{"title":"使用光子计数 X 射线探测器进行单次曝光双能量胸部成像时的对比度和量子噪声。","authors":"Jeffrey Dhari, Jesse Tanguay","doi":"10.1088/1361-6560/ad75df","DOIUrl":null,"url":null,"abstract":"<p><p><i>Objective.</i>Photon-counting x-ray detectors (PCDs) can produce dual-energy (DE) x-ray images of lung cancer in a single x-ray exposure. It is important to understand the factors that affect contrast, noise and the contrast-to-noise ratio (CNR). This study quantifies the dependence of CNR on tube voltage, energy threshold and patient thickness in single exposure, DE, bone-suppressed thoracic imaging with PCDs, and elucidates how the fundamental processes inherent in x-ray detection by PCDs contribute to CNR degradation.<i>Approach.</i>We modeled the DE CNR for five theoretical PCDs, ranging from an ideal PCD that detects every primary photon in the correct energy bin while rejecting all scattered radiation to a non-ideal PCD that suffers from charge-sharing and electronic noise, and detects scatter. CNR was computed as a function of tube voltage and high energy threshold for average and larger-than-average patients. Model predictions were compared with experimental data extracted from images acquired using a cadmium telluride (CdTe) PCD with two energy bins and analog charge summing for charge-sharing suppression. The imaging phantom simulated attenuation, scatter and contrast in lung nodule imaging. We quantified CNR improvements achievable with anti-correlated noise reduction (ACNR) and measured the range of exposure rates over which pulse pile-up is negligible.<i>Main Results.</i>The realistic model predicted overall trends observed in the experimental data. CNR improvements with ACNR were approximately five-fold, and modeled CNR-enhancements were on average within 10% of experiment. CNR increased modestly (i.e.<20%) when increasing the tube voltage from 90 kV to 130 kV. Optimal energy thresholds ranged from 50 keV to 70 keV across all tube voltages and patient thicknesses with and without ACNR. Quantum efficiency, electronic noise, charge sharing and scatter degraded CNR by ~50%. Charge sharing and scatter had the largest effect on CNR, degrading it by ~30% and ~15% respectively. Dead-time losses were less than 5% for patient exposure rates within the range of clinical exposure rates.<i>Significance.</i>In this study, we (1) employed analytical and computational models to assess the impact of different factors on CNR in single-exposure DE imaging with PCDs, (2) evaluated the accuracy of these models in predicting experimental trends, (3) quantified improvements in CNR achievable through ACNR and (4) determined the range of patient exposure rates at which pulse pile-up can be considered negligible. To the best of our knowledge, this study represents the first systematic investigation of single-exposure DE imaging of lung nodules with PCDs.</p>","PeriodicalId":20185,"journal":{"name":"Physics in medicine and biology","volume":null,"pages":null},"PeriodicalIF":3.3000,"publicationDate":"2024-09-25","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Contrast and quantum noise in single-exposure dual-energy thoracic imaging with photon-counting x-ray detectors.\",\"authors\":\"Jeffrey Dhari, Jesse Tanguay\",\"doi\":\"10.1088/1361-6560/ad75df\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p><i>Objective.</i>Photon-counting x-ray detectors (PCDs) can produce dual-energy (DE) x-ray images of lung cancer in a single x-ray exposure. It is important to understand the factors that affect contrast, noise and the contrast-to-noise ratio (CNR). This study quantifies the dependence of CNR on tube voltage, energy threshold and patient thickness in single exposure, DE, bone-suppressed thoracic imaging with PCDs, and elucidates how the fundamental processes inherent in x-ray detection by PCDs contribute to CNR degradation.<i>Approach.</i>We modeled the DE CNR for five theoretical PCDs, ranging from an ideal PCD that detects every primary photon in the correct energy bin while rejecting all scattered radiation to a non-ideal PCD that suffers from charge-sharing and electronic noise, and detects scatter. CNR was computed as a function of tube voltage and high energy threshold for average and larger-than-average patients. Model predictions were compared with experimental data extracted from images acquired using a cadmium telluride (CdTe) PCD with two energy bins and analog charge summing for charge-sharing suppression. The imaging phantom simulated attenuation, scatter and contrast in lung nodule imaging. We quantified CNR improvements achievable with anti-correlated noise reduction (ACNR) and measured the range of exposure rates over which pulse pile-up is negligible.<i>Main Results.</i>The realistic model predicted overall trends observed in the experimental data. CNR improvements with ACNR were approximately five-fold, and modeled CNR-enhancements were on average within 10% of experiment. CNR increased modestly (i.e.<20%) when increasing the tube voltage from 90 kV to 130 kV. Optimal energy thresholds ranged from 50 keV to 70 keV across all tube voltages and patient thicknesses with and without ACNR. Quantum efficiency, electronic noise, charge sharing and scatter degraded CNR by ~50%. Charge sharing and scatter had the largest effect on CNR, degrading it by ~30% and ~15% respectively. Dead-time losses were less than 5% for patient exposure rates within the range of clinical exposure rates.<i>Significance.</i>In this study, we (1) employed analytical and computational models to assess the impact of different factors on CNR in single-exposure DE imaging with PCDs, (2) evaluated the accuracy of these models in predicting experimental trends, (3) quantified improvements in CNR achievable through ACNR and (4) determined the range of patient exposure rates at which pulse pile-up can be considered negligible. To the best of our knowledge, this study represents the first systematic investigation of single-exposure DE imaging of lung nodules with PCDs.</p>\",\"PeriodicalId\":20185,\"journal\":{\"name\":\"Physics in medicine and biology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":3.3000,\"publicationDate\":\"2024-09-25\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Physics in medicine and biology\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1088/1361-6560/ad75df\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ENGINEERING, BIOMEDICAL\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Physics in medicine and biology","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1361-6560/ad75df","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, BIOMEDICAL","Score":null,"Total":0}
引用次数: 0
摘要
目的:光子计数 X 射线探测器(PCD)可在一次 X 射线曝光中生成肺癌的双能量(DE)X 射线图像。本研究量化了使用 PCD 进行单次曝光、双能量、骨抑制胸部成像时对比度-噪声比(CNR)与管电压、能量阈值和患者厚度的关系,并阐明了 PCD 检测 X 射线的固有过程如何导致 CNR 下降:我们对五种理论 PCD 的 DE CNR 进行了建模,这些 PCD 既有在正确的能量区间检测到每个光子并拒绝散射的理想 PCD,也有受电荷共享和电子噪声影响并检测到散射的非理想 PCD。模型预测结果与使用碲化镉 PCD 采集的图像实验数据进行了比较。成像模型模拟了肺结节成像中的衰减、散射和对比度。我们量化了反相关降噪(AcNR)可实现的 CNR 改进,并测量了脉冲堆积可忽略不计的曝光率范围:在最佳能量阈值下,使用或不使用 ACNR 的模型 CNR 与实验 CNR 的差距在 10%以内。使用 ACNR 时,CNR 提高了约五倍。CNR 增加
Contrast and quantum noise in single-exposure dual-energy thoracic imaging with photon-counting x-ray detectors.
Objective.Photon-counting x-ray detectors (PCDs) can produce dual-energy (DE) x-ray images of lung cancer in a single x-ray exposure. It is important to understand the factors that affect contrast, noise and the contrast-to-noise ratio (CNR). This study quantifies the dependence of CNR on tube voltage, energy threshold and patient thickness in single exposure, DE, bone-suppressed thoracic imaging with PCDs, and elucidates how the fundamental processes inherent in x-ray detection by PCDs contribute to CNR degradation.Approach.We modeled the DE CNR for five theoretical PCDs, ranging from an ideal PCD that detects every primary photon in the correct energy bin while rejecting all scattered radiation to a non-ideal PCD that suffers from charge-sharing and electronic noise, and detects scatter. CNR was computed as a function of tube voltage and high energy threshold for average and larger-than-average patients. Model predictions were compared with experimental data extracted from images acquired using a cadmium telluride (CdTe) PCD with two energy bins and analog charge summing for charge-sharing suppression. The imaging phantom simulated attenuation, scatter and contrast in lung nodule imaging. We quantified CNR improvements achievable with anti-correlated noise reduction (ACNR) and measured the range of exposure rates over which pulse pile-up is negligible.Main Results.The realistic model predicted overall trends observed in the experimental data. CNR improvements with ACNR were approximately five-fold, and modeled CNR-enhancements were on average within 10% of experiment. CNR increased modestly (i.e.<20%) when increasing the tube voltage from 90 kV to 130 kV. Optimal energy thresholds ranged from 50 keV to 70 keV across all tube voltages and patient thicknesses with and without ACNR. Quantum efficiency, electronic noise, charge sharing and scatter degraded CNR by ~50%. Charge sharing and scatter had the largest effect on CNR, degrading it by ~30% and ~15% respectively. Dead-time losses were less than 5% for patient exposure rates within the range of clinical exposure rates.Significance.In this study, we (1) employed analytical and computational models to assess the impact of different factors on CNR in single-exposure DE imaging with PCDs, (2) evaluated the accuracy of these models in predicting experimental trends, (3) quantified improvements in CNR achievable through ACNR and (4) determined the range of patient exposure rates at which pulse pile-up can be considered negligible. To the best of our knowledge, this study represents the first systematic investigation of single-exposure DE imaging of lung nodules with PCDs.
期刊介绍:
The development and application of theoretical, computational and experimental physics to medicine, physiology and biology. Topics covered are: therapy physics (including ionizing and non-ionizing radiation); biomedical imaging (e.g. x-ray, magnetic resonance, ultrasound, optical and nuclear imaging); image-guided interventions; image reconstruction and analysis (including kinetic modelling); artificial intelligence in biomedical physics and analysis; nanoparticles in imaging and therapy; radiobiology; radiation protection and patient dose monitoring; radiation dosimetry