplitude of the trapping field as well as increasing the extent of ion excitation partly mitigate these adverse space charge effects; nevertheless, the usable operational range of the simulated FT-QIT system for analytical applications remains rather narrow.</p>","PeriodicalId":672,"journal":{"name":"Journal of the American Society for Mass Spectrometry","volume":" ","pages":"2969-2983"},"PeriodicalIF":3.1000,"publicationDate":"2024-12-04","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Simulation of Space Charge Effects in Fourier Transform Quadrupole Ion Traps (FT-QITs).\",\"authors\":\"Walter Wißdorf, Marco Thinius, Thorsten Benter\",\"doi\":\"10.1021/jasms.4c00296\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>We present ion dynamics simulations regarding the effect of space charge on the performance of Fourier transform quadrupole ion traps (FT-QITs) with special attention to signal stability, mass resolving power, and sensitivity. Ion trajectory simulations within an idealized QIT geometry are performed by applying a dedicated application (QITSim) using an in-house developed open simulation framework (IDSimF). Image current detection transients are generated by the application and are subsequently transformed into frequency spectra of ion secular motion. Such frequency spectra are the basis for the calculation of mass spectra in FT-QIT instruments. The simulation results are used to assess the extent of space charge induced effects regarding the analytical performance of FT-QIT systems. The simulation results for two Cl<sup>+</sup> and seven Xe<sup>+</sup> isotope ions exhibit diverse space charge induced phenomena. Most prominently, complete isotope signal fusion, quantitative individual signal suppression, and severe signal distortions are observed even at low absolute numbers of trapped ions. Furthermore, significant shifts of the secular oscillation frequencies occur, even when the signal shape appears to be mostly unaffected and when the frequency separation for trapped ions with different masses is large. This significantly limits the applicability of FT-QIT systems for high resolution mass spectrometry. An increase of the RF amplitude of the trapping field as well as increasing the extent of ion excitation partly mitigate these adverse space charge effects; nevertheless, the usable operational range of the simulated FT-QIT system for analytical applications remains rather narrow.</p>\",\"PeriodicalId\":672,\"journal\":{\"name\":\"Journal of the American Society for Mass Spectrometry\",\"volume\":\" \",\"pages\":\"2969-2983\"},\"PeriodicalIF\":3.1000,\"publicationDate\":\"2024-12-04\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of the American Society for Mass Spectrometry\",\"FirstCategoryId\":\"92\",\"ListUrlMain\":\"https://doi.org/10.1021/jasms.4c00296\",\"RegionNum\":2,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/11/6 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"BIOCHEMICAL RESEARCH METHODS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of the American Society for Mass Spectrometry","FirstCategoryId":"92","ListUrlMain":"https://doi.org/10.1021/jasms.4c00296","RegionNum":2,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/11/6 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0

摘要

我们介绍了有关空间电荷对傅立叶变换四极杆离子阱(FT-QIT)性能影响的离子动力学模拟,特别关注信号稳定性、质量分辨能力和灵敏度。通过使用内部开发的开放式仿真框架(IDSimF),应用专用应用程序(QITSim)在理想化的 QIT 几何结构内进行离子轨迹仿真。图像电流检测瞬态由应用程序生成,随后转化为离子瞬时运动的频率谱。这种频率谱是计算 FT-QIT 仪器质谱的基础。模拟结果用于评估空间电荷对 FT-QIT 系统分析性能的影响程度。两个 Cl+ 和七个 Xe+ 同位素离子的模拟结果显示出多种空间电荷诱导现象。最突出的是,即使被捕获离子的绝对数量较少,也能观察到完全的同位素信号融合、定量的单个信号抑制和严重的信号失真。此外,即使信号形状看起来基本不受影响,而且不同质量的被捕获离子的频率间隔很大时,也会出现世俗振荡频率的明显偏移。这极大地限制了 FT-QIT 系统在高分辨率质谱分析中的应用。提高俘获场的射频振幅以及增加离子激发范围可以部分缓解这些不利的空间电荷效应;不过,模拟 FT-QIT 系统在分析应用中的可用操作范围仍然相当狭窄。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Simulation of Space Charge Effects in Fourier Transform Quadrupole Ion Traps (FT-QITs).

We present ion dynamics simulations regarding the effect of space charge on the performance of Fourier transform quadrupole ion traps (FT-QITs) with special attention to signal stability, mass resolving power, and sensitivity. Ion trajectory simulations within an idealized QIT geometry are performed by applying a dedicated application (QITSim) using an in-house developed open simulation framework (IDSimF). Image current detection transients are generated by the application and are subsequently transformed into frequency spectra of ion secular motion. Such frequency spectra are the basis for the calculation of mass spectra in FT-QIT instruments. The simulation results are used to assess the extent of space charge induced effects regarding the analytical performance of FT-QIT systems. The simulation results for two Cl+ and seven Xe+ isotope ions exhibit diverse space charge induced phenomena. Most prominently, complete isotope signal fusion, quantitative individual signal suppression, and severe signal distortions are observed even at low absolute numbers of trapped ions. Furthermore, significant shifts of the secular oscillation frequencies occur, even when the signal shape appears to be mostly unaffected and when the frequency separation for trapped ions with different masses is large. This significantly limits the applicability of FT-QIT systems for high resolution mass spectrometry. An increase of the RF amplitude of the trapping field as well as increasing the extent of ion excitation partly mitigate these adverse space charge effects; nevertheless, the usable operational range of the simulated FT-QIT system for analytical applications remains rather narrow.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
5.50
自引率
9.40%
发文量
257
审稿时长
1 months
期刊介绍: The Journal of the American Society for Mass Spectrometry presents research papers covering all aspects of mass spectrometry, incorporating coverage of fields of scientific inquiry in which mass spectrometry can play a role. Comprehensive in scope, the journal publishes papers on both fundamentals and applications of mass spectrometry. Fundamental subjects include instrumentation principles, design, and demonstration, structures and chemical properties of gas-phase ions, studies of thermodynamic properties, ion spectroscopy, chemical kinetics, mechanisms of ionization, theories of ion fragmentation, cluster ions, and potential energy surfaces. In addition to full papers, the journal offers Communications, Application Notes, and Accounts and Perspectives
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信