{"title":"在多通道阴道圆筒脉冲剂量率近距离放射治疗中使用无机闪烁探测器进行体内剂量测定:脉冲剂量率近距离放射治疗的剂量测定","authors":"Peter D. Georgi , Søren K. Nielsen , Anders T. Hansen , Harald Spejlborg , Susanne Rylander , Jacob Lindegaard , Simon Buus , Christian Wulff , Primoz Petric , Kari Tanderup , Jacob G. Johansen","doi":"10.1016/j.phro.2024.100638","DOIUrl":null,"url":null,"abstract":"<div><h3>Background and purpose</h3><p>In vivo dosimetry is not standard in brachytherapy and some errors go undetected. The aim of this study was to evaluate the accuracy of multi-channel vaginal cylinder pulsed dose-rate brachytherapy using in vivo dosimetry.</p></div><div><h3>Materials and methods</h3><p>In vivo dosimetry data was collected during the years 2019–2022 for 22 patients (32 fractions) receiving multi-channel cylinder pulsed dose-rate brachytherapy. An inorganic scintillation detector was inserted in a cylinder channel. Each fraction was analysed as independent data sets. In vivo dosimetry-based source-tracking was used to determine the relative source-to-detector position. Measured dose was compared to planned and re-calculated source-tracking based doses. Assuming no change in organ and applicator geometry throughout treatment, the planned and source-tracking based dose distributions were compared in select volumes via γ-index analysis and dose-volume-histograms.</p></div><div><h3>Results</h3><p>The mean ± SD planned vs. measured dose deviations in the first pulse were 0.8 <span><math><mrow><mo>±</mo></mrow></math></span> 5.9 %. In 31/32 fractions the deviation was within the combined in vivo dosimetry uncertainty (averaging 9.7 %, <em>k =</em> 2) and planning dose calculation uncertainty (1.6 %, <em>k =</em> 2). The dwell-position offsets were < 2 mm for 88 % of channels, with the largest being 5.1 mm (4.0 mm uncertainty, <em>k =</em> 2). 3 %/2 mm γ pass-rates averaged 97.0 % (clinical target volume (CTV)), 100.0 % (rectum), 99.9 % (bladder). The mean ± SD deviation was −1.<span><math><mrow><mn>1</mn></mrow></math></span> ± 2.9 % for CTV D98, and −0.2 ± 0.9 % and −1.2 ± 2.5 %, for bladder and rectum D2cm<sup>3</sup> respectively, indicating good agreement between intended and delivered dose.</p></div><div><h3>Conclusions</h3><p>In vivo dosimetry verified accurate and stable dose delivery in multi-channel vaginal cylinder based pulsed dose-rate brachytherapy.</p></div>","PeriodicalId":36850,"journal":{"name":"Physics and Imaging in Radiation Oncology","volume":"32 ","pages":"Article 100638"},"PeriodicalIF":3.4000,"publicationDate":"2024-08-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S2405631624001088/pdfft?md5=e582505d93f2167330a052ddaf354c3b&pid=1-s2.0-S2405631624001088-main.pdf","citationCount":"0","resultStr":"{\"title\":\"In vivo dosimetry with an inorganic scintillation detector during multi-channel vaginal cylinder pulsed dose-rate brachytherapy: Dosimetry for pulsed dose-rate brachytherapy\",\"authors\":\"Peter D. Georgi , Søren K. Nielsen , Anders T. Hansen , Harald Spejlborg , Susanne Rylander , Jacob Lindegaard , Simon Buus , Christian Wulff , Primoz Petric , Kari Tanderup , Jacob G. Johansen\",\"doi\":\"10.1016/j.phro.2024.100638\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Background and purpose</h3><p>In vivo dosimetry is not standard in brachytherapy and some errors go undetected. The aim of this study was to evaluate the accuracy of multi-channel vaginal cylinder pulsed dose-rate brachytherapy using in vivo dosimetry.</p></div><div><h3>Materials and methods</h3><p>In vivo dosimetry data was collected during the years 2019–2022 for 22 patients (32 fractions) receiving multi-channel cylinder pulsed dose-rate brachytherapy. An inorganic scintillation detector was inserted in a cylinder channel. Each fraction was analysed as independent data sets. In vivo dosimetry-based source-tracking was used to determine the relative source-to-detector position. Measured dose was compared to planned and re-calculated source-tracking based doses. Assuming no change in organ and applicator geometry throughout treatment, the planned and source-tracking based dose distributions were compared in select volumes via γ-index analysis and dose-volume-histograms.</p></div><div><h3>Results</h3><p>The mean ± SD planned vs. measured dose deviations in the first pulse were 0.8 <span><math><mrow><mo>±</mo></mrow></math></span> 5.9 %. In 31/32 fractions the deviation was within the combined in vivo dosimetry uncertainty (averaging 9.7 %, <em>k =</em> 2) and planning dose calculation uncertainty (1.6 %, <em>k =</em> 2). The dwell-position offsets were < 2 mm for 88 % of channels, with the largest being 5.1 mm (4.0 mm uncertainty, <em>k =</em> 2). 3 %/2 mm γ pass-rates averaged 97.0 % (clinical target volume (CTV)), 100.0 % (rectum), 99.9 % (bladder). The mean ± SD deviation was −1.<span><math><mrow><mn>1</mn></mrow></math></span> ± 2.9 % for CTV D98, and −0.2 ± 0.9 % and −1.2 ± 2.5 %, for bladder and rectum D2cm<sup>3</sup> respectively, indicating good agreement between intended and delivered dose.</p></div><div><h3>Conclusions</h3><p>In vivo dosimetry verified accurate and stable dose delivery in multi-channel vaginal cylinder based pulsed dose-rate brachytherapy.</p></div>\",\"PeriodicalId\":36850,\"journal\":{\"name\":\"Physics and Imaging in Radiation Oncology\",\"volume\":\"32 \",\"pages\":\"Article 100638\"},\"PeriodicalIF\":3.4000,\"publicationDate\":\"2024-08-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S2405631624001088/pdfft?md5=e582505d93f2167330a052ddaf354c3b&pid=1-s2.0-S2405631624001088-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Physics and Imaging in Radiation Oncology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S2405631624001088\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ONCOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics and Imaging in Radiation Oncology","FirstCategoryId":"1085","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S2405631624001088","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ONCOLOGY","Score":null,"Total":0}
In vivo dosimetry with an inorganic scintillation detector during multi-channel vaginal cylinder pulsed dose-rate brachytherapy: Dosimetry for pulsed dose-rate brachytherapy
Background and purpose
In vivo dosimetry is not standard in brachytherapy and some errors go undetected. The aim of this study was to evaluate the accuracy of multi-channel vaginal cylinder pulsed dose-rate brachytherapy using in vivo dosimetry.
Materials and methods
In vivo dosimetry data was collected during the years 2019–2022 for 22 patients (32 fractions) receiving multi-channel cylinder pulsed dose-rate brachytherapy. An inorganic scintillation detector was inserted in a cylinder channel. Each fraction was analysed as independent data sets. In vivo dosimetry-based source-tracking was used to determine the relative source-to-detector position. Measured dose was compared to planned and re-calculated source-tracking based doses. Assuming no change in organ and applicator geometry throughout treatment, the planned and source-tracking based dose distributions were compared in select volumes via γ-index analysis and dose-volume-histograms.
Results
The mean ± SD planned vs. measured dose deviations in the first pulse were 0.8 5.9 %. In 31/32 fractions the deviation was within the combined in vivo dosimetry uncertainty (averaging 9.7 %, k = 2) and planning dose calculation uncertainty (1.6 %, k = 2). The dwell-position offsets were < 2 mm for 88 % of channels, with the largest being 5.1 mm (4.0 mm uncertainty, k = 2). 3 %/2 mm γ pass-rates averaged 97.0 % (clinical target volume (CTV)), 100.0 % (rectum), 99.9 % (bladder). The mean ± SD deviation was −1. ± 2.9 % for CTV D98, and −0.2 ± 0.9 % and −1.2 ± 2.5 %, for bladder and rectum D2cm3 respectively, indicating good agreement between intended and delivered dose.
Conclusions
In vivo dosimetry verified accurate and stable dose delivery in multi-channel vaginal cylinder based pulsed dose-rate brachytherapy.