{"title":"UmboMic: PVDF 悬臂麦克风 *","authors":"Aaron J Yeiser, Emma F Wawrzynek, John Z Zhang, Lukas Graf, Christopher I McHugh, Ioannis Kymissis, Elizabeth S Olson, Jeffrey H Lang and Hideko Heidi Nakajima","doi":"10.1088/1361-6439/ad5c6d","DOIUrl":null,"url":null,"abstract":"Objective. We present the ‘UmboMic,’ a prototype piezoelectric cantilever microphone designed for future use with totally-implantable cochlear implants. Methods. The UmboMic sensor is made from polyvinylidene difluoride (PVDF) because of its low Young’s modulus and biocompatibility. The sensor is designed to fit in the middle ear and measure the motion of the underside of the eardrum at the umbo. To maximize its performance, we developed a low noise charge amplifier in tandem with the UmboMic sensor. This paper presents the performance of the UmboMic sensor and amplifier in fresh cadaveric human temporal bones. Results. When tested in human temporal bones, the UmboMic apparatus achieves an equivalent input noise of 32.3 dB SPL over the frequency range 100 Hz–7 kHz, good linearity, and a flat frequency response to within 10 dB from about 100 Hz–6 kHz. Conclusion. These results demonstrate the feasibility of a PVDF-based microphone when paired with a low-noise amplifier. The reported UmboMic apparatus is comparable in performance to a conventional hearing aid microphone. Significance. The proof-of-concept UmboMic apparatus is a promising step towards creating a totally-implantable cochlear implant. A completely internal system would enhance the quality of life of cochlear implant users.","PeriodicalId":16346,"journal":{"name":"Journal of Micromechanics and Microengineering","volume":"47 1","pages":""},"PeriodicalIF":2.4000,"publicationDate":"2024-07-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"The UmboMic: a PVDF cantilever microphone *\",\"authors\":\"Aaron J Yeiser, Emma F Wawrzynek, John Z Zhang, Lukas Graf, Christopher I McHugh, Ioannis Kymissis, Elizabeth S Olson, Jeffrey H Lang and Hideko Heidi Nakajima\",\"doi\":\"10.1088/1361-6439/ad5c6d\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective. We present the ‘UmboMic,’ a prototype piezoelectric cantilever microphone designed for future use with totally-implantable cochlear implants. Methods. The UmboMic sensor is made from polyvinylidene difluoride (PVDF) because of its low Young’s modulus and biocompatibility. The sensor is designed to fit in the middle ear and measure the motion of the underside of the eardrum at the umbo. To maximize its performance, we developed a low noise charge amplifier in tandem with the UmboMic sensor. This paper presents the performance of the UmboMic sensor and amplifier in fresh cadaveric human temporal bones. Results. When tested in human temporal bones, the UmboMic apparatus achieves an equivalent input noise of 32.3 dB SPL over the frequency range 100 Hz–7 kHz, good linearity, and a flat frequency response to within 10 dB from about 100 Hz–6 kHz. Conclusion. These results demonstrate the feasibility of a PVDF-based microphone when paired with a low-noise amplifier. The reported UmboMic apparatus is comparable in performance to a conventional hearing aid microphone. Significance. The proof-of-concept UmboMic apparatus is a promising step towards creating a totally-implantable cochlear implant. A completely internal system would enhance the quality of life of cochlear implant users.\",\"PeriodicalId\":16346,\"journal\":{\"name\":\"Journal of Micromechanics and Microengineering\",\"volume\":\"47 1\",\"pages\":\"\"},\"PeriodicalIF\":2.4000,\"publicationDate\":\"2024-07-17\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Micromechanics and Microengineering\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1088/1361-6439/ad5c6d\",\"RegionNum\":4,\"RegionCategory\":\"工程技术\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"ENGINEERING, ELECTRICAL & ELECTRONIC\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Micromechanics and Microengineering","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1361-6439/ad5c6d","RegionNum":4,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"ENGINEERING, ELECTRICAL & ELECTRONIC","Score":null,"Total":0}
引用次数: 0

摘要

目的。我们展示的 "UmboMic "是一种压电悬臂麦克风原型,设计用于未来的全植入式人工耳蜗。方法。UmboMic 传感器由聚偏二氟乙烯(PVDF)制成,因为它的杨氏模量低且具有生物相容性。该传感器设计用于安装在中耳内,测量鼓膜下缘的运动。为了最大限度地提高其性能,我们开发了一种与 UmboMic 传感器配套使用的低噪声电荷放大器。本文介绍了 UmboMic 传感器和放大器在新鲜人体颞骨中的性能。结果。在人体颞骨中进行测试时,UmboMic 设备在 100 Hz-7 kHz 频率范围内实现了 32.3 dB SPL 的等效输入噪声、良好的线性度和 10 dB 以内的平坦频率响应(约 100 Hz-6 kHz)。结论这些结果证明了基于 PVDF 的麦克风与低噪声放大器搭配使用的可行性。所报告的 UmboMic 设备在性能上与传统助听器麦克风相当。意义重大。概念验证型 UmboMic 设备是向制造完全可植入式人工耳蜗迈出的充满希望的一步。一个完全内置的系统将提高人工耳蜗用户的生活质量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
The UmboMic: a PVDF cantilever microphone *
Objective. We present the ‘UmboMic,’ a prototype piezoelectric cantilever microphone designed for future use with totally-implantable cochlear implants. Methods. The UmboMic sensor is made from polyvinylidene difluoride (PVDF) because of its low Young’s modulus and biocompatibility. The sensor is designed to fit in the middle ear and measure the motion of the underside of the eardrum at the umbo. To maximize its performance, we developed a low noise charge amplifier in tandem with the UmboMic sensor. This paper presents the performance of the UmboMic sensor and amplifier in fresh cadaveric human temporal bones. Results. When tested in human temporal bones, the UmboMic apparatus achieves an equivalent input noise of 32.3 dB SPL over the frequency range 100 Hz–7 kHz, good linearity, and a flat frequency response to within 10 dB from about 100 Hz–6 kHz. Conclusion. These results demonstrate the feasibility of a PVDF-based microphone when paired with a low-noise amplifier. The reported UmboMic apparatus is comparable in performance to a conventional hearing aid microphone. Significance. The proof-of-concept UmboMic apparatus is a promising step towards creating a totally-implantable cochlear implant. A completely internal system would enhance the quality of life of cochlear implant users.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Micromechanics and Microengineering
Journal of Micromechanics and Microengineering 工程技术-材料科学:综合
CiteScore
4.50
自引率
4.30%
发文量
136
审稿时长
2.8 months
期刊介绍: Journal of Micromechanics and Microengineering (JMM) primarily covers experimental work, however relevant modelling papers are considered where supported by experimental data. The journal is focussed on all aspects of: -nano- and micro- mechanical systems -nano- and micro- electomechanical systems -nano- and micro- electrical and mechatronic systems -nano- and micro- engineering -nano- and micro- scale science Please note that we do not publish materials papers with no obvious application or link to nano- or micro-engineering. Below are some examples of the topics that are included within the scope of the journal: -MEMS and NEMS: Including sensors, optical MEMS/NEMS, RF MEMS/NEMS, etc. -Fabrication techniques and manufacturing: Including micromachining, etching, lithography, deposition, patterning, self-assembly, 3d printing, inkjet printing. -Packaging and Integration technologies. -Materials, testing, and reliability. -Micro- and nano-fluidics: Including optofluidics, acoustofluidics, droplets, microreactors, organ-on-a-chip. -Lab-on-a-chip and micro- and nano-total analysis systems. -Biomedical systems and devices: Including bio MEMS, biosensors, assays, organ-on-a-chip, drug delivery, cells, biointerfaces. -Energy and power: Including power MEMS/NEMS, energy harvesters, actuators, microbatteries. -Electronics: Including flexible electronics, wearable electronics, interface electronics. -Optical systems. -Robotics.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信