{"title":"[便携式超声诊断设备与 CBCT 在盆腔前列腺放射治疗中的膀胱容积测量精度比较]。","authors":"Takahiro Saika, Kenji Matsumoto","doi":"10.6009/jjrt.2024-1461","DOIUrl":null,"url":null,"abstract":"<p><strong>Purpose: </strong>We evaluated the measurement accuracy of the CubeScan BioCon-900 (here after BioCon-900), a portable ultrasound imaging diagnostic device capable of measuring bladder volume, to determine if it can accurately assess bladder volume before intensity-modulated radiation therapy (IMRT) for prostate cancer.</p><p><strong>Methods: </strong>Bladder volume was measured from kV-Cone Beam computed tomography (CBCT) images obtained immediately before the administration of IMRT for prostate cancer using Halcyon. The bladder volume measured from kV-CBCT images (23 patients, total number of scans: 139) immediately before the IMRT procedure was used as the reference value. The difference between the bladder volume measured by the BioCon-900 was then calculated.</p><p><strong>Results: </strong>The bladder volume measured from kV-CBCT images was 117.5±49.4 cc. In contrast, the bladder volume obtained with BioCon-900 was 104.1±48.6 ml, resulting in an absolute mean difference of 18.4% and a correlation coefficient of 0.881. The measurements by BioCon-900 tended to be, on average, 11% smaller than the bladder volume measured by kV-CBCT imaging.</p><p><strong>Conclusion: </strong>kV-CBCT images demonstrated a strong positive correlation between bladder volume and bladder urine output obtained with BioCon-900. BioCon-900 can be used before kV-CBCT imaging to accurately and conveniently assess bladder volume.</p>","PeriodicalId":74309,"journal":{"name":"Nihon Hoshasen Gijutsu Gakkai zasshi","volume":" ","pages":"760-765"},"PeriodicalIF":0.0000,"publicationDate":"2024-07-20","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Comparison of Bladder Volume Measurement Accuracy between Portable Ultrasound Diagnostic Equipment and CBCT for Pelvic Prostate Radiotherapy].\",\"authors\":\"Takahiro Saika, Kenji Matsumoto\",\"doi\":\"10.6009/jjrt.2024-1461\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Purpose: </strong>We evaluated the measurement accuracy of the CubeScan BioCon-900 (here after BioCon-900), a portable ultrasound imaging diagnostic device capable of measuring bladder volume, to determine if it can accurately assess bladder volume before intensity-modulated radiation therapy (IMRT) for prostate cancer.</p><p><strong>Methods: </strong>Bladder volume was measured from kV-Cone Beam computed tomography (CBCT) images obtained immediately before the administration of IMRT for prostate cancer using Halcyon. The bladder volume measured from kV-CBCT images (23 patients, total number of scans: 139) immediately before the IMRT procedure was used as the reference value. The difference between the bladder volume measured by the BioCon-900 was then calculated.</p><p><strong>Results: </strong>The bladder volume measured from kV-CBCT images was 117.5±49.4 cc. In contrast, the bladder volume obtained with BioCon-900 was 104.1±48.6 ml, resulting in an absolute mean difference of 18.4% and a correlation coefficient of 0.881. The measurements by BioCon-900 tended to be, on average, 11% smaller than the bladder volume measured by kV-CBCT imaging.</p><p><strong>Conclusion: </strong>kV-CBCT images demonstrated a strong positive correlation between bladder volume and bladder urine output obtained with BioCon-900. BioCon-900 can be used before kV-CBCT imaging to accurately and conveniently assess bladder volume.</p>\",\"PeriodicalId\":74309,\"journal\":{\"name\":\"Nihon Hoshasen Gijutsu Gakkai zasshi\",\"volume\":\" \",\"pages\":\"760-765\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-07-20\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nihon Hoshasen Gijutsu Gakkai zasshi\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.6009/jjrt.2024-1461\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/6/26 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nihon Hoshasen Gijutsu Gakkai zasshi","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.6009/jjrt.2024-1461","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/26 0:00:00","PubModel":"Epub","JCR":"","JCRName":"","Score":null,"Total":0}
[Comparison of Bladder Volume Measurement Accuracy between Portable Ultrasound Diagnostic Equipment and CBCT for Pelvic Prostate Radiotherapy].
Purpose: We evaluated the measurement accuracy of the CubeScan BioCon-900 (here after BioCon-900), a portable ultrasound imaging diagnostic device capable of measuring bladder volume, to determine if it can accurately assess bladder volume before intensity-modulated radiation therapy (IMRT) for prostate cancer.
Methods: Bladder volume was measured from kV-Cone Beam computed tomography (CBCT) images obtained immediately before the administration of IMRT for prostate cancer using Halcyon. The bladder volume measured from kV-CBCT images (23 patients, total number of scans: 139) immediately before the IMRT procedure was used as the reference value. The difference between the bladder volume measured by the BioCon-900 was then calculated.
Results: The bladder volume measured from kV-CBCT images was 117.5±49.4 cc. In contrast, the bladder volume obtained with BioCon-900 was 104.1±48.6 ml, resulting in an absolute mean difference of 18.4% and a correlation coefficient of 0.881. The measurements by BioCon-900 tended to be, on average, 11% smaller than the bladder volume measured by kV-CBCT imaging.
Conclusion: kV-CBCT images demonstrated a strong positive correlation between bladder volume and bladder urine output obtained with BioCon-900. BioCon-900 can be used before kV-CBCT imaging to accurately and conveniently assess bladder volume.