125I 近距离放射粒子水中吸收剂量绝对测量室的开发和性能测试

Zhongbin Hang, Yuntao Liu, Mingzhe Song, Kexin Wei, Hongyu Wang, Qingran Li, Ning Ni, Xu Zhao, Tao Sun, Fei Gao
{"title":"125I 近距离放射粒子水中吸收剂量绝对测量室的开发和性能测试","authors":"Zhongbin Hang, Yuntao Liu, Mingzhe Song, Kexin Wei, Hongyu Wang, Qingran Li, Ning Ni, Xu Zhao, Tao Sun, Fei Gao","doi":"10.1088/1748-0221/19/07/p07026","DOIUrl":null,"url":null,"abstract":"\n The application of more and more low-energy, low-dose-rate\n photon radiation seeds in brachytherapy. According to the TG43 report\n and TG43-U1 report issued by the American Association of Physicists\n in Medicine, both apparent activity and air kerma strength are\n physical characterization parameters of brachytherapy seed\n strength,which are needed to convert into water absorbed dose at the\n depth of 1 cm. At present, the method of obtaining the water\n absorbed dose is to convert the air kerma strength, the uncertainty\n of results exceeds 5% (k = 1), which could reduce the cure rate\n of the treatment. In order to address this issue ,the China\n Institute of Atomic Energy has designed a device that can directly\n replicate the water absorbed dose of 125I brachytherapy\n seeds,which is an extrapolation chamber embedded with water\n equivalent material. In order to evaluate the performance of the\n extrapolation chamber, a series of experiments were carried out,\n including partial pressure test, leakage current, saturation curve,\n response linearity, collection area and zero point position. The\n test results show that the performance parameters of the\n extrapolation chamber meet the relevant requirements.","PeriodicalId":507814,"journal":{"name":"Journal of Instrumentation","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-07-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Development and performance testing of a chamber for absorbed dose in water absolute measurement for 125I brachytherapy seeds\",\"authors\":\"Zhongbin Hang, Yuntao Liu, Mingzhe Song, Kexin Wei, Hongyu Wang, Qingran Li, Ning Ni, Xu Zhao, Tao Sun, Fei Gao\",\"doi\":\"10.1088/1748-0221/19/07/p07026\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"\\n The application of more and more low-energy, low-dose-rate\\n photon radiation seeds in brachytherapy. According to the TG43 report\\n and TG43-U1 report issued by the American Association of Physicists\\n in Medicine, both apparent activity and air kerma strength are\\n physical characterization parameters of brachytherapy seed\\n strength,which are needed to convert into water absorbed dose at the\\n depth of 1 cm. At present, the method of obtaining the water\\n absorbed dose is to convert the air kerma strength, the uncertainty\\n of results exceeds 5% (k = 1), which could reduce the cure rate\\n of the treatment. In order to address this issue ,the China\\n Institute of Atomic Energy has designed a device that can directly\\n replicate the water absorbed dose of 125I brachytherapy\\n seeds,which is an extrapolation chamber embedded with water\\n equivalent material. In order to evaluate the performance of the\\n extrapolation chamber, a series of experiments were carried out,\\n including partial pressure test, leakage current, saturation curve,\\n response linearity, collection area and zero point position. The\\n test results show that the performance parameters of the\\n extrapolation chamber meet the relevant requirements.\",\"PeriodicalId\":507814,\"journal\":{\"name\":\"Journal of Instrumentation\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-07-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Instrumentation\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1088/1748-0221/19/07/p07026\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Instrumentation","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1088/1748-0221/19/07/p07026","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

近距离治疗中应用越来越多的低能量、低剂量率光子辐射粒子。根据美国医学物理学家协会(American Association of Physicists in Medicine)发布的 TG43 报告和 TG43-U1 报告,表观活度和空气热玛强度都是近距离放射粒子强度的物理表征参数,需要换算成 1 厘米深度的水吸收剂量。目前,获取水吸收剂量的方法是将空气热玛强度换算成水吸收剂量,其结果的不确定性超过 5%(k = 1),可能会降低治疗的治愈率。针对这一问题,中国原子能科学研究院设计了一种可直接复制 125I 近距离放射粒子水吸收剂量的装置,即嵌入水当量材料的外推室。为了评估外推室的性能,进行了一系列实验,包括分压测试、泄漏电流、饱和曲线、响应线性度、收集面积和零点位置。测试结果表明,外推室的性能参数符合相关要求。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Development and performance testing of a chamber for absorbed dose in water absolute measurement for 125I brachytherapy seeds
The application of more and more low-energy, low-dose-rate photon radiation seeds in brachytherapy. According to the TG43 report and TG43-U1 report issued by the American Association of Physicists in Medicine, both apparent activity and air kerma strength are physical characterization parameters of brachytherapy seed strength,which are needed to convert into water absorbed dose at the depth of 1 cm. At present, the method of obtaining the water absorbed dose is to convert the air kerma strength, the uncertainty of results exceeds 5% (k = 1), which could reduce the cure rate of the treatment. In order to address this issue ,the China Institute of Atomic Energy has designed a device that can directly replicate the water absorbed dose of 125I brachytherapy seeds,which is an extrapolation chamber embedded with water equivalent material. In order to evaluate the performance of the extrapolation chamber, a series of experiments were carried out, including partial pressure test, leakage current, saturation curve, response linearity, collection area and zero point position. The test results show that the performance parameters of the extrapolation chamber meet the relevant requirements.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信