ICD 动态灵敏度导致心室异位搏动后的功能性心室欠敏感。

IF 1.7 4区 医学 Q3 CARDIAC & CARDIOVASCULAR SYSTEMS
Antonino Micari, Paolo Bellocchi, Pasquale Crea
{"title":"ICD 动态灵敏度导致心室异位搏动后的功能性心室欠敏感。","authors":"Antonino Micari, Paolo Bellocchi, Pasquale Crea","doi":"10.1111/pace.15099","DOIUrl":null,"url":null,"abstract":"<p><p>An 80-year-old patient, previously implanted for idiopathic sustained ventricular tachycardia with a Rivacor VR-T DX implantable cardioverter defibrillator (ICD Biotronik, Berlin, Germany), came to our pacemaker clinic for palpitations. Electrocardiogram (ECG) showed sinus bradycardia with interpolated ventricular ectopic beats. A ventricular-paced event occurred with a coupling interval shorter than the expected one suggesting ventricular undersensing. The phenomenon can be explained by the default ICD sensitivity setting proposed by the manufacturer.</p>","PeriodicalId":54653,"journal":{"name":"Pace-Pacing and Clinical Electrophysiology","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-10-24","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Functional Ventricular Undersensing After Ventricular Ectopic Beats Due to ICD Dynamic Sensitivity.\",\"authors\":\"Antonino Micari, Paolo Bellocchi, Pasquale Crea\",\"doi\":\"10.1111/pace.15099\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>An 80-year-old patient, previously implanted for idiopathic sustained ventricular tachycardia with a Rivacor VR-T DX implantable cardioverter defibrillator (ICD Biotronik, Berlin, Germany), came to our pacemaker clinic for palpitations. Electrocardiogram (ECG) showed sinus bradycardia with interpolated ventricular ectopic beats. A ventricular-paced event occurred with a coupling interval shorter than the expected one suggesting ventricular undersensing. The phenomenon can be explained by the default ICD sensitivity setting proposed by the manufacturer.</p>\",\"PeriodicalId\":54653,\"journal\":{\"name\":\"Pace-Pacing and Clinical Electrophysiology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-10-24\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Pace-Pacing and Clinical Electrophysiology\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1111/pace.15099\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pace-Pacing and Clinical Electrophysiology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.15099","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0

摘要

一位 80 岁的患者曾因特发性持续室性心动过速而植入 Rivacor VR-T DX 植入式心脏转复除颤器(ICD Biotronik,德国柏林),后因心悸来到我们的起搏器诊所就诊。心电图(ECG)显示窦性心动过缓,伴有室性异位搏动。心室搏动事件发生时的耦合间期短于预期,表明心室感应不足。这一现象可以用制造商建议的 ICD 默认灵敏度设置来解释。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Functional Ventricular Undersensing After Ventricular Ectopic Beats Due to ICD Dynamic Sensitivity.

An 80-year-old patient, previously implanted for idiopathic sustained ventricular tachycardia with a Rivacor VR-T DX implantable cardioverter defibrillator (ICD Biotronik, Berlin, Germany), came to our pacemaker clinic for palpitations. Electrocardiogram (ECG) showed sinus bradycardia with interpolated ventricular ectopic beats. A ventricular-paced event occurred with a coupling interval shorter than the expected one suggesting ventricular undersensing. The phenomenon can be explained by the default ICD sensitivity setting proposed by the manufacturer.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Pace-Pacing and Clinical Electrophysiology
Pace-Pacing and Clinical Electrophysiology 医学-工程:生物医学
CiteScore
2.70
自引率
5.60%
发文量
209
审稿时长
2-4 weeks
期刊介绍: Pacing and Clinical Electrophysiology (PACE) is the foremost peer-reviewed journal in the field of pacing and implantable cardioversion defibrillation, publishing over 50% of all English language articles in its field, featuring original, review, and didactic papers, and case reports related to daily practice. Articles also include editorials, book reviews, Musings on humane topics relevant to medical practice, electrophysiology (EP) rounds, device rounds, and information concerning the quality of devices used in the practice of the specialty.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信