{"title":"血细胞比容水平对特定血糖仪准确性的影响:一项基于医院的研究。","authors":"Huan Nguyen Pham, Phuc Nguyen Huu Pham, Hang Thi Phan, Long Thang Cao, Hau Thi Thu Thoi, Tuyen Dang Thanh Do, Tuyet Thi Anh Truong","doi":"10.1111/jdi.14276","DOIUrl":null,"url":null,"abstract":"<div>\n \n \n <section>\n \n <h3> Aims/Introduction</h3>\n \n <p>Blood glucose meters are commonly used at the bedside, but most of the meters used in Hung Vuong Hospital (Ho Chi Minh City, Vietnam) are built for self-monitoring and might not be suitable for determining glucose levels in patients. In this study, we aimed to validate the performance of six frequently used meters in our hospital using the Clinical & Laboratory Standards Institute (CLSI) standard, and investigate the hematocrit impact on the accuracy of these meters.</p>\n </section>\n \n <section>\n \n <h3> Materials and Methods</h3>\n \n <p>A total of 135 pregnant women who underwent a 75-g oral glucose tolerance test consented to participate in the study at Hung Vuong Hospital. Whole blood glucose levels were measured in duplicate using meters, and hematocrit levels were measured using an Alinity h-series analyzer. Within 5 min, plasma glucose levels were measured twice in a row using the Cobas c502 reference analyzer. For accuracy and precision, the hematocrit effect was assed using CLSI POCT12-A3.</p>\n </section>\n \n <section>\n \n <h3> Results</h3>\n \n <p>Out of six evaluated meters, three meters qualified. For CLSI criterion at glucose concentration of 5.55 mmol/L, Accu-Chek Inform II, Accu-Chek Performa and OneTouch VerioVue achieved 97.31%, 98.08% and 99.62%, respectively. For CLSI criterion at 4.17 mmol/L, these three achieved 100%. Accu-Chek Inform II and Accu-Chek Performa showed an inverse correlation between glucose level and hematocrit with slopes of −0.500 (95% confidence interval −0.678 to −0.322) and −0.396 (95% confidence interval −0.569 to −0.224), whereas OneTouch VerioVue was not affected by hematocrit, with a slope of 0.207 (95% confidence interval −0.026 to 0.440).</p>\n </section>\n \n <section>\n \n <h3> Conclusions</h3>\n \n <p>Blood glucose meters' measurements can be affected by hematocrit, and might provide readings not within an acceptable bias. Medical organizations need to verify or validate before using on patients.</p>\n </section>\n </div>","PeriodicalId":51250,"journal":{"name":"Journal of Diabetes Investigation","volume":"15 10","pages":"1472-1482"},"PeriodicalIF":3.1000,"publicationDate":"2024-08-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11442750/pdf/","citationCount":"0","resultStr":"{\"title\":\"Impact of hematocrit levels on the accuracy of specific blood glucose meters: A hospital-based study\",\"authors\":\"Huan Nguyen Pham, Phuc Nguyen Huu Pham, Hang Thi Phan, Long Thang Cao, Hau Thi Thu Thoi, Tuyen Dang Thanh Do, Tuyet Thi Anh Truong\",\"doi\":\"10.1111/jdi.14276\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div>\\n \\n \\n <section>\\n \\n <h3> Aims/Introduction</h3>\\n \\n <p>Blood glucose meters are commonly used at the bedside, but most of the meters used in Hung Vuong Hospital (Ho Chi Minh City, Vietnam) are built for self-monitoring and might not be suitable for determining glucose levels in patients. In this study, we aimed to validate the performance of six frequently used meters in our hospital using the Clinical & Laboratory Standards Institute (CLSI) standard, and investigate the hematocrit impact on the accuracy of these meters.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Materials and Methods</h3>\\n \\n <p>A total of 135 pregnant women who underwent a 75-g oral glucose tolerance test consented to participate in the study at Hung Vuong Hospital. Whole blood glucose levels were measured in duplicate using meters, and hematocrit levels were measured using an Alinity h-series analyzer. Within 5 min, plasma glucose levels were measured twice in a row using the Cobas c502 reference analyzer. For accuracy and precision, the hematocrit effect was assed using CLSI POCT12-A3.</p>\\n </section>\\n \\n <section>\\n \\n <h3> Results</h3>\\n \\n <p>Out of six evaluated meters, three meters qualified. For CLSI criterion at glucose concentration of 5.55 mmol/L, Accu-Chek Inform II, Accu-Chek Performa and OneTouch VerioVue achieved 97.31%, 98.08% and 99.62%, respectively. For CLSI criterion at 4.17 mmol/L, these three achieved 100%. Accu-Chek Inform II and Accu-Chek Performa showed an inverse correlation between glucose level and hematocrit with slopes of −0.500 (95% confidence interval −0.678 to −0.322) and −0.396 (95% confidence interval −0.569 to −0.224), whereas OneTouch VerioVue was not affected by hematocrit, with a slope of 0.207 (95% confidence interval −0.026 to 0.440).</p>\\n </section>\\n \\n <section>\\n \\n <h3> Conclusions</h3>\\n \\n <p>Blood glucose meters' measurements can be affected by hematocrit, and might provide readings not within an acceptable bias. Medical organizations need to verify or validate before using on patients.</p>\\n </section>\\n </div>\",\"PeriodicalId\":51250,\"journal\":{\"name\":\"Journal of Diabetes Investigation\",\"volume\":\"15 10\",\"pages\":\"1472-1482\"},\"PeriodicalIF\":3.1000,\"publicationDate\":\"2024-08-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11442750/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Diabetes Investigation\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://onlinelibrary.wiley.com/doi/10.1111/jdi.14276\",\"RegionNum\":3,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ENDOCRINOLOGY & METABOLISM\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Diabetes Investigation","FirstCategoryId":"3","ListUrlMain":"https://onlinelibrary.wiley.com/doi/10.1111/jdi.14276","RegionNum":3,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENDOCRINOLOGY & METABOLISM","Score":null,"Total":0}
Impact of hematocrit levels on the accuracy of specific blood glucose meters: A hospital-based study
Aims/Introduction
Blood glucose meters are commonly used at the bedside, but most of the meters used in Hung Vuong Hospital (Ho Chi Minh City, Vietnam) are built for self-monitoring and might not be suitable for determining glucose levels in patients. In this study, we aimed to validate the performance of six frequently used meters in our hospital using the Clinical & Laboratory Standards Institute (CLSI) standard, and investigate the hematocrit impact on the accuracy of these meters.
Materials and Methods
A total of 135 pregnant women who underwent a 75-g oral glucose tolerance test consented to participate in the study at Hung Vuong Hospital. Whole blood glucose levels were measured in duplicate using meters, and hematocrit levels were measured using an Alinity h-series analyzer. Within 5 min, plasma glucose levels were measured twice in a row using the Cobas c502 reference analyzer. For accuracy and precision, the hematocrit effect was assed using CLSI POCT12-A3.
Results
Out of six evaluated meters, three meters qualified. For CLSI criterion at glucose concentration of 5.55 mmol/L, Accu-Chek Inform II, Accu-Chek Performa and OneTouch VerioVue achieved 97.31%, 98.08% and 99.62%, respectively. For CLSI criterion at 4.17 mmol/L, these three achieved 100%. Accu-Chek Inform II and Accu-Chek Performa showed an inverse correlation between glucose level and hematocrit with slopes of −0.500 (95% confidence interval −0.678 to −0.322) and −0.396 (95% confidence interval −0.569 to −0.224), whereas OneTouch VerioVue was not affected by hematocrit, with a slope of 0.207 (95% confidence interval −0.026 to 0.440).
Conclusions
Blood glucose meters' measurements can be affected by hematocrit, and might provide readings not within an acceptable bias. Medical organizations need to verify or validate before using on patients.
期刊介绍:
Journal of Diabetes Investigation is your core diabetes journal from Asia; the official journal of the Asian Association for the Study of Diabetes (AASD). The journal publishes original research, country reports, commentaries, reviews, mini-reviews, case reports, letters, as well as editorials and news. Embracing clinical and experimental research in diabetes and related areas, the Journal of Diabetes Investigation includes aspects of prevention, treatment, as well as molecular aspects and pathophysiology. Translational research focused on the exchange of ideas between clinicians and researchers is also welcome. Journal of Diabetes Investigation is indexed by Science Citation Index Expanded (SCIE).