{"title":"[数字振镜扫描系统在二氧化碳点阵激光安全改进中的应用]。","authors":"Ming Li, Wei Feng, Songhua Liu","doi":"10.12455/j.issn.1671-7104.230568","DOIUrl":null,"url":null,"abstract":"<p><p>An analog PID controller-based galvanometer scanner is widely used by fractional laser medical equipment (FLME) to scan lasers across tissue surfaces, achieving the desired therapeutic effect. This type of driver, primarily composed of passive components and operational amplifiers, can only accept commands from the central controller of the FLME, with a simple hardware circuit-based fault diagnosis; thus, the safety of the FLME is compromised. To address these issues, the failure mechanisms of galvanometers and their impact on the safety of FLME are thoroughly analyzed first. Then, an adaptive limit protection method, a coil open circuit fault diagnosis, a communication timeout protection based on two handshakes, and a galvanometer control timeout protection are proposed, respectively, based on a digital driver platform, to supplement the deficiencies in the original fault diagnosis and protection system. This ensures the safety of the FLME. Finally, the effectiveness of the proposed strategies is validated through experiments.</p>","PeriodicalId":52535,"journal":{"name":"中国医疗器械杂志","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"2024-05-30","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"[Application of Digital Galvanometer Scanner System for CO<sub>2</sub> Fractional Laser Safety Improvement].\",\"authors\":\"Ming Li, Wei Feng, Songhua Liu\",\"doi\":\"10.12455/j.issn.1671-7104.230568\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>An analog PID controller-based galvanometer scanner is widely used by fractional laser medical equipment (FLME) to scan lasers across tissue surfaces, achieving the desired therapeutic effect. This type of driver, primarily composed of passive components and operational amplifiers, can only accept commands from the central controller of the FLME, with a simple hardware circuit-based fault diagnosis; thus, the safety of the FLME is compromised. To address these issues, the failure mechanisms of galvanometers and their impact on the safety of FLME are thoroughly analyzed first. Then, an adaptive limit protection method, a coil open circuit fault diagnosis, a communication timeout protection based on two handshakes, and a galvanometer control timeout protection are proposed, respectively, based on a digital driver platform, to supplement the deficiencies in the original fault diagnosis and protection system. This ensures the safety of the FLME. Finally, the effectiveness of the proposed strategies is validated through experiments.</p>\",\"PeriodicalId\":52535,\"journal\":{\"name\":\"中国医疗器械杂志\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-05-30\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"中国医疗器械杂志\",\"FirstCategoryId\":\"1087\",\"ListUrlMain\":\"https://doi.org/10.12455/j.issn.1671-7104.230568\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"中国医疗器械杂志","FirstCategoryId":"1087","ListUrlMain":"https://doi.org/10.12455/j.issn.1671-7104.230568","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
[Application of Digital Galvanometer Scanner System for CO2 Fractional Laser Safety Improvement].
An analog PID controller-based galvanometer scanner is widely used by fractional laser medical equipment (FLME) to scan lasers across tissue surfaces, achieving the desired therapeutic effect. This type of driver, primarily composed of passive components and operational amplifiers, can only accept commands from the central controller of the FLME, with a simple hardware circuit-based fault diagnosis; thus, the safety of the FLME is compromised. To address these issues, the failure mechanisms of galvanometers and their impact on the safety of FLME are thoroughly analyzed first. Then, an adaptive limit protection method, a coil open circuit fault diagnosis, a communication timeout protection based on two handshakes, and a galvanometer control timeout protection are proposed, respectively, based on a digital driver platform, to supplement the deficiencies in the original fault diagnosis and protection system. This ensures the safety of the FLME. Finally, the effectiveness of the proposed strategies is validated through experiments.