{"title":"氪离子激光发射波长下阱式探测器绝对光谱响应度老化评估","authors":"M. Dury, Subrena Harris, Teresa M Goodman, W. Servantes, Malcolm White","doi":"10.1088/1681-7575/ad615d","DOIUrl":null,"url":null,"abstract":"\n We have assessed the absolute spectral responsivities of three-element trap detectors calibrated against the NPL primary standard cryogenic radiometer at eight krypton ion laser emission wavelengths over an approximate period of twenty-years. As expected, we observed that the absolute spectral responsivities decrease with time. Additionally, we found that as the wavelength decreases, the ageing rate increases. At 356.4 nm, an average responsivity decrease of around -1800 ppm/year is reported. While at longer wavelengths, the slowest rate of ageing, approximately 40 ppm/year, occurs at 799.3 nm. Interestingly, the year-on-year changes exhibit a relatively linear trend, allowing us to predict the future performance of the trap detectors. It is important to note, however, that predicted performance cannot be guaranteed and regular recalibration against a primary standard is essential, especially if low calibration uncertainties are required.","PeriodicalId":18444,"journal":{"name":"Metrologia","volume":null,"pages":null},"PeriodicalIF":2.1000,"publicationDate":"2024-07-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"An Assessment of the Ageing of the Absolute Spectral Responsivity of Trap Detectors at Krypton Ion Laser Emission Wavelengths\",\"authors\":\"M. Dury, Subrena Harris, Teresa M Goodman, W. Servantes, Malcolm White\",\"doi\":\"10.1088/1681-7575/ad615d\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"\\n We have assessed the absolute spectral responsivities of three-element trap detectors calibrated against the NPL primary standard cryogenic radiometer at eight krypton ion laser emission wavelengths over an approximate period of twenty-years. As expected, we observed that the absolute spectral responsivities decrease with time. Additionally, we found that as the wavelength decreases, the ageing rate increases. At 356.4 nm, an average responsivity decrease of around -1800 ppm/year is reported. While at longer wavelengths, the slowest rate of ageing, approximately 40 ppm/year, occurs at 799.3 nm. Interestingly, the year-on-year changes exhibit a relatively linear trend, allowing us to predict the future performance of the trap detectors. It is important to note, however, that predicted performance cannot be guaranteed and regular recalibration against a primary standard is essential, especially if low calibration uncertainties are required.\",\"PeriodicalId\":18444,\"journal\":{\"name\":\"Metrologia\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":2.1000,\"publicationDate\":\"2024-07-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Metrologia\",\"FirstCategoryId\":\"5\",\"ListUrlMain\":\"https://doi.org/10.1088/1681-7575/ad615d\",\"RegionNum\":3,\"RegionCategory\":\"工程技术\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"INSTRUMENTS & INSTRUMENTATION\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Metrologia","FirstCategoryId":"5","ListUrlMain":"https://doi.org/10.1088/1681-7575/ad615d","RegionNum":3,"RegionCategory":"工程技术","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"INSTRUMENTS & INSTRUMENTATION","Score":null,"Total":0}
引用次数: 0

摘要

我们评估了三元素阱探测器的绝对光谱响应度,这些探测器是根据 NPL 初级标准低温辐射计在八个氪离子激光发射波长下进行校准的,时间跨度大约为二十年。不出所料,我们观察到绝对光谱响应度随着时间的推移而降低。此外,我们还发现随着波长的减少,老化率也会增加。据报告,在 356.4 纳米波长处,平均响应度下降约为-1800 ppm/年。而在波长较长的 799.3 纳米波段,老化速度最慢,约为每年 40 ppm。有趣的是,每年的变化呈现出相对线性的趋势,使我们能够预测捕获器探测器未来的性能。但需要注意的是,预测的性能是无法保证的,因此必须根据主要标准定期重新校准,尤其是在要求校准不确定性较低的情况下。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
An Assessment of the Ageing of the Absolute Spectral Responsivity of Trap Detectors at Krypton Ion Laser Emission Wavelengths
We have assessed the absolute spectral responsivities of three-element trap detectors calibrated against the NPL primary standard cryogenic radiometer at eight krypton ion laser emission wavelengths over an approximate period of twenty-years. As expected, we observed that the absolute spectral responsivities decrease with time. Additionally, we found that as the wavelength decreases, the ageing rate increases. At 356.4 nm, an average responsivity decrease of around -1800 ppm/year is reported. While at longer wavelengths, the slowest rate of ageing, approximately 40 ppm/year, occurs at 799.3 nm. Interestingly, the year-on-year changes exhibit a relatively linear trend, allowing us to predict the future performance of the trap detectors. It is important to note, however, that predicted performance cannot be guaranteed and regular recalibration against a primary standard is essential, especially if low calibration uncertainties are required.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Metrologia
Metrologia 工程技术-物理:应用
CiteScore
2.80
自引率
25.00%
发文量
137
审稿时长
12 months
期刊介绍: Published 6 times per year, Metrologia covers the fundamentals of measurements, particularly those dealing with the seven base units of the International System of Units (metre, kilogram, second, ampere, kelvin, candela, mole) or proposals to replace them. The journal also publishes papers that contribute to the solution of difficult measurement problems and improve the accuracy of derived units and constants that are of fundamental importance to physics. In addition to regular papers, the journal publishes review articles, issues devoted to single topics of timely interest and occasional conference proceedings. Letters to the Editor and Short Communications (generally three pages or less) are also considered.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信