{"title":"四种酶联免疫吸附试剂盒测量血清内皮素-1的准确性值得怀疑","authors":"Minqi Cai, Qian Zheng, Yiqiang Chen, Siyuan Liu, Huimin Zhu, Bing Bai","doi":"10.1515/labmed-2024-0028","DOIUrl":null,"url":null,"abstract":"Abstract Objectives Altered serum Netrin-1 levels have been widely reported in cancer and other clinical diseases and they are often measured by commercial ELISA kits. However, we found the questionable results using these kits and therefore performed this simple study to evaluate their accuracy in detection of serum Netrin-1. Methods Four commonly used commercial kits were collected. The kit standards were serially diluted or spiked into serum samples. The cells with confirmed expression of Netrin-1 and their culture medium, as well as the Netrin-1 controls of each kit were used for the kits to detect. The cell lysate samples and the kit controls were also blotted on a nitrocellulose membrane for detection antibodies of each kit to probe. Results Detection of the Netrin-1 standards in serum by each kit were all affected. Only one kit was able to detect Netrin-1 in the cell lysate or medium. No ELISA kits could detect all Netrin-1 controls of the four kits. None of the detection antibodies correctly probed Netrin-1 in the dot blot. Conclusions The accuracy of these four Netrin-1 ELISA kits is under question. Reported serum Netrin-1 levels based on measurements by these kits need be carefully interpreted.","PeriodicalId":55986,"journal":{"name":"Journal of Laboratory Medicine","volume":null,"pages":null},"PeriodicalIF":1.1000,"publicationDate":"2024-06-10","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Questionable accuracy of four ELISA kits in serum Netrin-1 measurement\",\"authors\":\"Minqi Cai, Qian Zheng, Yiqiang Chen, Siyuan Liu, Huimin Zhu, Bing Bai\",\"doi\":\"10.1515/labmed-2024-0028\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract Objectives Altered serum Netrin-1 levels have been widely reported in cancer and other clinical diseases and they are often measured by commercial ELISA kits. However, we found the questionable results using these kits and therefore performed this simple study to evaluate their accuracy in detection of serum Netrin-1. Methods Four commonly used commercial kits were collected. The kit standards were serially diluted or spiked into serum samples. The cells with confirmed expression of Netrin-1 and their culture medium, as well as the Netrin-1 controls of each kit were used for the kits to detect. The cell lysate samples and the kit controls were also blotted on a nitrocellulose membrane for detection antibodies of each kit to probe. Results Detection of the Netrin-1 standards in serum by each kit were all affected. Only one kit was able to detect Netrin-1 in the cell lysate or medium. No ELISA kits could detect all Netrin-1 controls of the four kits. None of the detection antibodies correctly probed Netrin-1 in the dot blot. Conclusions The accuracy of these four Netrin-1 ELISA kits is under question. Reported serum Netrin-1 levels based on measurements by these kits need be carefully interpreted.\",\"PeriodicalId\":55986,\"journal\":{\"name\":\"Journal of Laboratory Medicine\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.1000,\"publicationDate\":\"2024-06-10\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Laboratory Medicine\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1515/labmed-2024-0028\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"MEDICAL LABORATORY TECHNOLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Laboratory Medicine","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1515/labmed-2024-0028","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"MEDICAL LABORATORY TECHNOLOGY","Score":null,"Total":0}
引用次数: 0

摘要

摘要 目的 癌症和其他临床疾病中血清 Netrin-1 水平的变化已被广泛报道,通常使用商业 ELISA 试剂盒进行测量。然而,我们发现使用这些试剂盒的结果值得商榷,因此我们进行了这项简单的研究,以评估这些试剂盒检测血清 Netrin-1 的准确性。方法 收集了四种常用的商业试剂盒。将试剂盒标准品系列稀释或添加到血清样本中。用已确认表达 Netrin-1 的细胞及其培养基以及每种试剂盒的 Netrin-1 对照组进行检测。细胞裂解物样本和试剂盒对照还被印在硝酸纤维素膜上,供各试剂盒的检测抗体探针使用。结果 各试剂盒对血清中 Netrin-1 标准品的检测均受到影响。只有一种试剂盒能检测出细胞裂解液或培养基中的 Netrin-1。四种试剂盒中没有一种 ELISA 试剂盒能检测出所有 Netrin-1 对照品。没有一种检测抗体能在点印迹中正确探测到 Netrin-1。结论 这四种 Netrin-1 酶联免疫吸附试剂盒的准确性受到质疑。根据这些试剂盒的测量结果报告的血清 Netrin-1 水平需要仔细解读。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Questionable accuracy of four ELISA kits in serum Netrin-1 measurement
Abstract Objectives Altered serum Netrin-1 levels have been widely reported in cancer and other clinical diseases and they are often measured by commercial ELISA kits. However, we found the questionable results using these kits and therefore performed this simple study to evaluate their accuracy in detection of serum Netrin-1. Methods Four commonly used commercial kits were collected. The kit standards were serially diluted or spiked into serum samples. The cells with confirmed expression of Netrin-1 and their culture medium, as well as the Netrin-1 controls of each kit were used for the kits to detect. The cell lysate samples and the kit controls were also blotted on a nitrocellulose membrane for detection antibodies of each kit to probe. Results Detection of the Netrin-1 standards in serum by each kit were all affected. Only one kit was able to detect Netrin-1 in the cell lysate or medium. No ELISA kits could detect all Netrin-1 controls of the four kits. None of the detection antibodies correctly probed Netrin-1 in the dot blot. Conclusions The accuracy of these four Netrin-1 ELISA kits is under question. Reported serum Netrin-1 levels based on measurements by these kits need be carefully interpreted.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Laboratory Medicine
Journal of Laboratory Medicine Mathematics-Discrete Mathematics and Combinatorics
CiteScore
2.50
自引率
0.00%
发文量
39
审稿时长
10 weeks
期刊介绍: The Journal of Laboratory Medicine (JLM) is a bi-monthly published journal that reports on the latest developments in laboratory medicine. Particular focus is placed on the diagnostic aspects of the clinical laboratory, although technical, regulatory, and educational topics are equally covered. The Journal specializes in the publication of high-standard, competent and timely review articles on clinical, methodological and pathogenic aspects of modern laboratory diagnostics. These reviews are critically reviewed by expert reviewers and JLM’s Associate Editors who are specialists in the various subdisciplines of laboratory medicine. In addition, JLM publishes original research articles, case reports, point/counterpoint articles and letters to the editor, all of which are peer reviewed by at least two experts in the field.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信