OMIP-069 第 2 版:用于人体外周血主要细胞亚群深度免疫分型的 40 色全光谱流式细胞仪面板更新版

IF 2.5 4区 生物学 Q3 BIOCHEMICAL RESEARCH METHODS
Lily M. Park, Joanne Lannigan, Quentin Low, Maria C. Jaimes, Diana L. Bonilla
{"title":"OMIP-069 第 2 版:用于人体外周血主要细胞亚群深度免疫分型的 40 色全光谱流式细胞仪面板更新版","authors":"Lily M. Park,&nbsp;Joanne Lannigan,&nbsp;Quentin Low,&nbsp;Maria C. Jaimes,&nbsp;Diana L. Bonilla","doi":"10.1002/cyto.a.24898","DOIUrl":null,"url":null,"abstract":"<p>Optimized Multicolor Immunofluorescence Panel (OMIP)-069 was the first optimized flow cytometry panel demonstrating that 40 different fluorochromes can be effectively used in combination, without compromising the resolution of each individual marker [<span>1</span>]. Since its publication, the panel has been adopted in some laboratories (personal communications and [<span>2, 3</span>]). Throughout its adoption, challenges that limited the ability to use the full panel have been reported.</p><p>The challenges can be grouped into two main categories: reagent availability and reagent performance. Concerning reagent availability, CD20 Pacific Orange™ (Thermo Fisher Cat. MHCD2030), CD25 Phycoerythrin (PE) Alexa Fluor™ 700 (Thermo Fisher Cat. MHCD2524), CD24 PE-Alexa Fluor™ 610 (Thermo Fisher Cat. MHCD2422), and CD127 Allophycocyanin (APC) R700 (BD Biosciences Cat. 565185) have been reported to be often on back order. Regarding reagent performance, HLA-DR PE/Fire™ 810 (Biolegend Cat. 307683) and TCRγδ Peridinin-Chlorophyll-Protein (PerCP) eFluor® 710 (Thermo Fisher Cat. 46–9959-42), have shown technical issues, due to tandem degradation or unexpected spectrum signatures, respectively. Finally, it was also documented that spread between fluorescein isothiocyanate (FITC) and Brilliant™ Blue (BB) BB515, the most challenging fluorochrome combination in the panel (similarity index 0.98; spillover spread value of BB515 into FITC 29), was not always consistent with the one reported in the publication, making it difficult to use that fluorochrome pair.</p><p>These issues were investigated, and solutions were identified to enable the use of the full 40-color panel without compromises in performance. First, regarding the PE/Fire 810 tandem stability, the manufacturer stated the fluorochrome meets specifications and the authors did not observe tandem degradation when the reagent was used before its expiration date. To better understand the issues reported by the readership, a stability test was performed with exposure to fixation and light. No stability issues were observed with fixation with paraformaldehyde (1% or 4% in Phosphate buffered saline (PBS)). However, changes in the PE/Fire 810 spectrum and higher spillover values into PE were observed starting at 2 h of light exposure. Tandem breakdown can have serious impacts on the data, in this case specifically with any PE conjugates. For comparison, the PE/Cyanine 7 (Cy7) reagent included in the panel was tested in parallel and exhibited a more stable behavior <b>(</b>Figure 1A<b>).</b> The data suggests that PE/Fire 810 might need to be handled with even more care than other tandems. Next, multiple spectra were observed when using TCRγδ PerCP-eFluor710, without consistency across lots <b>(</b>Figure 1B<b>).</b> Data unmixing and TCRγδ population resolution are highly impacted by the presence of multiple spectra <b>(</b>Figure 1C<b>).</b> Finally, we observed that the use of BB515 and FITC in combination leads to increase in spread not only for those two fluorochromes, but also from Spark Blue 550 and PE. The spread is reduced when either BB515 or FITC are removed <b>(</b>Figure 1D<b>)</b>.</p><p>To overcome the described performance issues and the reagent availability limitations, a rigorous investigation was done to identify fluorochromes and reagents that could be used as alternatives. The goal was to preserve all the markers included in the original publication and to ensure that the panel had equal or better performance. The steps used to identify alternative fluorochromes were the same as described in the original publication: (a) the spectrum needed to be unique, (b) new fluorochromes could not introduce significant spread into the other fluorochromes already in the panel, and (c) the brightness needed to be adequate to address the antigen densities of the new fluorochrome pairs. In addition, stability, tendency to aggregate, and interaction with other fluorochromes were evaluated. Five reagents had a direct fluorochrome replacement with new fluorochromes having similar emission spectrum and similar or higher CD4 resolution compared with the original fluor: Spark Violet 538 for Pacific Orange; cFluor blue yellow green (BYG) BYG710 for PE-Alexa Fluor 700; cFluor R720 for APC-R700; and PerCP-Vio 700 for PerCP-eFluor 710. cFluor Yellow Green (YG) YG610 was identified as an option to replace PE-Alexa Fluor 610, with low excitation by the blue laser, reducing the spread into blue laser excited fluorochromes with similar emission. To overcome the high spread between FITC and BB515, FITC was replaced by cFluor B532. We observed a significantly lower similarity index and lower spillover spread value between BB515 and cFluor B532 (similarity index of 0.97 for BB515-FITC vs. 0.89 for BB515-cFluor 532; [Figure S2A,B]). With careful PE/Fire 810 handling in sample staining, we did not observe stability issues during the development of OMIP-069. However, because of the performance reports we received regarding this reagent, we evaluated alternative fluorochromes to replace it. As mentioned before, we focused on fluorochromes with similar 561 nm excitation and far-red emission, stable to exposure to light and fixation, and with minimal spread impact into other fluorochromes. cFluor BYG750 was selected as a replacement. Resolution was first assessed by measuring stain index of peripheral blood mononuclear cells (PBMCs) labeled with anti-human CD4 conjugated to the new fluorochromes, as shown in Figure S1. Overall, spread was reduced in critical areas, such as the impact on BB515, when using original panel (Spillover spreading value (SSV) from cFluor B532 into BB515 10.99 [Figure S2C]) versus the revised panel (SSV from FITC into BB515 2.95, [Figure S2D]). Figure S2E shows the reduced spread into BB515 and Spark Blue 550 when FITC is replaced using CD4 single stained cells. Similar observations were seen for the markers in the panel when using cells stained with the full 40-color fluorochrome combination (Figure S2F<b>)</b>.</p><p>Reagents of the same specificities and equivalent clones conjugated to the new fluorochromes were then tested. As shown in Figure 2, most of the new conjugates showed a similar resolution and pattern at optimal titer compared with the predicate reagent. However, the staining patterns for CD57 cFluor B532, CD24 cFluor YG610, and TCRγδ PerCP-Vio 700 were slightly different compared to the original reagents (highlighted with asterisks in Figure 5 and Figure S5). For CD57 cFluor B532, we evaluated its performance compared to PE (considered as a gold standard), FITC and Spark Blue 515 conjugates, all from the same clone. Cells were also stained with CD3 to evaluate expression in T and non-T cells. (Figure 3A,B). The staining pattern obtained with CD57 cFluor B532 was equivalent to the ones with the PE and Spark Blue 550 reagents, whereas the original CD57 FITC exhibited a lower percentage of positive cells in both compartments. We next investigated the differences in performance in a multicolor panel that allowed us to identify different T cell subsets. We found that CD57 cFluor B532 performed very similarly to the PE conjugate across T-cell subsets. Based on all these findings, CD57 cFluor B532 was identified as a suitable and superior reagent for the panel compared to the original CD57 FITC. Concerning CD24 cFluor YG610, co-staining with CD19, to identify B cells, and comparison with the original reagent revealed that the staining was specific and provided adequate resolution, and that the new reagent had less non-specific binding to non-B cells and hence included in the final panel (Figure 3C). Finally regarding TCRγδ, although changes in the staining pattern were observed between the original and replacement reagents, the frequencies of positive events remained unchanged (Figure S5). Of note, it was not possible to keep the clone consistent for this marker, but both clones (B1.1-original panel vs. REA591-revised panel) are pan identifiers of all circulating TCRγδ. We hypothesize that the two clones are binding to different conformational epitopes causing the differences in staining profiles, with three diagonals varying in median fluorescence intensity (MFI) observed using the B1.1 clone, in contrast to one single diagonal using REA591 clone. Confirmation that delta variants Vd1 and Vd2 are equally identified with both clones is presented in OMIP-xxx by the expression of TIGIT, CD38, CD27, and CD45RA within the CD3+ TCRγδ+ population.</p><p>As a result of this careful evaluation, the following reagents were identified as suitable alternatives: CD20 Pacific Orange™ was replaced by CD20 Spark Violet™ 538 (Biolegend Cat. 302,373), CD25 PE Alexa Fluor™ 700 by CD25 cFluor® BYG710 (Cytek Biosciences Cat. R7-20660), CD24 PE Alexa Fluor™ 610 by CD24 cFluor® YG610 (Cytek Biosciences Cat. R7-20658), TCRγδ PerCP-eFluor™ 710 by TCRγδ PerCP-Vio® 700 (Miltenyi Cat. 130–113-514), and CD127 APC-R700 by CD127 cFluor® R720 (Cytek Biosciences Cat. R7-20664) [<span>3</span>]. Two specificities were reassigned: HLA-DR PE/Fire™ 810 was replaced by HLA-DR BV480 (BD Biosciences Cat. 752499) and IgD BV480 by IgD cFluor® BYG750 (Cytek Biosciences Cat. R7-20662). Final panel designs (Figure S3A,B) are shown. A list of reagents suggested for this revised panel is included in Table 1. Titration results for the new eight reagents are shown in Figure S4.</p><p>After the replacement fluorochromes and reagents were thoroughly tested in the single-color scenario, their performance was evaluated in fully stained samples comparing the two 40-color combinations (the one originally published in OMIP-069, and the revised one with the replacement fluorochromes). Panel optimization was conducted as described in the original paper, including evaluation of spillover spread, marker resolution assessment, and evaluation of the ability to quantify all described populations by manual gating. A side-by-side comparison of the gating strategy used in OMIP-069 was included for both panels, displaying all populations of interest, all makers in the panel and, highlighting in blue frames, all reagent modifications <b>(</b>Figure S5<b>)</b>. Definition of gate boundaries and identification of subsets of interest using manual gating was equally straightforward. Careful evaluation of performance across different donors allowed us to conclude that resolution obtained with this new combination of fluorochromes is comparable with the original panel. Of note, during assay optimization, modifications were introduced to the staining protocol as well as to the gating strategy to avoid double counting events. For detailed information please refer to OMIP-XXX [<span>4</span>]. In addition, two computational algorithms were used to evaluate if both panels can resolve similarly the heterogeneity of immune subsets identifiable with this combination of markers by high-dimensional analysis. FlowSOM was used for clustering [<span>5</span>] and Uniform Manifold Approximation Projection (UMAP) for dimensionality reduction [<span>6</span>]. Following the analysis strategy presented in OMIP-069, Figure 4 shows representative UMAP plots, overlaid with 34 cell clusters for the entire CD45+ population comparing both panels (original and revised). With both panels, all the populations gated manually can be identified as unique clusters, including subsets identified at different frequencies, from high (natural killer [NK] cell populations, B cell memory subsets, monocytes, CD4+ or CD8+naïve, and memory T cell subsets) to low (γδ T cells, dendritic cells [DC] subsets, basophils, or innate lymphoid cells [ILCs]). UMAP visualization simplified the performance comparison between the original and revised panel, showing similar phenotype across corresponding islands, as well as correlation in island location, size, and proximity in the lymphoid and myeloid sub compartments. Figure 5 shows in detail the performance of highly overlapping fluorochrome pairs in both panels, revealing high resolution for all the newly tested combinations.</p><p>We report that with the identified fluorochrome substitutions, the resolution of all cellular subsets identified with OMIP-069 was preserved, and in some cases improved. We carefully determined that the updated version of the panel has comparable performance, optimal resolution, and robust identification of all subsets of interest. Moreover, based on the experience with OMIP-069, revisions of OMIPs over time seem suitable and critical to keep panels relevant, address issues encountered by the readership or the authors, and benefit from advancements, such as the development of new reagents after publication.</p><p><b>Lily M. Park:</b> Methodology; validation; software; data curation; formal analysis. <b>Joanne Lannigan:</b> Conceptualization; writing – review and editing; data curation. <b>Quentin Low:</b> Methodology; data curation; visualization; formal analysis. <b>Maria C. Jaimes:</b> Conceptualization; methodology; funding acquisition; formal analysis; project administration; resources; supervision; writing – original draft; writing – review and editing. <b>Diana L. Bonilla:</b> Writing – original draft; writing – review and editing; conceptualization; methodology; formal analysis; visualization; supervision; data curation; software; validation.</p>","PeriodicalId":11068,"journal":{"name":"Cytometry Part A","volume":"105 11","pages":"791-799"},"PeriodicalIF":2.5000,"publicationDate":"2024-09-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://onlinelibrary.wiley.com/doi/epdf/10.1002/cyto.a.24898","citationCount":"0","resultStr":"{\"title\":\"OMIP-069 version 2: Update to the 40-color full Spectrum flow cytometry panel for deep immunophenotyping of major cell subsets in human peripheral blood\",\"authors\":\"Lily M. Park,&nbsp;Joanne Lannigan,&nbsp;Quentin Low,&nbsp;Maria C. Jaimes,&nbsp;Diana L. Bonilla\",\"doi\":\"10.1002/cyto.a.24898\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p>Optimized Multicolor Immunofluorescence Panel (OMIP)-069 was the first optimized flow cytometry panel demonstrating that 40 different fluorochromes can be effectively used in combination, without compromising the resolution of each individual marker [<span>1</span>]. Since its publication, the panel has been adopted in some laboratories (personal communications and [<span>2, 3</span>]). Throughout its adoption, challenges that limited the ability to use the full panel have been reported.</p><p>The challenges can be grouped into two main categories: reagent availability and reagent performance. Concerning reagent availability, CD20 Pacific Orange™ (Thermo Fisher Cat. MHCD2030), CD25 Phycoerythrin (PE) Alexa Fluor™ 700 (Thermo Fisher Cat. MHCD2524), CD24 PE-Alexa Fluor™ 610 (Thermo Fisher Cat. MHCD2422), and CD127 Allophycocyanin (APC) R700 (BD Biosciences Cat. 565185) have been reported to be often on back order. Regarding reagent performance, HLA-DR PE/Fire™ 810 (Biolegend Cat. 307683) and TCRγδ Peridinin-Chlorophyll-Protein (PerCP) eFluor® 710 (Thermo Fisher Cat. 46–9959-42), have shown technical issues, due to tandem degradation or unexpected spectrum signatures, respectively. Finally, it was also documented that spread between fluorescein isothiocyanate (FITC) and Brilliant™ Blue (BB) BB515, the most challenging fluorochrome combination in the panel (similarity index 0.98; spillover spread value of BB515 into FITC 29), was not always consistent with the one reported in the publication, making it difficult to use that fluorochrome pair.</p><p>These issues were investigated, and solutions were identified to enable the use of the full 40-color panel without compromises in performance. First, regarding the PE/Fire 810 tandem stability, the manufacturer stated the fluorochrome meets specifications and the authors did not observe tandem degradation when the reagent was used before its expiration date. To better understand the issues reported by the readership, a stability test was performed with exposure to fixation and light. No stability issues were observed with fixation with paraformaldehyde (1% or 4% in Phosphate buffered saline (PBS)). However, changes in the PE/Fire 810 spectrum and higher spillover values into PE were observed starting at 2 h of light exposure. Tandem breakdown can have serious impacts on the data, in this case specifically with any PE conjugates. For comparison, the PE/Cyanine 7 (Cy7) reagent included in the panel was tested in parallel and exhibited a more stable behavior <b>(</b>Figure 1A<b>).</b> The data suggests that PE/Fire 810 might need to be handled with even more care than other tandems. Next, multiple spectra were observed when using TCRγδ PerCP-eFluor710, without consistency across lots <b>(</b>Figure 1B<b>).</b> Data unmixing and TCRγδ population resolution are highly impacted by the presence of multiple spectra <b>(</b>Figure 1C<b>).</b> Finally, we observed that the use of BB515 and FITC in combination leads to increase in spread not only for those two fluorochromes, but also from Spark Blue 550 and PE. The spread is reduced when either BB515 or FITC are removed <b>(</b>Figure 1D<b>)</b>.</p><p>To overcome the described performance issues and the reagent availability limitations, a rigorous investigation was done to identify fluorochromes and reagents that could be used as alternatives. The goal was to preserve all the markers included in the original publication and to ensure that the panel had equal or better performance. The steps used to identify alternative fluorochromes were the same as described in the original publication: (a) the spectrum needed to be unique, (b) new fluorochromes could not introduce significant spread into the other fluorochromes already in the panel, and (c) the brightness needed to be adequate to address the antigen densities of the new fluorochrome pairs. In addition, stability, tendency to aggregate, and interaction with other fluorochromes were evaluated. Five reagents had a direct fluorochrome replacement with new fluorochromes having similar emission spectrum and similar or higher CD4 resolution compared with the original fluor: Spark Violet 538 for Pacific Orange; cFluor blue yellow green (BYG) BYG710 for PE-Alexa Fluor 700; cFluor R720 for APC-R700; and PerCP-Vio 700 for PerCP-eFluor 710. cFluor Yellow Green (YG) YG610 was identified as an option to replace PE-Alexa Fluor 610, with low excitation by the blue laser, reducing the spread into blue laser excited fluorochromes with similar emission. To overcome the high spread between FITC and BB515, FITC was replaced by cFluor B532. We observed a significantly lower similarity index and lower spillover spread value between BB515 and cFluor B532 (similarity index of 0.97 for BB515-FITC vs. 0.89 for BB515-cFluor 532; [Figure S2A,B]). With careful PE/Fire 810 handling in sample staining, we did not observe stability issues during the development of OMIP-069. However, because of the performance reports we received regarding this reagent, we evaluated alternative fluorochromes to replace it. As mentioned before, we focused on fluorochromes with similar 561 nm excitation and far-red emission, stable to exposure to light and fixation, and with minimal spread impact into other fluorochromes. cFluor BYG750 was selected as a replacement. Resolution was first assessed by measuring stain index of peripheral blood mononuclear cells (PBMCs) labeled with anti-human CD4 conjugated to the new fluorochromes, as shown in Figure S1. Overall, spread was reduced in critical areas, such as the impact on BB515, when using original panel (Spillover spreading value (SSV) from cFluor B532 into BB515 10.99 [Figure S2C]) versus the revised panel (SSV from FITC into BB515 2.95, [Figure S2D]). Figure S2E shows the reduced spread into BB515 and Spark Blue 550 when FITC is replaced using CD4 single stained cells. Similar observations were seen for the markers in the panel when using cells stained with the full 40-color fluorochrome combination (Figure S2F<b>)</b>.</p><p>Reagents of the same specificities and equivalent clones conjugated to the new fluorochromes were then tested. As shown in Figure 2, most of the new conjugates showed a similar resolution and pattern at optimal titer compared with the predicate reagent. However, the staining patterns for CD57 cFluor B532, CD24 cFluor YG610, and TCRγδ PerCP-Vio 700 were slightly different compared to the original reagents (highlighted with asterisks in Figure 5 and Figure S5). For CD57 cFluor B532, we evaluated its performance compared to PE (considered as a gold standard), FITC and Spark Blue 515 conjugates, all from the same clone. Cells were also stained with CD3 to evaluate expression in T and non-T cells. (Figure 3A,B). The staining pattern obtained with CD57 cFluor B532 was equivalent to the ones with the PE and Spark Blue 550 reagents, whereas the original CD57 FITC exhibited a lower percentage of positive cells in both compartments. We next investigated the differences in performance in a multicolor panel that allowed us to identify different T cell subsets. We found that CD57 cFluor B532 performed very similarly to the PE conjugate across T-cell subsets. Based on all these findings, CD57 cFluor B532 was identified as a suitable and superior reagent for the panel compared to the original CD57 FITC. Concerning CD24 cFluor YG610, co-staining with CD19, to identify B cells, and comparison with the original reagent revealed that the staining was specific and provided adequate resolution, and that the new reagent had less non-specific binding to non-B cells and hence included in the final panel (Figure 3C). Finally regarding TCRγδ, although changes in the staining pattern were observed between the original and replacement reagents, the frequencies of positive events remained unchanged (Figure S5). Of note, it was not possible to keep the clone consistent for this marker, but both clones (B1.1-original panel vs. REA591-revised panel) are pan identifiers of all circulating TCRγδ. We hypothesize that the two clones are binding to different conformational epitopes causing the differences in staining profiles, with three diagonals varying in median fluorescence intensity (MFI) observed using the B1.1 clone, in contrast to one single diagonal using REA591 clone. Confirmation that delta variants Vd1 and Vd2 are equally identified with both clones is presented in OMIP-xxx by the expression of TIGIT, CD38, CD27, and CD45RA within the CD3+ TCRγδ+ population.</p><p>As a result of this careful evaluation, the following reagents were identified as suitable alternatives: CD20 Pacific Orange™ was replaced by CD20 Spark Violet™ 538 (Biolegend Cat. 302,373), CD25 PE Alexa Fluor™ 700 by CD25 cFluor® BYG710 (Cytek Biosciences Cat. R7-20660), CD24 PE Alexa Fluor™ 610 by CD24 cFluor® YG610 (Cytek Biosciences Cat. R7-20658), TCRγδ PerCP-eFluor™ 710 by TCRγδ PerCP-Vio® 700 (Miltenyi Cat. 130–113-514), and CD127 APC-R700 by CD127 cFluor® R720 (Cytek Biosciences Cat. R7-20664) [<span>3</span>]. Two specificities were reassigned: HLA-DR PE/Fire™ 810 was replaced by HLA-DR BV480 (BD Biosciences Cat. 752499) and IgD BV480 by IgD cFluor® BYG750 (Cytek Biosciences Cat. R7-20662). Final panel designs (Figure S3A,B) are shown. A list of reagents suggested for this revised panel is included in Table 1. Titration results for the new eight reagents are shown in Figure S4.</p><p>After the replacement fluorochromes and reagents were thoroughly tested in the single-color scenario, their performance was evaluated in fully stained samples comparing the two 40-color combinations (the one originally published in OMIP-069, and the revised one with the replacement fluorochromes). Panel optimization was conducted as described in the original paper, including evaluation of spillover spread, marker resolution assessment, and evaluation of the ability to quantify all described populations by manual gating. A side-by-side comparison of the gating strategy used in OMIP-069 was included for both panels, displaying all populations of interest, all makers in the panel and, highlighting in blue frames, all reagent modifications <b>(</b>Figure S5<b>)</b>. Definition of gate boundaries and identification of subsets of interest using manual gating was equally straightforward. Careful evaluation of performance across different donors allowed us to conclude that resolution obtained with this new combination of fluorochromes is comparable with the original panel. Of note, during assay optimization, modifications were introduced to the staining protocol as well as to the gating strategy to avoid double counting events. For detailed information please refer to OMIP-XXX [<span>4</span>]. In addition, two computational algorithms were used to evaluate if both panels can resolve similarly the heterogeneity of immune subsets identifiable with this combination of markers by high-dimensional analysis. FlowSOM was used for clustering [<span>5</span>] and Uniform Manifold Approximation Projection (UMAP) for dimensionality reduction [<span>6</span>]. Following the analysis strategy presented in OMIP-069, Figure 4 shows representative UMAP plots, overlaid with 34 cell clusters for the entire CD45+ population comparing both panels (original and revised). With both panels, all the populations gated manually can be identified as unique clusters, including subsets identified at different frequencies, from high (natural killer [NK] cell populations, B cell memory subsets, monocytes, CD4+ or CD8+naïve, and memory T cell subsets) to low (γδ T cells, dendritic cells [DC] subsets, basophils, or innate lymphoid cells [ILCs]). UMAP visualization simplified the performance comparison between the original and revised panel, showing similar phenotype across corresponding islands, as well as correlation in island location, size, and proximity in the lymphoid and myeloid sub compartments. Figure 5 shows in detail the performance of highly overlapping fluorochrome pairs in both panels, revealing high resolution for all the newly tested combinations.</p><p>We report that with the identified fluorochrome substitutions, the resolution of all cellular subsets identified with OMIP-069 was preserved, and in some cases improved. We carefully determined that the updated version of the panel has comparable performance, optimal resolution, and robust identification of all subsets of interest. Moreover, based on the experience with OMIP-069, revisions of OMIPs over time seem suitable and critical to keep panels relevant, address issues encountered by the readership or the authors, and benefit from advancements, such as the development of new reagents after publication.</p><p><b>Lily M. Park:</b> Methodology; validation; software; data curation; formal analysis. <b>Joanne Lannigan:</b> Conceptualization; writing – review and editing; data curation. <b>Quentin Low:</b> Methodology; data curation; visualization; formal analysis. <b>Maria C. Jaimes:</b> Conceptualization; methodology; funding acquisition; formal analysis; project administration; resources; supervision; writing – original draft; writing – review and editing. <b>Diana L. Bonilla:</b> Writing – original draft; writing – review and editing; conceptualization; methodology; formal analysis; visualization; supervision; data curation; software; validation.</p>\",\"PeriodicalId\":11068,\"journal\":{\"name\":\"Cytometry Part A\",\"volume\":\"105 11\",\"pages\":\"791-799\"},\"PeriodicalIF\":2.5000,\"publicationDate\":\"2024-09-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://onlinelibrary.wiley.com/doi/epdf/10.1002/cyto.a.24898\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Cytometry Part A\",\"FirstCategoryId\":\"99\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1002/cyto.a.24898\",\"RegionNum\":4,\"RegionCategory\":\"生物学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"BIOCHEMICAL RESEARCH METHODS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cytometry Part A","FirstCategoryId":"99","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1002/cyto.a.24898","RegionNum":4,"RegionCategory":"生物学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"BIOCHEMICAL RESEARCH METHODS","Score":null,"Total":0}
引用次数: 0

摘要

优化的多色免疫荧光面板(OMIP)-069 是第一个优化的流式细胞仪面板,它证明了 40 种不同的荧光素可以有效地组合使用,而不会影响每个标记物的分辨率[1]。自发布以来,一些实验室已经采用了该面板(个人通信和[2, 3])。这些挑战可分为两大类:试剂的可用性和试剂的性能。在试剂供应方面,CD20 Pacific Orange™ (Thermo Fisher Cat. MHCD2030)、CD25 Phycoerythrin (PE) Alexa Fluor™ 700 (Thermo Fisher Cat. MHCD2524)、CD24 PE-Alexa Fluor™ 610 (Thermo Fisher Cat. MHCD2422)和 CD127 Allophycocyanin (APC) R700 (BD Biosciences Cat. 565185)据报道经常处于滞销状态。在试剂性能方面,HLA-DR PE/Fire™ 810(Biolegend Cat.307683)和 TCRγδ Peridinin-Chlorophyll-Protein (PerCP) eFluor® 710(Thermo Fisher Cat.46-9959-42)分别出现了串联降解或意外光谱特征等技术问题。最后,还发现异硫氰酸荧光素(FITC)和亮蓝(BB)BB515 之间的扩散(相似性指数为 0.98;BB515 对 FITC 的溢出扩散值为 29)并不总是与出版物中报告的一致,因此很难使用这对荧光素。首先,关于 PE/Fire 810 的串联稳定性,生产商称该荧光染料符合规格要求,作者在试剂有效期前使用时也未观察到串联降解现象。为了更好地了解读者反映的问题,我们进行了一次暴露于固定和光照下的稳定性测试。用多聚甲醛(1% 或 4% 磷酸盐缓冲盐水 (PBS))固定时未发现稳定性问题。但是,从光照 2 小时开始,PE/Fire 810 光谱发生变化,PE 中的溢出值升高。串联分解会对数据产生严重影响,在这种情况下,任何 PE 共轭物都是如此。为了进行比较,我们同时测试了面板中的 PE/Cyanine 7 (Cy7)试剂,其表现更为稳定(图 1A)。这些数据表明,处理 PE/Fire 810 时可能需要比处理其他串联试剂更加小心。接下来,在使用 TCRγδ PerCP-eFluor710 时观察到多个光谱,不同批次的光谱不一致(图 1B)。多光谱的存在严重影响了数据非混合和 TCRγδ 群体分辨率(图 1C)。最后,我们观察到,结合使用 BB515 和 FITC 不仅会增加这两种荧光色素的扩散,还会增加 Spark Blue 550 和 PE 的扩散。为了克服上述性能问题和试剂供应的限制,我们进行了严格的调查,以确定可用作替代品的荧光色素和试剂。我们的目标是保留原始出版物中包含的所有标记物,并确保面板具有相同或更好的性能。用于确定替代荧光素的步骤与原始出版物中描述的相同:(a) 光谱必须是唯一的;(b) 新荧光素不能给面板中的其他荧光素带来明显的扩散;(c) 亮度必须足以应对新荧光素对的抗原密度。此外,还对试剂的稳定性、聚集倾向以及与其他荧光色素的相互作用进行了评估。有五种试剂直接用新的荧光染料替代了原有的荧光染料,新的荧光染料具有相似的发射光谱和相似或更高的 CD4 分辨率:cFluor Yellow Green (YG) YG610 被认为是替代 PE-Alexa Fluor 610 的一种选择,它受蓝色激光的激发较低,从而减少了蓝色激光激发的荧光色素与相似发射的扩散。为了克服 FITC 和 BB515 之间的高扩散,我们用 cFluor B532 取代了 FITC。我们观察到,BB515 和 cFluor B532 之间的相似指数和溢出扩散值都明显较低(BB515-FITC 的相似指数为 0.97,而 BB515-cFluor 532 的相似指数为 0.89;[图 S2A,B])。由于在样本染色过程中对 PE/Fire 810 进行了仔细处理,我们在 OMIP-069 的开发过程中没有发现稳定性问题。然而,由于我们收到的有关该试剂的性能报告,我们评估了替代该试剂的荧光素。
本文章由计算机程序翻译,如有差异,请以英文原文为准。

OMIP-069 version 2: Update to the 40-color full Spectrum flow cytometry panel for deep immunophenotyping of major cell subsets in human peripheral blood

OMIP-069 version 2: Update to the 40-color full Spectrum flow cytometry panel for deep immunophenotyping of major cell subsets in human peripheral blood

Optimized Multicolor Immunofluorescence Panel (OMIP)-069 was the first optimized flow cytometry panel demonstrating that 40 different fluorochromes can be effectively used in combination, without compromising the resolution of each individual marker [1]. Since its publication, the panel has been adopted in some laboratories (personal communications and [2, 3]). Throughout its adoption, challenges that limited the ability to use the full panel have been reported.

The challenges can be grouped into two main categories: reagent availability and reagent performance. Concerning reagent availability, CD20 Pacific Orange™ (Thermo Fisher Cat. MHCD2030), CD25 Phycoerythrin (PE) Alexa Fluor™ 700 (Thermo Fisher Cat. MHCD2524), CD24 PE-Alexa Fluor™ 610 (Thermo Fisher Cat. MHCD2422), and CD127 Allophycocyanin (APC) R700 (BD Biosciences Cat. 565185) have been reported to be often on back order. Regarding reagent performance, HLA-DR PE/Fire™ 810 (Biolegend Cat. 307683) and TCRγδ Peridinin-Chlorophyll-Protein (PerCP) eFluor® 710 (Thermo Fisher Cat. 46–9959-42), have shown technical issues, due to tandem degradation or unexpected spectrum signatures, respectively. Finally, it was also documented that spread between fluorescein isothiocyanate (FITC) and Brilliant™ Blue (BB) BB515, the most challenging fluorochrome combination in the panel (similarity index 0.98; spillover spread value of BB515 into FITC 29), was not always consistent with the one reported in the publication, making it difficult to use that fluorochrome pair.

These issues were investigated, and solutions were identified to enable the use of the full 40-color panel without compromises in performance. First, regarding the PE/Fire 810 tandem stability, the manufacturer stated the fluorochrome meets specifications and the authors did not observe tandem degradation when the reagent was used before its expiration date. To better understand the issues reported by the readership, a stability test was performed with exposure to fixation and light. No stability issues were observed with fixation with paraformaldehyde (1% or 4% in Phosphate buffered saline (PBS)). However, changes in the PE/Fire 810 spectrum and higher spillover values into PE were observed starting at 2 h of light exposure. Tandem breakdown can have serious impacts on the data, in this case specifically with any PE conjugates. For comparison, the PE/Cyanine 7 (Cy7) reagent included in the panel was tested in parallel and exhibited a more stable behavior (Figure 1A). The data suggests that PE/Fire 810 might need to be handled with even more care than other tandems. Next, multiple spectra were observed when using TCRγδ PerCP-eFluor710, without consistency across lots (Figure 1B). Data unmixing and TCRγδ population resolution are highly impacted by the presence of multiple spectra (Figure 1C). Finally, we observed that the use of BB515 and FITC in combination leads to increase in spread not only for those two fluorochromes, but also from Spark Blue 550 and PE. The spread is reduced when either BB515 or FITC are removed (Figure 1D).

To overcome the described performance issues and the reagent availability limitations, a rigorous investigation was done to identify fluorochromes and reagents that could be used as alternatives. The goal was to preserve all the markers included in the original publication and to ensure that the panel had equal or better performance. The steps used to identify alternative fluorochromes were the same as described in the original publication: (a) the spectrum needed to be unique, (b) new fluorochromes could not introduce significant spread into the other fluorochromes already in the panel, and (c) the brightness needed to be adequate to address the antigen densities of the new fluorochrome pairs. In addition, stability, tendency to aggregate, and interaction with other fluorochromes were evaluated. Five reagents had a direct fluorochrome replacement with new fluorochromes having similar emission spectrum and similar or higher CD4 resolution compared with the original fluor: Spark Violet 538 for Pacific Orange; cFluor blue yellow green (BYG) BYG710 for PE-Alexa Fluor 700; cFluor R720 for APC-R700; and PerCP-Vio 700 for PerCP-eFluor 710. cFluor Yellow Green (YG) YG610 was identified as an option to replace PE-Alexa Fluor 610, with low excitation by the blue laser, reducing the spread into blue laser excited fluorochromes with similar emission. To overcome the high spread between FITC and BB515, FITC was replaced by cFluor B532. We observed a significantly lower similarity index and lower spillover spread value between BB515 and cFluor B532 (similarity index of 0.97 for BB515-FITC vs. 0.89 for BB515-cFluor 532; [Figure S2A,B]). With careful PE/Fire 810 handling in sample staining, we did not observe stability issues during the development of OMIP-069. However, because of the performance reports we received regarding this reagent, we evaluated alternative fluorochromes to replace it. As mentioned before, we focused on fluorochromes with similar 561 nm excitation and far-red emission, stable to exposure to light and fixation, and with minimal spread impact into other fluorochromes. cFluor BYG750 was selected as a replacement. Resolution was first assessed by measuring stain index of peripheral blood mononuclear cells (PBMCs) labeled with anti-human CD4 conjugated to the new fluorochromes, as shown in Figure S1. Overall, spread was reduced in critical areas, such as the impact on BB515, when using original panel (Spillover spreading value (SSV) from cFluor B532 into BB515 10.99 [Figure S2C]) versus the revised panel (SSV from FITC into BB515 2.95, [Figure S2D]). Figure S2E shows the reduced spread into BB515 and Spark Blue 550 when FITC is replaced using CD4 single stained cells. Similar observations were seen for the markers in the panel when using cells stained with the full 40-color fluorochrome combination (Figure S2F).

Reagents of the same specificities and equivalent clones conjugated to the new fluorochromes were then tested. As shown in Figure 2, most of the new conjugates showed a similar resolution and pattern at optimal titer compared with the predicate reagent. However, the staining patterns for CD57 cFluor B532, CD24 cFluor YG610, and TCRγδ PerCP-Vio 700 were slightly different compared to the original reagents (highlighted with asterisks in Figure 5 and Figure S5). For CD57 cFluor B532, we evaluated its performance compared to PE (considered as a gold standard), FITC and Spark Blue 515 conjugates, all from the same clone. Cells were also stained with CD3 to evaluate expression in T and non-T cells. (Figure 3A,B). The staining pattern obtained with CD57 cFluor B532 was equivalent to the ones with the PE and Spark Blue 550 reagents, whereas the original CD57 FITC exhibited a lower percentage of positive cells in both compartments. We next investigated the differences in performance in a multicolor panel that allowed us to identify different T cell subsets. We found that CD57 cFluor B532 performed very similarly to the PE conjugate across T-cell subsets. Based on all these findings, CD57 cFluor B532 was identified as a suitable and superior reagent for the panel compared to the original CD57 FITC. Concerning CD24 cFluor YG610, co-staining with CD19, to identify B cells, and comparison with the original reagent revealed that the staining was specific and provided adequate resolution, and that the new reagent had less non-specific binding to non-B cells and hence included in the final panel (Figure 3C). Finally regarding TCRγδ, although changes in the staining pattern were observed between the original and replacement reagents, the frequencies of positive events remained unchanged (Figure S5). Of note, it was not possible to keep the clone consistent for this marker, but both clones (B1.1-original panel vs. REA591-revised panel) are pan identifiers of all circulating TCRγδ. We hypothesize that the two clones are binding to different conformational epitopes causing the differences in staining profiles, with three diagonals varying in median fluorescence intensity (MFI) observed using the B1.1 clone, in contrast to one single diagonal using REA591 clone. Confirmation that delta variants Vd1 and Vd2 are equally identified with both clones is presented in OMIP-xxx by the expression of TIGIT, CD38, CD27, and CD45RA within the CD3+ TCRγδ+ population.

As a result of this careful evaluation, the following reagents were identified as suitable alternatives: CD20 Pacific Orange™ was replaced by CD20 Spark Violet™ 538 (Biolegend Cat. 302,373), CD25 PE Alexa Fluor™ 700 by CD25 cFluor® BYG710 (Cytek Biosciences Cat. R7-20660), CD24 PE Alexa Fluor™ 610 by CD24 cFluor® YG610 (Cytek Biosciences Cat. R7-20658), TCRγδ PerCP-eFluor™ 710 by TCRγδ PerCP-Vio® 700 (Miltenyi Cat. 130–113-514), and CD127 APC-R700 by CD127 cFluor® R720 (Cytek Biosciences Cat. R7-20664) [3]. Two specificities were reassigned: HLA-DR PE/Fire™ 810 was replaced by HLA-DR BV480 (BD Biosciences Cat. 752499) and IgD BV480 by IgD cFluor® BYG750 (Cytek Biosciences Cat. R7-20662). Final panel designs (Figure S3A,B) are shown. A list of reagents suggested for this revised panel is included in Table 1. Titration results for the new eight reagents are shown in Figure S4.

After the replacement fluorochromes and reagents were thoroughly tested in the single-color scenario, their performance was evaluated in fully stained samples comparing the two 40-color combinations (the one originally published in OMIP-069, and the revised one with the replacement fluorochromes). Panel optimization was conducted as described in the original paper, including evaluation of spillover spread, marker resolution assessment, and evaluation of the ability to quantify all described populations by manual gating. A side-by-side comparison of the gating strategy used in OMIP-069 was included for both panels, displaying all populations of interest, all makers in the panel and, highlighting in blue frames, all reagent modifications (Figure S5). Definition of gate boundaries and identification of subsets of interest using manual gating was equally straightforward. Careful evaluation of performance across different donors allowed us to conclude that resolution obtained with this new combination of fluorochromes is comparable with the original panel. Of note, during assay optimization, modifications were introduced to the staining protocol as well as to the gating strategy to avoid double counting events. For detailed information please refer to OMIP-XXX [4]. In addition, two computational algorithms were used to evaluate if both panels can resolve similarly the heterogeneity of immune subsets identifiable with this combination of markers by high-dimensional analysis. FlowSOM was used for clustering [5] and Uniform Manifold Approximation Projection (UMAP) for dimensionality reduction [6]. Following the analysis strategy presented in OMIP-069, Figure 4 shows representative UMAP plots, overlaid with 34 cell clusters for the entire CD45+ population comparing both panels (original and revised). With both panels, all the populations gated manually can be identified as unique clusters, including subsets identified at different frequencies, from high (natural killer [NK] cell populations, B cell memory subsets, monocytes, CD4+ or CD8+naïve, and memory T cell subsets) to low (γδ T cells, dendritic cells [DC] subsets, basophils, or innate lymphoid cells [ILCs]). UMAP visualization simplified the performance comparison between the original and revised panel, showing similar phenotype across corresponding islands, as well as correlation in island location, size, and proximity in the lymphoid and myeloid sub compartments. Figure 5 shows in detail the performance of highly overlapping fluorochrome pairs in both panels, revealing high resolution for all the newly tested combinations.

We report that with the identified fluorochrome substitutions, the resolution of all cellular subsets identified with OMIP-069 was preserved, and in some cases improved. We carefully determined that the updated version of the panel has comparable performance, optimal resolution, and robust identification of all subsets of interest. Moreover, based on the experience with OMIP-069, revisions of OMIPs over time seem suitable and critical to keep panels relevant, address issues encountered by the readership or the authors, and benefit from advancements, such as the development of new reagents after publication.

Lily M. Park: Methodology; validation; software; data curation; formal analysis. Joanne Lannigan: Conceptualization; writing – review and editing; data curation. Quentin Low: Methodology; data curation; visualization; formal analysis. Maria C. Jaimes: Conceptualization; methodology; funding acquisition; formal analysis; project administration; resources; supervision; writing – original draft; writing – review and editing. Diana L. Bonilla: Writing – original draft; writing – review and editing; conceptualization; methodology; formal analysis; visualization; supervision; data curation; software; validation.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Cytometry Part A
Cytometry Part A 生物-生化研究方法
CiteScore
8.10
自引率
13.50%
发文量
183
审稿时长
4-8 weeks
期刊介绍: Cytometry Part A, the journal of quantitative single-cell analysis, features original research reports and reviews of innovative scientific studies employing quantitative single-cell measurement, separation, manipulation, and modeling techniques, as well as original articles on mechanisms of molecular and cellular functions obtained by cytometry techniques. The journal welcomes submissions from multiple research fields that fully embrace the study of the cytome: Biomedical Instrumentation Engineering Biophotonics Bioinformatics Cell Biology Computational Biology Data Science Immunology Parasitology Microbiology Neuroscience Cancer Stem Cells Tissue Regeneration.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信