{"title":"皮下植入式心律转复除颤器患者的合成 V7 QRS 振幅和过敏发作。","authors":"Takashi Okajima, Shinji Ishikawa, Satoshi Yanagisawa, Takayuki Okamoto, Yusuke Uemura, Kenji Takemoto, Yasuya Inden, Toyoaki Murohara, Masato Watarai","doi":"10.1111/pace.15086","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Patients with subcutaneous implantable cardioverter-defibrillators (S-ICDs) experience an oversensing episode (OS) more frequently than those with transvenous ICDs. However, no established electrocardiography (ECG) parameters can accurately detect an OS. This study aimed to evaluate the incidence of an OS in real-world clinical practice and the association of synthesized 18-lead ECG (syn18-ECG) parameters with an OS.</p><p><strong>Methods: </strong>We retrospectively included 21 consecutive patients who underwent S-ICD implantation and collected syn18-ECG parameters. We placed the generator in a deep posterior position and defined an OS as an inappropriate charging episode caused by cardiac or noncardiac signals. A SMART pass filter and two tachyarrhythmia zones were programed.</p><p><strong>Results: </strong>The most frequent underlying heart disease was Brugada/J wave syndrome (n = 7). During a median follow-up period of 1188 days, an OS was observed in six patients (28.6%). The QRS amplitude in synthesized V7 lead (synV7) was significantly lower in the OS group than in the non-OS group (0.59 ± 0.17 vs. 0.91 ± 0.35 mV, p = 0.019). The optimal cutoff value of synV7 QRS amplitude was 0.61 mV, with a sensitivity of 80.0% and a specificity of 83.7% for predicting an OS. Univariate logistic analysis showed that a synV7 QRS amplitude of <0.61 mV was only associated with an OS (odd ratio, 20.0; 95% confidence interval, 1.66-241.72; p = 0.018).</p><p><strong>Conclusions: </strong>In patients with S-ICDs, an OS was not a rare complication during long-term follow-up. A low synV7 QRS amplitude was associated with a high OS incidence.</p>","PeriodicalId":54653,"journal":{"name":"Pace-Pacing and Clinical Electrophysiology","volume":null,"pages":null},"PeriodicalIF":1.7000,"publicationDate":"2024-10-05","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Synthesized V7 QRS Amplitude and Oversensing Episodes in Patients With Subcutaneous Implantable Cardioverter-Defibrillators.\",\"authors\":\"Takashi Okajima, Shinji Ishikawa, Satoshi Yanagisawa, Takayuki Okamoto, Yusuke Uemura, Kenji Takemoto, Yasuya Inden, Toyoaki Murohara, Masato Watarai\",\"doi\":\"10.1111/pace.15086\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Patients with subcutaneous implantable cardioverter-defibrillators (S-ICDs) experience an oversensing episode (OS) more frequently than those with transvenous ICDs. However, no established electrocardiography (ECG) parameters can accurately detect an OS. This study aimed to evaluate the incidence of an OS in real-world clinical practice and the association of synthesized 18-lead ECG (syn18-ECG) parameters with an OS.</p><p><strong>Methods: </strong>We retrospectively included 21 consecutive patients who underwent S-ICD implantation and collected syn18-ECG parameters. We placed the generator in a deep posterior position and defined an OS as an inappropriate charging episode caused by cardiac or noncardiac signals. A SMART pass filter and two tachyarrhythmia zones were programed.</p><p><strong>Results: </strong>The most frequent underlying heart disease was Brugada/J wave syndrome (n = 7). During a median follow-up period of 1188 days, an OS was observed in six patients (28.6%). The QRS amplitude in synthesized V7 lead (synV7) was significantly lower in the OS group than in the non-OS group (0.59 ± 0.17 vs. 0.91 ± 0.35 mV, p = 0.019). The optimal cutoff value of synV7 QRS amplitude was 0.61 mV, with a sensitivity of 80.0% and a specificity of 83.7% for predicting an OS. Univariate logistic analysis showed that a synV7 QRS amplitude of <0.61 mV was only associated with an OS (odd ratio, 20.0; 95% confidence interval, 1.66-241.72; p = 0.018).</p><p><strong>Conclusions: </strong>In patients with S-ICDs, an OS was not a rare complication during long-term follow-up. A low synV7 QRS amplitude was associated with a high OS incidence.</p>\",\"PeriodicalId\":54653,\"journal\":{\"name\":\"Pace-Pacing and Clinical Electrophysiology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.7000,\"publicationDate\":\"2024-10-05\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Pace-Pacing and Clinical Electrophysiology\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1111/pace.15086\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Pace-Pacing and Clinical Electrophysiology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1111/pace.15086","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
引用次数: 0
摘要
背景:与使用经静脉 ICD 的患者相比,使用皮下植入式心律转复除颤器(S-ICD)的患者更经常出现超感发作(OS)。然而,目前还没有既定的心电图(ECG)参数能准确检测出过感。本研究旨在评估实际临床实践中 OS 的发生率以及合成 18 导联心电图(syn18-ECG)参数与 OS 的关联:我们回顾性地纳入了 21 例连续接受 S-ICD 植入术的患者,并收集了 syn18-ECG 参数。我们将发生器置于深后位,并将 OS 定义为由心脏或非心脏信号引起的不适当充电事件。我们对 SMART 通滤波器和两个快速性心律失常区进行了编程:最常见的潜在心脏病是 Brugada/J 波综合征(7 例)。在中位 1188 天的随访期间,6 名患者(28.6%)出现了 OS。OS 组合成 V7 导联(synV7)的 QRS 波幅明显低于非 OS 组(0.59 ± 0.17 vs. 0.91 ± 0.35 mV,p = 0.019)。synV7 QRS 波幅的最佳临界值为 0.61 mV,预测 OS 的灵敏度为 80.0%,特异度为 83.7%。单变量逻辑分析表明,synV7 QRS振幅为0.61 mV时,OS的预测灵敏度为80.0%,特异度为83.7%:在长期随访过程中,S-ICD 患者发生 OS 的并发症并不罕见。低 synV7 QRS 振幅与高 OS 发生率相关。
Synthesized V7 QRS Amplitude and Oversensing Episodes in Patients With Subcutaneous Implantable Cardioverter-Defibrillators.
Background: Patients with subcutaneous implantable cardioverter-defibrillators (S-ICDs) experience an oversensing episode (OS) more frequently than those with transvenous ICDs. However, no established electrocardiography (ECG) parameters can accurately detect an OS. This study aimed to evaluate the incidence of an OS in real-world clinical practice and the association of synthesized 18-lead ECG (syn18-ECG) parameters with an OS.
Methods: We retrospectively included 21 consecutive patients who underwent S-ICD implantation and collected syn18-ECG parameters. We placed the generator in a deep posterior position and defined an OS as an inappropriate charging episode caused by cardiac or noncardiac signals. A SMART pass filter and two tachyarrhythmia zones were programed.
Results: The most frequent underlying heart disease was Brugada/J wave syndrome (n = 7). During a median follow-up period of 1188 days, an OS was observed in six patients (28.6%). The QRS amplitude in synthesized V7 lead (synV7) was significantly lower in the OS group than in the non-OS group (0.59 ± 0.17 vs. 0.91 ± 0.35 mV, p = 0.019). The optimal cutoff value of synV7 QRS amplitude was 0.61 mV, with a sensitivity of 80.0% and a specificity of 83.7% for predicting an OS. Univariate logistic analysis showed that a synV7 QRS amplitude of <0.61 mV was only associated with an OS (odd ratio, 20.0; 95% confidence interval, 1.66-241.72; p = 0.018).
Conclusions: In patients with S-ICDs, an OS was not a rare complication during long-term follow-up. A low synV7 QRS amplitude was associated with a high OS incidence.
期刊介绍:
Pacing and Clinical Electrophysiology (PACE) is the foremost peer-reviewed journal in the field of pacing and implantable cardioversion defibrillation, publishing over 50% of all English language articles in its field, featuring original, review, and didactic papers, and case reports related to daily practice. Articles also include editorials, book reviews, Musings on humane topics relevant to medical practice, electrophysiology (EP) rounds, device rounds, and information concerning the quality of devices used in the practice of the specialty.