{"title":"在儿科环境中,两种不同的 PCA 泵在低流速时因极点位置而导致输液不准确的可能性--一项实验研究。","authors":"Judith E Riemer, Nathalie C Greber, Melanie Frühauf, Markus Weiss, Tobias Piegeler","doi":"10.1007/s10877-024-01233-3","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>Vertical displacement of infusion pumps used in patient-controlled analgesia can cause irregularities in drug delivery and is especially crucial at low flow rates, which are commonly used in pediatrics. There is only scarce data available regarding the extent of these inaccuracies. The current in vitro study therefore aimed at a comparison of the performance of two commonly used PCA pumps at different pole positions due to vertical displacement.</p><p><strong>Methods: </strong>The Syramed<sup>®</sup> µSP6000 Chroma syringe infusion pump featuring a stepper motor drive was compared to the CADD<sup>®</sup>-Solis pump utilizing a linear peristaltic pump system at two different flow rates (0.3 ml/h and 1 ml/h) and three different levels of height (0, + 50 and - 50 cm). Flow patterns and delivered volumes were measured after every change in position and infusion boluses, retrograde aspiration volumes and zero-drug delivery time were calculated.</p><p><strong>Results: </strong>The Syramed<sup>®</sup> pump was more susceptible to vertical displacement than the CADD<sup>®</sup>-Solis pump and showed overall greater inaccuracies in the delivered volumes as well as higher infusion boluses, retrograde aspiration volumes and zero-drug delivery time at both flow rates.</p><p><strong>Conclusion: </strong>The observed differences between the pumps might be explained by the higher compliance of this syringe pump and the diverse working mechanisms. Overall, the CADD<sup>®</sup>-Solis pump might be considered a preferable option for patient-controlled analgesia in children. It is nonetheless essential for medical staff to be aware of the effects of vertical displacement of PCA pumps and to minimize these displacements as much as possible.</p>","PeriodicalId":15513,"journal":{"name":"Journal of Clinical Monitoring and Computing","volume":" ","pages":""},"PeriodicalIF":2.0000,"publicationDate":"2024-10-12","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Susceptibility of two different PCA pumps to inaccurate delivery associated with pole position at low flow-rates in a pediatric setting - an experimental study.\",\"authors\":\"Judith E Riemer, Nathalie C Greber, Melanie Frühauf, Markus Weiss, Tobias Piegeler\",\"doi\":\"10.1007/s10877-024-01233-3\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>Vertical displacement of infusion pumps used in patient-controlled analgesia can cause irregularities in drug delivery and is especially crucial at low flow rates, which are commonly used in pediatrics. There is only scarce data available regarding the extent of these inaccuracies. The current in vitro study therefore aimed at a comparison of the performance of two commonly used PCA pumps at different pole positions due to vertical displacement.</p><p><strong>Methods: </strong>The Syramed<sup>®</sup> µSP6000 Chroma syringe infusion pump featuring a stepper motor drive was compared to the CADD<sup>®</sup>-Solis pump utilizing a linear peristaltic pump system at two different flow rates (0.3 ml/h and 1 ml/h) and three different levels of height (0, + 50 and - 50 cm). Flow patterns and delivered volumes were measured after every change in position and infusion boluses, retrograde aspiration volumes and zero-drug delivery time were calculated.</p><p><strong>Results: </strong>The Syramed<sup>®</sup> pump was more susceptible to vertical displacement than the CADD<sup>®</sup>-Solis pump and showed overall greater inaccuracies in the delivered volumes as well as higher infusion boluses, retrograde aspiration volumes and zero-drug delivery time at both flow rates.</p><p><strong>Conclusion: </strong>The observed differences between the pumps might be explained by the higher compliance of this syringe pump and the diverse working mechanisms. Overall, the CADD<sup>®</sup>-Solis pump might be considered a preferable option for patient-controlled analgesia in children. It is nonetheless essential for medical staff to be aware of the effects of vertical displacement of PCA pumps and to minimize these displacements as much as possible.</p>\",\"PeriodicalId\":15513,\"journal\":{\"name\":\"Journal of Clinical Monitoring and Computing\",\"volume\":\" \",\"pages\":\"\"},\"PeriodicalIF\":2.0000,\"publicationDate\":\"2024-10-12\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Clinical Monitoring and Computing\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1007/s10877-024-01233-3\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ANESTHESIOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Clinical Monitoring and Computing","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1007/s10877-024-01233-3","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ANESTHESIOLOGY","Score":null,"Total":0}
Susceptibility of two different PCA pumps to inaccurate delivery associated with pole position at low flow-rates in a pediatric setting - an experimental study.
Purpose: Vertical displacement of infusion pumps used in patient-controlled analgesia can cause irregularities in drug delivery and is especially crucial at low flow rates, which are commonly used in pediatrics. There is only scarce data available regarding the extent of these inaccuracies. The current in vitro study therefore aimed at a comparison of the performance of two commonly used PCA pumps at different pole positions due to vertical displacement.
Methods: The Syramed® µSP6000 Chroma syringe infusion pump featuring a stepper motor drive was compared to the CADD®-Solis pump utilizing a linear peristaltic pump system at two different flow rates (0.3 ml/h and 1 ml/h) and three different levels of height (0, + 50 and - 50 cm). Flow patterns and delivered volumes were measured after every change in position and infusion boluses, retrograde aspiration volumes and zero-drug delivery time were calculated.
Results: The Syramed® pump was more susceptible to vertical displacement than the CADD®-Solis pump and showed overall greater inaccuracies in the delivered volumes as well as higher infusion boluses, retrograde aspiration volumes and zero-drug delivery time at both flow rates.
Conclusion: The observed differences between the pumps might be explained by the higher compliance of this syringe pump and the diverse working mechanisms. Overall, the CADD®-Solis pump might be considered a preferable option for patient-controlled analgesia in children. It is nonetheless essential for medical staff to be aware of the effects of vertical displacement of PCA pumps and to minimize these displacements as much as possible.
期刊介绍:
The Journal of Clinical Monitoring and Computing is a clinical journal publishing papers related to technology in the fields of anaesthesia, intensive care medicine, emergency medicine, and peri-operative medicine.
The journal has links with numerous specialist societies, including editorial board representatives from the European Society for Computing and Technology in Anaesthesia and Intensive Care (ESCTAIC), the Society for Technology in Anesthesia (STA), the Society for Complex Acute Illness (SCAI) and the NAVAt (NAVigating towards your Anaestheisa Targets) group.
The journal publishes original papers, narrative and systematic reviews, technological notes, letters to the editor, editorial or commentary papers, and policy statements or guidelines from national or international societies. The journal encourages debate on published papers and technology, including letters commenting on previous publications or technological concerns. The journal occasionally publishes special issues with technological or clinical themes, or reports and abstracts from scientificmeetings. Special issues proposals should be sent to the Editor-in-Chief. Specific details of types of papers, and the clinical and technological content of papers considered within scope can be found in instructions for authors.