{"title":"通过全输出同步电击,使不断增大的电击线圈阻抗正常化。","authors":"Takahide Kadosaka, Masaya Watanabe, Motoki Nakao, Taro Koya, Taro Temma, Toshihisa Anzai","doi":"10.1111/jce.16416","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>Impedance is a crucial parameter in cardiovascular implantable electronic devices (CIEDs). Clinically, most CIEDs measure impedance using low voltage sub-threshold measurement (LVSM). Although the LVSM of shock impedance (LVSM-SI) is generally comparable with high voltage shock impedance (HVSI), LVSM-SI might be inaccurate if peri-lead tissue degeneration occurs.</p><p><strong>Methods and results: </strong>We present a case of elevated LVSM-SI occurring 8 years post-lead implantation, possibly attributed to encapsulation of the right ventricular lead coil. After 0.1 J shock was delivered, a full output synchronized shock was administered to measure HVSI, revealing a normal value. Furthermore, LVSM-SI was normalized and maintained within the normal range during long-term follow-up.</p><p><strong>Conclusion: </strong>Our findings suggest conducting a full-output synchronized shock test to assess HVSI when abnormal LVSM-SI is detected in the remote phase post-ICD implantation, which may be considered to help normalize LVSM shock impedance.</p>","PeriodicalId":15178,"journal":{"name":"Journal of Cardiovascular Electrophysiology","volume":null,"pages":null},"PeriodicalIF":2.3000,"publicationDate":"2024-08-21","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Normalization of increasing shocking coil impedance with full output synchronized shock.\",\"authors\":\"Takahide Kadosaka, Masaya Watanabe, Motoki Nakao, Taro Koya, Taro Temma, Toshihisa Anzai\",\"doi\":\"10.1111/jce.16416\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>Impedance is a crucial parameter in cardiovascular implantable electronic devices (CIEDs). Clinically, most CIEDs measure impedance using low voltage sub-threshold measurement (LVSM). Although the LVSM of shock impedance (LVSM-SI) is generally comparable with high voltage shock impedance (HVSI), LVSM-SI might be inaccurate if peri-lead tissue degeneration occurs.</p><p><strong>Methods and results: </strong>We present a case of elevated LVSM-SI occurring 8 years post-lead implantation, possibly attributed to encapsulation of the right ventricular lead coil. After 0.1 J shock was delivered, a full output synchronized shock was administered to measure HVSI, revealing a normal value. Furthermore, LVSM-SI was normalized and maintained within the normal range during long-term follow-up.</p><p><strong>Conclusion: </strong>Our findings suggest conducting a full-output synchronized shock test to assess HVSI when abnormal LVSM-SI is detected in the remote phase post-ICD implantation, which may be considered to help normalize LVSM shock impedance.</p>\",\"PeriodicalId\":15178,\"journal\":{\"name\":\"Journal of Cardiovascular Electrophysiology\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.3000,\"publicationDate\":\"2024-08-21\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Cardiovascular Electrophysiology\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1111/jce.16416\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Cardiovascular Electrophysiology","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1111/jce.16416","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
Normalization of increasing shocking coil impedance with full output synchronized shock.
Introduction: Impedance is a crucial parameter in cardiovascular implantable electronic devices (CIEDs). Clinically, most CIEDs measure impedance using low voltage sub-threshold measurement (LVSM). Although the LVSM of shock impedance (LVSM-SI) is generally comparable with high voltage shock impedance (HVSI), LVSM-SI might be inaccurate if peri-lead tissue degeneration occurs.
Methods and results: We present a case of elevated LVSM-SI occurring 8 years post-lead implantation, possibly attributed to encapsulation of the right ventricular lead coil. After 0.1 J shock was delivered, a full output synchronized shock was administered to measure HVSI, revealing a normal value. Furthermore, LVSM-SI was normalized and maintained within the normal range during long-term follow-up.
Conclusion: Our findings suggest conducting a full-output synchronized shock test to assess HVSI when abnormal LVSM-SI is detected in the remote phase post-ICD implantation, which may be considered to help normalize LVSM shock impedance.
期刊介绍:
Journal of Cardiovascular Electrophysiology (JCE) keeps its readership well informed of the latest developments in the study and management of arrhythmic disorders. Edited by Bradley P. Knight, M.D., and a distinguished international editorial board, JCE is the leading journal devoted to the study of the electrophysiology of the heart.