[便携式超声诊断设备与 CBCT 在盆腔前列腺放射治疗中的膀胱容积测量精度比较]。

Nihon Hoshasen Gijutsu Gakkai zasshi Pub Date : 2024-07-20 Epub Date: 2024-06-26 DOI:10.6009/jjrt.2024-1461
Takahiro Saika, Kenji Matsumoto
{"title":"[便携式超声诊断设备与 CBCT 在盆腔前列腺放射治疗中的膀胱容积测量精度比较]。","authors":"Takahiro Saika, Kenji Matsumoto","doi":"10.6009/jjrt.2024-1461","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>We evaluated the measurement accuracy of the CubeScan BioCon-900 (here after BioCon-900), a portable ultrasound imaging diagnostic device capable of measuring bladder volume, to determine if it can accurately assess bladder volume before intensity-modulated radiation therapy (IMRT) for prostate cancer.</p><p><strong>Methods: </strong>Bladder volume was measured from kV-Cone Beam computed tomography (CBCT) images obtained immediately before the administration of IMRT for prostate cancer using Halcyon. The bladder volume measured from kV-CBCT images (23 patients, total number of scans: 139) immediately before the IMRT procedure was used as the reference value. The difference between the bladder volume measured by the BioCon-900 was then calculated.</p><p><strong>Results: </strong>The bladder volume measured from kV-CBCT images was 117.5±49.4 cc. In contrast, the bladder volume obtained with BioCon-900 was 104.1±48.6 ml, resulting in an absolute mean difference of 18.4% and a correlation coefficient of 0.881. The measurements by BioCon-900 tended to be, on average, 11% smaller than the bladder volume measured by kV-CBCT imaging.</p><p><strong>Conclusion: </strong>kV-CBCT images demonstrated a strong positive correlation between bladder volume and bladder urine output obtained with BioCon-900. BioCon-900 can be used before kV-CBCT imaging to accurately and conveniently assess bladder volume.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":" ","pages":"760-765"},"PeriodicalIF":0.0000,"publicationDate":"2024-07-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Comparison of Bladder Volume Measurement Accuracy between Portable Ultrasound Diagnostic Equipment and CBCT for Pelvic Prostate Radiotherapy].\",\"authors\":\"Takahiro Saika, Kenji Matsumoto\",\"doi\":\"10.6009/jjrt.2024-1461\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>We evaluated the measurement accuracy of the CubeScan BioCon-900 (here after BioCon-900), a portable ultrasound imaging diagnostic device capable of measuring bladder volume, to determine if it can accurately assess bladder volume before intensity-modulated radiation therapy (IMRT) for prostate cancer.</p><p><strong>Methods: </strong>Bladder volume was measured from kV-Cone Beam computed tomography (CBCT) images obtained immediately before the administration of IMRT for prostate cancer using Halcyon. The bladder volume measured from kV-CBCT images (23 patients, total number of scans: 139) immediately before the IMRT procedure was used as the reference value. The difference between the bladder volume measured by the BioCon-900 was then calculated.</p><p><strong>Results: </strong>The bladder volume measured from kV-CBCT images was 117.5±49.4 cc. In contrast, the bladder volume obtained with BioCon-900 was 104.1±48.6 ml, resulting in an absolute mean difference of 18.4% and a correlation coefficient of 0.881. The measurements by BioCon-900 tended to be, on average, 11% smaller than the bladder volume measured by kV-CBCT imaging.</p><p><strong>Conclusion: </strong>kV-CBCT images demonstrated a strong positive correlation between bladder volume and bladder urine output obtained with BioCon-900. BioCon-900 can be used before kV-CBCT imaging to accurately and conveniently assess bladder volume.</p>\",\"PeriodicalId\":74309,\"journal\":{\"name\":\"Nihon Hoshasen Gijutsu Gakkai zasshi\",\"volume\":\" \",\"pages\":\"760-765\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-07-20\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nihon Hoshasen Gijutsu Gakkai zasshi\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.6009/jjrt.2024-1461\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/6/26 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.2024-1461","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/26 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:我们评估了可测量膀胱容积的便携式超声成像诊断设备 CubeScan BioCon-900(以下简称 BioCon-900)的测量准确性,以确定它能否在前列腺癌调强放射治疗(IMRT)前准确评估膀胱容积:在使用 Halcyon 进行前列腺癌 IMRT 治疗前,通过 kV-Cone Beam 计算机断层扫描 (CBCT) 图像测量膀胱容积。将 IMRT 术前通过 kV-CBCT 图像测量的膀胱容积(23 名患者,扫描总数:139)作为参考值。然后计算 BioCon-900 测量的膀胱容积之间的差异:结果:通过 kV-CBCT 图像测量的膀胱容积为 117.5±49.4 cc。而 BioCon-900 测得的膀胱容积为 104.1±48.6 毫升,绝对平均值相差 18.4%,相关系数为 0.881。结论:KV-CBCT 图像显示,BioCon-900 测得的膀胱容量与膀胱尿量之间存在很强的正相关性。BioCon-900 可在 kV-CBCT 成像前使用,以准确、方便地评估膀胱容量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
[Comparison of Bladder Volume Measurement Accuracy between Portable Ultrasound Diagnostic Equipment and CBCT for Pelvic Prostate Radiotherapy].

Purpose: We evaluated the measurement accuracy of the CubeScan BioCon-900 (here after BioCon-900), a portable ultrasound imaging diagnostic device capable of measuring bladder volume, to determine if it can accurately assess bladder volume before intensity-modulated radiation therapy (IMRT) for prostate cancer.

Methods: Bladder volume was measured from kV-Cone Beam computed tomography (CBCT) images obtained immediately before the administration of IMRT for prostate cancer using Halcyon. The bladder volume measured from kV-CBCT images (23 patients, total number of scans: 139) immediately before the IMRT procedure was used as the reference value. The difference between the bladder volume measured by the BioCon-900 was then calculated.

Results: The bladder volume measured from kV-CBCT images was 117.5±49.4 cc. In contrast, the bladder volume obtained with BioCon-900 was 104.1±48.6 ml, resulting in an absolute mean difference of 18.4% and a correlation coefficient of 0.881. The measurements by BioCon-900 tended to be, on average, 11% smaller than the bladder volume measured by kV-CBCT imaging.

Conclusion: kV-CBCT images demonstrated a strong positive correlation between bladder volume and bladder urine output obtained with BioCon-900. BioCon-900 can be used before kV-CBCT imaging to accurately and conveniently assess bladder volume.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信