{"title":"3Dπ:三维正电子成像,一种使用掺氙液态氩闪烁体的新型全身 PET 扫描仪","authors":"Azam Zabihi, Xinran Li, Alejandro Ramirez, Manuel D. Da Rocha Rolo, Davide Franco, Federico Gabriele, Cristiano Galbiati, Michela Lai, Daniel R. Marlow, Andrew Renshaw, Shawn Westerdale, Masayuki Wada","doi":"arxiv-2408.14645","DOIUrl":null,"url":null,"abstract":"Objective: This paper introduces a novel PET imaging methodology called\n3-dimensional positron imaging (3D{\\pi}), which integrates total-body (TB)\ncoverage, time-of-flight (TOF) technology, ultra-low dose imaging capabilities,\nand ultra-fast readout electronics inspired by emerging technology from the\nDarkSide collaboration. Approach: The study evaluates the performance of\n3D{\\pi} using Monte Carlo simulations based on NEMA NU 2-2018 protocols. The\nmethodology employs a homogenous, monolithic scintillator composed of liquid\nargon (LAr) doped with xenon (Xe) with silicon photomultipliers (SiPM)\noperating at cryogenic temperatures. Main results: Significant enhancements in\nsystem performance are observed, with the 3D{\\pi} system achieving a noise\nequivalent count rate (NECR) of 3.2 Mcps which is approximately two times\nhigher than uEXPLORER's peak NECR (1.5 Mcps) at 17.3 (kBq/mL). Spatial\nresolution measurements show an average FWHM of 2.7 mm across both axial\npositions. The system exhibits superior sensitivity, with values reaching 373\nkcps/MBq with a line source at the center of the field of view. Additionally,\n3D{\\pi} achieves a TOF resolution of 151 ps at 5.3 kBq/mL, highlighting its\npotential to produce high-quality images with reduced noise levels.\nSignificance: The study underscores the potential of 3D{\\pi} in improving PET\nimaging performance, offering the potential for shorter scan times and reduced\nradiation exposure for patients. The Xe-doped LAr offers advantages such as\nfast scintillation, enhanced light yield, and cost-effectiveness. Future\nresearch will focus on optimizing system geometry and further refining\nreconstruction algorithms to exploit the strengths of 3D{\\pi} for clinical\napplications.","PeriodicalId":501378,"journal":{"name":"arXiv - PHYS - Medical Physics","volume":"1 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-08-26","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"3Dπ: Three-Dimensional Positron Imaging, A Novel Total-Body PET Scanner Using Xenon-Doped Liquid Argon Scintillator\",\"authors\":\"Azam Zabihi, Xinran Li, Alejandro Ramirez, Manuel D. Da Rocha Rolo, Davide Franco, Federico Gabriele, Cristiano Galbiati, Michela Lai, Daniel R. Marlow, Andrew Renshaw, Shawn Westerdale, Masayuki Wada\",\"doi\":\"arxiv-2408.14645\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective: This paper introduces a novel PET imaging methodology called\\n3-dimensional positron imaging (3D{\\\\pi}), which integrates total-body (TB)\\ncoverage, time-of-flight (TOF) technology, ultra-low dose imaging capabilities,\\nand ultra-fast readout electronics inspired by emerging technology from the\\nDarkSide collaboration. Approach: The study evaluates the performance of\\n3D{\\\\pi} using Monte Carlo simulations based on NEMA NU 2-2018 protocols. The\\nmethodology employs a homogenous, monolithic scintillator composed of liquid\\nargon (LAr) doped with xenon (Xe) with silicon photomultipliers (SiPM)\\noperating at cryogenic temperatures. Main results: Significant enhancements in\\nsystem performance are observed, with the 3D{\\\\pi} system achieving a noise\\nequivalent count rate (NECR) of 3.2 Mcps which is approximately two times\\nhigher than uEXPLORER's peak NECR (1.5 Mcps) at 17.3 (kBq/mL). Spatial\\nresolution measurements show an average FWHM of 2.7 mm across both axial\\npositions. The system exhibits superior sensitivity, with values reaching 373\\nkcps/MBq with a line source at the center of the field of view. Additionally,\\n3D{\\\\pi} achieves a TOF resolution of 151 ps at 5.3 kBq/mL, highlighting its\\npotential to produce high-quality images with reduced noise levels.\\nSignificance: The study underscores the potential of 3D{\\\\pi} in improving PET\\nimaging performance, offering the potential for shorter scan times and reduced\\nradiation exposure for patients. The Xe-doped LAr offers advantages such as\\nfast scintillation, enhanced light yield, and cost-effectiveness. Future\\nresearch will focus on optimizing system geometry and further refining\\nreconstruction algorithms to exploit the strengths of 3D{\\\\pi} for clinical\\napplications.\",\"PeriodicalId\":501378,\"journal\":{\"name\":\"arXiv - PHYS - Medical Physics\",\"volume\":\"1 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-08-26\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"arXiv - PHYS - Medical Physics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/arxiv-2408.14645\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"arXiv - PHYS - Medical Physics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/arxiv-2408.14645","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0
摘要
目的:本文介绍了一种名为三维正电子成像(3D{\pi})的新型 PET 成像方法,它集成了全身(TB)覆盖、飞行时间(TOF)技术、超低剂量成像能力和超快读出电子装置,其灵感来自于黑暗面合作组织(DarkSide collaboration)的新兴技术。方法:该研究使用基于NEMA NU 2-2018协议的蒙特卡罗模拟来评估3D{pi}的性能。该方法采用了由掺杂氙(Xe)的液氩(LAr)和在低温下工作的硅光电倍增管(SiPM)组成的同质单片闪烁器。主要结果:观察到系统性能显著提高,3D{pi}系统的噪声等效计数率(NECR)达到了3.2 Mcps,比uEXPLORER在17.3(kBq/mL)时的峰值NECR(1.5 Mcps)高出约两倍。空间分辨率测量显示,两个轴向位置的平均 FWHM 为 2.7 毫米。该系统显示出卓越的灵敏度,在视场中心的线源上,灵敏度值达到 373kcps/MBq。此外,3D{pi} 在 5.3 kBq/mL 时实现了 151 ps 的 TOF 分辨率,凸显了其在降低噪声水平的同时生成高质量图像的潜力:这项研究强调了 3D{pi} 在提高 PET 成像性能方面的潜力,为缩短扫描时间和减少患者的辐射暴露提供了可能。掺Xe的LAr具有快速闪烁、提高光产率和成本效益等优势。未来的研究将侧重于优化系统几何结构和进一步完善重建算法,以发挥三维{pi}在临床应用中的优势。
3Dπ: Three-Dimensional Positron Imaging, A Novel Total-Body PET Scanner Using Xenon-Doped Liquid Argon Scintillator
Objective: This paper introduces a novel PET imaging methodology called
3-dimensional positron imaging (3D{\pi}), which integrates total-body (TB)
coverage, time-of-flight (TOF) technology, ultra-low dose imaging capabilities,
and ultra-fast readout electronics inspired by emerging technology from the
DarkSide collaboration. Approach: The study evaluates the performance of
3D{\pi} using Monte Carlo simulations based on NEMA NU 2-2018 protocols. The
methodology employs a homogenous, monolithic scintillator composed of liquid
argon (LAr) doped with xenon (Xe) with silicon photomultipliers (SiPM)
operating at cryogenic temperatures. Main results: Significant enhancements in
system performance are observed, with the 3D{\pi} system achieving a noise
equivalent count rate (NECR) of 3.2 Mcps which is approximately two times
higher than uEXPLORER's peak NECR (1.5 Mcps) at 17.3 (kBq/mL). Spatial
resolution measurements show an average FWHM of 2.7 mm across both axial
positions. The system exhibits superior sensitivity, with values reaching 373
kcps/MBq with a line source at the center of the field of view. Additionally,
3D{\pi} achieves a TOF resolution of 151 ps at 5.3 kBq/mL, highlighting its
potential to produce high-quality images with reduced noise levels.
Significance: The study underscores the potential of 3D{\pi} in improving PET
imaging performance, offering the potential for shorter scan times and reduced
radiation exposure for patients. The Xe-doped LAr offers advantages such as
fast scintillation, enhanced light yield, and cost-effectiveness. Future
research will focus on optimizing system geometry and further refining
reconstruction algorithms to exploit the strengths of 3D{\pi} for clinical
applications.