Qiao-Wei Du , Fan Xiao , Lin Zheng , Ren-dong Chen , Li-Nan Dong , Fang-Yi Liu , Zhi-Gang Cheng , Jie Yu , Ping Liang
{"title":"增强型冷却系统对更多球形消融区的重要性:数值模拟、体内外验证","authors":"Qiao-Wei Du , Fan Xiao , Lin Zheng , Ren-dong Chen , Li-Nan Dong , Fang-Yi Liu , Zhi-Gang Cheng , Jie Yu , Ping Liang","doi":"10.1016/j.cmpb.2024.108383","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>This study aimed to investigate the efficacy of a small-gauge microwave ablation antenna (MWA) with an enhanced cooling system (ECS) for generating more spherical ablation zones.</p></div><div><h3>Methods</h3><p>A comparison was made between two types of microwave ablation antennas, one with ECS and the other with a conventional cooling system (CCS). The finite element method was used to simulate in vivo ablation. Two types of antennas were used to create MWA zones for 5, 8, 10 min at 50, 60, and 80 W in <em>ex vivo</em> bovine livers (n = 6) and 5 min at 60 W <em>in vivo</em> porcine livers (n = 16). The overtreatment ratio, ablation aspect ratio, carbonization area, and other characteristcs of antennas were measured and compared using numerical simulation and gross pathologic examination.</p></div><div><h3>Results</h3><p>In numerical simulation, the ECS antenna demonstrated a lower overtreatment ratio than the CCS antenna (1.38 vs 1.43 at 50 W 5 min, 1.19 vs 1.35 at 50 W 8 min, 1.13 vs 1.32 at 50 W 10 min, 1.28 vs 1.38 at 60 W 5 min, 1.14 vs 1.32 at 60 W 8 min, 1.10 vs 1.30 at 60 W 10 min). The experiments revealed that the ECS antenna generated ablation zones with a more significant aspect ratio (0.92 ± 0.03 vs 0.72 ± 0.01 at 50 W 5 min, 0.95 ± 0.02 vs 0.70 ± 0.01 at 50 W 8 min, 0.96 ± 0.01 vs 0.71 ± 0.04 at 50 W 10 min, 0.96 ± 0.01 vs 0.73 ± 0.02 at 60 W 5 min, 0.94 ± 0.03 vs 0.71 ± 0.03 at 60 W 8 min, 0.96 ± 0.02 vs 0.69 ± 0.04 at 60 W 10 min) and a smaller carbonization area (0.00 ± 0.00 cm<sup>2</sup> vs 0.54 ± 0.06 cm<sup>2</sup> at 50 W 5 min, 0.13 ± 0.03 cm<sup>2</sup> vs 0.61 ± 0.09 cm<sup>2</sup> at 50 W 8 min, 0.23 ± 0.05 cm<sup>2</sup> vs 0.73 ± 0.05 m<sup>2</sup> at 50 W 10 min, 0.00 ± 0.00 cm<sup>2</sup> vs 1.59 ± 0.41 cm<sup>2</sup> at 60 W 5 min, 0.23 ± 0.22 cm<sup>2</sup> vs 2.11 ± 0.63 cm<sup>2</sup> at 60 W 8 min, 0.57 ± 0.09 cm<sup>2</sup> vs 2.55 ± 0.51 cm<sup>2</sup> at 60 W 10 min). Intraoperative ultrasound images revealed a hypoechoic area instead of a hyperechoic area near the antenna. Hematoxylin-eosin staining of the dissected tissue revealed a correlation between the edge of the ablation zone and that of the hypoechoic area.</p></div><div><h3>Conclusions</h3><p>The ECS antenna can produce more spherical ablation zones with less charring and a clearer intraoperative ultrasound image of the ablation area than the CCS antenna.</p></div>","PeriodicalId":10624,"journal":{"name":"Computer methods and programs in biomedicine","volume":"257 ","pages":"Article 108383"},"PeriodicalIF":4.9000,"publicationDate":"2024-08-23","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Importance of the enhanced cooling system for more spherical ablation zones: Numerical simulation, ex vivo and in vivo validation\",\"authors\":\"Qiao-Wei Du , Fan Xiao , Lin Zheng , Ren-dong Chen , Li-Nan Dong , Fang-Yi Liu , Zhi-Gang Cheng , Jie Yu , Ping Liang\",\"doi\":\"10.1016/j.cmpb.2024.108383\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Introduction</h3><p>This study aimed to investigate the efficacy of a small-gauge microwave ablation antenna (MWA) with an enhanced cooling system (ECS) for generating more spherical ablation zones.</p></div><div><h3>Methods</h3><p>A comparison was made between two types of microwave ablation antennas, one with ECS and the other with a conventional cooling system (CCS). The finite element method was used to simulate in vivo ablation. Two types of antennas were used to create MWA zones for 5, 8, 10 min at 50, 60, and 80 W in <em>ex vivo</em> bovine livers (n = 6) and 5 min at 60 W <em>in vivo</em> porcine livers (n = 16). The overtreatment ratio, ablation aspect ratio, carbonization area, and other characteristcs of antennas were measured and compared using numerical simulation and gross pathologic examination.</p></div><div><h3>Results</h3><p>In numerical simulation, the ECS antenna demonstrated a lower overtreatment ratio than the CCS antenna (1.38 vs 1.43 at 50 W 5 min, 1.19 vs 1.35 at 50 W 8 min, 1.13 vs 1.32 at 50 W 10 min, 1.28 vs 1.38 at 60 W 5 min, 1.14 vs 1.32 at 60 W 8 min, 1.10 vs 1.30 at 60 W 10 min). The experiments revealed that the ECS antenna generated ablation zones with a more significant aspect ratio (0.92 ± 0.03 vs 0.72 ± 0.01 at 50 W 5 min, 0.95 ± 0.02 vs 0.70 ± 0.01 at 50 W 8 min, 0.96 ± 0.01 vs 0.71 ± 0.04 at 50 W 10 min, 0.96 ± 0.01 vs 0.73 ± 0.02 at 60 W 5 min, 0.94 ± 0.03 vs 0.71 ± 0.03 at 60 W 8 min, 0.96 ± 0.02 vs 0.69 ± 0.04 at 60 W 10 min) and a smaller carbonization area (0.00 ± 0.00 cm<sup>2</sup> vs 0.54 ± 0.06 cm<sup>2</sup> at 50 W 5 min, 0.13 ± 0.03 cm<sup>2</sup> vs 0.61 ± 0.09 cm<sup>2</sup> at 50 W 8 min, 0.23 ± 0.05 cm<sup>2</sup> vs 0.73 ± 0.05 m<sup>2</sup> at 50 W 10 min, 0.00 ± 0.00 cm<sup>2</sup> vs 1.59 ± 0.41 cm<sup>2</sup> at 60 W 5 min, 0.23 ± 0.22 cm<sup>2</sup> vs 2.11 ± 0.63 cm<sup>2</sup> at 60 W 8 min, 0.57 ± 0.09 cm<sup>2</sup> vs 2.55 ± 0.51 cm<sup>2</sup> at 60 W 10 min). Intraoperative ultrasound images revealed a hypoechoic area instead of a hyperechoic area near the antenna. Hematoxylin-eosin staining of the dissected tissue revealed a correlation between the edge of the ablation zone and that of the hypoechoic area.</p></div><div><h3>Conclusions</h3><p>The ECS antenna can produce more spherical ablation zones with less charring and a clearer intraoperative ultrasound image of the ablation area than the CCS antenna.</p></div>\",\"PeriodicalId\":10624,\"journal\":{\"name\":\"Computer methods and programs in biomedicine\",\"volume\":\"257 \",\"pages\":\"Article 108383\"},\"PeriodicalIF\":4.9000,\"publicationDate\":\"2024-08-23\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Computer methods and programs in biomedicine\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S0169260724003766\",\"RegionNum\":2,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q1\",\"JCRName\":\"COMPUTER SCIENCE, INTERDISCIPLINARY APPLICATIONS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Computer methods and programs in biomedicine","FirstCategoryId":"5","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S0169260724003766","RegionNum":2,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q1","JCRName":"COMPUTER SCIENCE, INTERDISCIPLINARY APPLICATIONS","Score":null,"Total":0}
引用次数: 0
摘要
方法比较了两种类型的微波消融天线,一种是 ECS,另一种是传统冷却系统(CCS)。采用有限元法模拟体内消融。使用两种类型的天线在体外牛肝脏(n = 6)和体外猪肝脏(n = 16)中分别以 50、60 和 80 W 的功率创建 5、8 和 10 分钟的微波消融区,以 60 W 的功率创建 5 分钟的微波消融区。通过数值模拟和大体病理检查测量并比较了天线的过度处理比率、烧蚀纵横比、碳化面积和其他特征。结果在数值模拟中,ECS 天线的过度处理比低于 CCS 天线(50 W 5 分钟时为 1.38 比 1.43,50 W 8 分钟时为 1.19 比 1.35,50 W 10 分钟时为 1.13 比 1.32,60 W 5 分钟时为 1.28 比 1.38,60 W 8 分钟时为 1.14 比 1.32,60 W 10 分钟时为 1.10 比 1.30)。实验表明,ECS 天线产生的烧蚀区具有更显著的纵横比(50 W 5 分钟时为 0.92 ± 0.03 vs 0.72 ± 0.01,50 W 8 分钟时为 0.95 ± 0.02 vs 0.70 ± 0.01,60 W 10 分钟时为 0.92 ± 0.03 vs 0.72 ± 0.01)。01 (50 W 8 min), 0.96 ± 0.01 vs 0.71 ± 0.04 (50 W 10 min), 0.96 ± 0.01 vs 0.73 ± 0.02 (60 W 5 min), 0.94 ± 0.03 vs 0.71 ± 0.03 (60 W 8 min), 0.96 ± 0.02 vs 0.69 ± 0.04 (60 W 10 min)。04 在 60 W 10 分钟时),碳化面积较小(0.00 ± 0.00 cm2 vs 0.54 ± 0.06 cm2(50 W 5 分钟时),0.13 ± 0.03 cm2 vs 0.61 ± 0.09 cm2(50 W 8 分钟时),0.23 ± 0.05 cm2 vs 0.50 W 10 分钟时为 0.73 ± 0.05 m2,60 W 5 分钟时为 0.00 ± 0.00 cm2 vs 1.59 ± 0.41 cm2,60 W 8 分钟时为 0.23 ± 0.22 cm2 vs 2.11 ± 0.63 cm2,60 W 10 分钟时为 0.57 ± 0.09 cm2 vs 2.55 ± 0.51 cm2)。术中超声图像显示天线附近为低回声区,而非高回声区。结论与 CCS 天线相比,ECS 天线能产生更多的球形消融区,炭化更少,术中消融区的超声图像更清晰。
Importance of the enhanced cooling system for more spherical ablation zones: Numerical simulation, ex vivo and in vivo validation
Introduction
This study aimed to investigate the efficacy of a small-gauge microwave ablation antenna (MWA) with an enhanced cooling system (ECS) for generating more spherical ablation zones.
Methods
A comparison was made between two types of microwave ablation antennas, one with ECS and the other with a conventional cooling system (CCS). The finite element method was used to simulate in vivo ablation. Two types of antennas were used to create MWA zones for 5, 8, 10 min at 50, 60, and 80 W in ex vivo bovine livers (n = 6) and 5 min at 60 W in vivo porcine livers (n = 16). The overtreatment ratio, ablation aspect ratio, carbonization area, and other characteristcs of antennas were measured and compared using numerical simulation and gross pathologic examination.
Results
In numerical simulation, the ECS antenna demonstrated a lower overtreatment ratio than the CCS antenna (1.38 vs 1.43 at 50 W 5 min, 1.19 vs 1.35 at 50 W 8 min, 1.13 vs 1.32 at 50 W 10 min, 1.28 vs 1.38 at 60 W 5 min, 1.14 vs 1.32 at 60 W 8 min, 1.10 vs 1.30 at 60 W 10 min). The experiments revealed that the ECS antenna generated ablation zones with a more significant aspect ratio (0.92 ± 0.03 vs 0.72 ± 0.01 at 50 W 5 min, 0.95 ± 0.02 vs 0.70 ± 0.01 at 50 W 8 min, 0.96 ± 0.01 vs 0.71 ± 0.04 at 50 W 10 min, 0.96 ± 0.01 vs 0.73 ± 0.02 at 60 W 5 min, 0.94 ± 0.03 vs 0.71 ± 0.03 at 60 W 8 min, 0.96 ± 0.02 vs 0.69 ± 0.04 at 60 W 10 min) and a smaller carbonization area (0.00 ± 0.00 cm2 vs 0.54 ± 0.06 cm2 at 50 W 5 min, 0.13 ± 0.03 cm2 vs 0.61 ± 0.09 cm2 at 50 W 8 min, 0.23 ± 0.05 cm2 vs 0.73 ± 0.05 m2 at 50 W 10 min, 0.00 ± 0.00 cm2 vs 1.59 ± 0.41 cm2 at 60 W 5 min, 0.23 ± 0.22 cm2 vs 2.11 ± 0.63 cm2 at 60 W 8 min, 0.57 ± 0.09 cm2 vs 2.55 ± 0.51 cm2 at 60 W 10 min). Intraoperative ultrasound images revealed a hypoechoic area instead of a hyperechoic area near the antenna. Hematoxylin-eosin staining of the dissected tissue revealed a correlation between the edge of the ablation zone and that of the hypoechoic area.
Conclusions
The ECS antenna can produce more spherical ablation zones with less charring and a clearer intraoperative ultrasound image of the ablation area than the CCS antenna.
期刊介绍:
To encourage the development of formal computing methods, and their application in biomedical research and medical practice, by illustration of fundamental principles in biomedical informatics research; to stimulate basic research into application software design; to report the state of research of biomedical information processing projects; to report new computer methodologies applied in biomedical areas; the eventual distribution of demonstrable software to avoid duplication of effort; to provide a forum for discussion and improvement of existing software; to optimize contact between national organizations and regional user groups by promoting an international exchange of information on formal methods, standards and software in biomedicine.
Computer Methods and Programs in Biomedicine covers computing methodology and software systems derived from computing science for implementation in all aspects of biomedical research and medical practice. It is designed to serve: biochemists; biologists; geneticists; immunologists; neuroscientists; pharmacologists; toxicologists; clinicians; epidemiologists; psychiatrists; psychologists; cardiologists; chemists; (radio)physicists; computer scientists; programmers and systems analysts; biomedical, clinical, electrical and other engineers; teachers of medical informatics and users of educational software.