{"title":"高剂量率 192Ir 近距离放射治疗标准 BIPM.RI(I)-K8 与 VSL 和 BIPM 的参考空气灼烧率的关键比较","authors":"C Kessler, B J Jansen, J A de Pooter, P Aviles Lucas","doi":"10.1088/0026-1394/61/1a/06010","DOIUrl":null,"url":null,"abstract":"<title>Main text</title>A new key comparison of the standards for reference air kerma rate for <sup>192</sup>Ir high dose rate (HDR) brachytherapy sources of the Dutch Metrology Institute (VSL), The Netherlands, and the Bureau International des Poids et Mesures (BIPM) was carried out at the VSL in May 2019. The comparison result, based on the calibration coefficients for a transfer standard and expressed as a ratio of the VSL and the BIPM standards for reference air kerma rate, is 0.9926 with a combined standard uncertainty of 0.0055.To reach the main text of this paper, click on <ext-link xlink:href=\"https://www.bipm.org/documents/d/guest/bipm-ri-i-k8_vsl_2019\" xlink:type=\"simple\">Final Report</ext-link>. Note that this text is that which appears in Appendix B of the BIPM key comparison database <ext-link xlink:href=\"https://www.bipm.org/kcdb/\" xlink:type=\"simple\">https://www.bipm.org/kcdb/</ext-link>.The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).","PeriodicalId":18444,"journal":{"name":"Metrologia","volume":null,"pages":null},"PeriodicalIF":2.1000,"publicationDate":"2024-07-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Key comparison BIPM.RI(I)-K8 of high dose-rate 192Ir brachytherapy standards for reference air kerma rate of the VSL and the BIPM\",\"authors\":\"C Kessler, B J Jansen, J A de Pooter, P Aviles Lucas\",\"doi\":\"10.1088/0026-1394/61/1a/06010\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<title>Main text</title>A new key comparison of the standards for reference air kerma rate for <sup>192</sup>Ir high dose rate (HDR) brachytherapy sources of the Dutch Metrology Institute (VSL), The Netherlands, and the Bureau International des Poids et Mesures (BIPM) was carried out at the VSL in May 2019. The comparison result, based on the calibration coefficients for a transfer standard and expressed as a ratio of the VSL and the BIPM standards for reference air kerma rate, is 0.9926 with a combined standard uncertainty of 0.0055.To reach the main text of this paper, click on <ext-link xlink:href=\\\"https://www.bipm.org/documents/d/guest/bipm-ri-i-k8_vsl_2019\\\" xlink:type=\\\"simple\\\">Final Report</ext-link>. Note that this text is that which appears in Appendix B of the BIPM key comparison database <ext-link xlink:href=\\\"https://www.bipm.org/kcdb/\\\" xlink:type=\\\"simple\\\">https://www.bipm.org/kcdb/</ext-link>.The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).\",\"PeriodicalId\":18444,\"journal\":{\"name\":\"Metrologia\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.1000,\"publicationDate\":\"2024-07-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Metrologia\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1088/0026-1394/61/1a/06010\",\"RegionNum\":3,\"RegionCategory\":\"工程技术\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"INSTRUMENTS & INSTRUMENTATION\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Metrologia","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/0026-1394/61/1a/06010","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0

摘要

正文 2019年5月,荷兰计量研究院(VSL)和国际计量局(BIPM)对192Ir高剂量率(HDR)近距离放射源的参考空气开玛率标准进行了新的关键比对。比对结果以转移标准的校准系数为基础,以 VSL 和 BIPM 标准的参考空气开尔玛率之比表示,为 0.9926,综合标准不确定度为 0.0055。要阅读本文正文,请点击最终报告。请注意,此文本为 BIPM 关键比对数据库附录 B 中的文本 https://www.bipm.org/kcdb/.The 根据 CIPM 互认安排 (CIPM MRA) 的规定,最终报告已通过同行评审,并批准由 CCRI 出版。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Key comparison BIPM.RI(I)-K8 of high dose-rate 192Ir brachytherapy standards for reference air kerma rate of the VSL and the BIPM
Main textA new key comparison of the standards for reference air kerma rate for 192Ir high dose rate (HDR) brachytherapy sources of the Dutch Metrology Institute (VSL), The Netherlands, and the Bureau International des Poids et Mesures (BIPM) was carried out at the VSL in May 2019. The comparison result, based on the calibration coefficients for a transfer standard and expressed as a ratio of the VSL and the BIPM standards for reference air kerma rate, is 0.9926 with a combined standard uncertainty of 0.0055.To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database https://www.bipm.org/kcdb/.The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Metrologia
Metrologia 工程技术-物理:应用
CiteScore
2.80
自引率
25.00%
发文量
137
审稿时长
12 months
期刊介绍: Published 6 times per year, Metrologia covers the fundamentals of measurements, particularly those dealing with the seven base units of the International System of Units (metre, kilogram, second, ampere, kelvin, candela, mole) or proposals to replace them. The journal also publishes papers that contribute to the solution of difficult measurement problems and improve the accuracy of derived units and constants that are of fundamental importance to physics. In addition to regular papers, the journal publishes review articles, issues devoted to single topics of timely interest and occasional conference proceedings. Letters to the Editor and Short Communications (generally three pages or less) are also considered.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信