Magnum 金属非骨水泥全髋关节置换术:211 例病例的 8 至 18 年疗效。

Q1 Medicine
MUSCULOSKELETAL SURGERY Pub Date : 2024-12-01 Epub Date: 2024-06-04 DOI:10.1007/s12306-024-00831-3
D Gaillard-Campbell, T Gross
{"title":"Magnum 金属非骨水泥全髋关节置换术:211 例病例的 8 至 18 年疗效。","authors":"D Gaillard-Campbell, T Gross","doi":"10.1007/s12306-024-00831-3","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>Reports of adverse reactions to metal debris contributed in part to a decline in use of large-bearing metal-on-metal total hip devices. We hypothesize an optimal trunnion design may reduce risk of this failure mode in large-bearing total hip arthroplasty systems. The purpose of this study is to report mid- to long-term outcomes for a single-surgeon series of 211 total hip arthroplasties using the large-bearing Biomet Magnum metal-on-metal system.</p><p><strong>Materials and methods: </strong>Between December 2004 and January 2016, the primary surgeon performed 211 uncemented Magnum total hip arthroplasties in 181 patients. The average length of follow-up was 10.1 ± 3.5 years (range 8-18 years).</p><p><strong>Results: </strong>Using failure of any component as the endpoint, the overall survivorship rate was 98.1% at 10 years and 97.4% at 18 years. These eight failures (3.8% of cohort) included one case of adverse wear-related failure (0.5%), two cases of acetabular ingrowth failure (0.9%), three cases of trunnion corrosion (1.4%), one failure of late infection (0.5%), and one inappropriate revision of components for trochanteric nonunion without instability (0.5%). Excluding failed cases, all components were radiographically stable with no radiolucencies. Except for the one wear failure, ion testing revealed that 97.2% of cases were within optimal whole blood metal ion levels with the remaining ion test results within acceptable levels.</p><p><strong>Conclusions: </strong>With the uncemented Magnum metal-on-metal total hip, we achieved 97.4% 18-year implant survivorship, exceeding the NICE criteria and registry benchmarks for implant survivorship. We observed a trunnion corrosion rate of 1.4% and no cases of instability. The single case of adverse wear-related failure was caused by acetabular component malposition.</p>","PeriodicalId":18875,"journal":{"name":"MUSCULOSKELETAL SURGERY","volume":" ","pages":"449-457"},"PeriodicalIF":0.0000,"publicationDate":"2024-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Magnum metal-on-metal uncemented total hip replacement: 8- to 18-year outcomes of 211 cases.\",\"authors\":\"D Gaillard-Campbell, T Gross\",\"doi\":\"10.1007/s12306-024-00831-3\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>Reports of adverse reactions to metal debris contributed in part to a decline in use of large-bearing metal-on-metal total hip devices. We hypothesize an optimal trunnion design may reduce risk of this failure mode in large-bearing total hip arthroplasty systems. The purpose of this study is to report mid- to long-term outcomes for a single-surgeon series of 211 total hip arthroplasties using the large-bearing Biomet Magnum metal-on-metal system.</p><p><strong>Materials and methods: </strong>Between December 2004 and January 2016, the primary surgeon performed 211 uncemented Magnum total hip arthroplasties in 181 patients. The average length of follow-up was 10.1 ± 3.5 years (range 8-18 years).</p><p><strong>Results: </strong>Using failure of any component as the endpoint, the overall survivorship rate was 98.1% at 10 years and 97.4% at 18 years. These eight failures (3.8% of cohort) included one case of adverse wear-related failure (0.5%), two cases of acetabular ingrowth failure (0.9%), three cases of trunnion corrosion (1.4%), one failure of late infection (0.5%), and one inappropriate revision of components for trochanteric nonunion without instability (0.5%). Excluding failed cases, all components were radiographically stable with no radiolucencies. Except for the one wear failure, ion testing revealed that 97.2% of cases were within optimal whole blood metal ion levels with the remaining ion test results within acceptable levels.</p><p><strong>Conclusions: </strong>With the uncemented Magnum metal-on-metal total hip, we achieved 97.4% 18-year implant survivorship, exceeding the NICE criteria and registry benchmarks for implant survivorship. We observed a trunnion corrosion rate of 1.4% and no cases of instability. The single case of adverse wear-related failure was caused by acetabular component malposition.</p>\",\"PeriodicalId\":18875,\"journal\":{\"name\":\"MUSCULOSKELETAL SURGERY\",\"volume\":\" \",\"pages\":\"449-457\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"MUSCULOSKELETAL SURGERY\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1007/s12306-024-00831-3\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/6/4 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"MUSCULOSKELETAL SURGERY","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1007/s12306-024-00831-3","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/6/4 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

背景:有关金属碎片不良反应的报道在一定程度上导致了大承载金属全髋关节装置使用率的下降。我们假设最佳的耳轴设计可以降低大承重全髋关节置换系统出现这种故障模式的风险。本研究的目的是报告使用大承重Biomet Magnum金属全髋关节置换系统的211例全髋关节置换手术的中长期结果:2004年12月至2016年1月期间,主刀医生为181名患者实施了211例非骨水泥Magnum全髋关节置换术。平均随访时间为 10.1 ± 3.5 年(8-18 年不等):结果:以任何组件的失败作为终点,10 年的总体存活率为 98.1%,18 年的总体存活率为 97.4%。这8例失败病例(占队列的3.8%)包括1例与磨损相关的不良失败(0.5%)、2例髋臼嵌生失败(0.9%)、3例耳轴腐蚀(1.4%)、1例晚期感染失败(0.5%)以及1例因转子不连接而进行的不适当的组件翻修(0.5%)。除去失败的病例,所有组件在影像学上都是稳定的,没有放射性缺陷。除一个磨损失败的病例外,离子检测显示97.2%的病例全血金属离子含量在最佳范围内,其余离子检测结果均在可接受范围内:通过使用非骨水泥Magnum金属全髋关节,我们实现了97.4%的18年植入存活率,超过了NICE标准和植入存活率登记基准。我们观察到耳轴腐蚀率为 1.4%,没有不稳定病例。唯一一例与磨损相关的不良故障是由髋臼组件错位引起的。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Magnum metal-on-metal uncemented total hip replacement: 8- to 18-year outcomes of 211 cases.

Background: Reports of adverse reactions to metal debris contributed in part to a decline in use of large-bearing metal-on-metal total hip devices. We hypothesize an optimal trunnion design may reduce risk of this failure mode in large-bearing total hip arthroplasty systems. The purpose of this study is to report mid- to long-term outcomes for a single-surgeon series of 211 total hip arthroplasties using the large-bearing Biomet Magnum metal-on-metal system.

Materials and methods: Between December 2004 and January 2016, the primary surgeon performed 211 uncemented Magnum total hip arthroplasties in 181 patients. The average length of follow-up was 10.1 ± 3.5 years (range 8-18 years).

Results: Using failure of any component as the endpoint, the overall survivorship rate was 98.1% at 10 years and 97.4% at 18 years. These eight failures (3.8% of cohort) included one case of adverse wear-related failure (0.5%), two cases of acetabular ingrowth failure (0.9%), three cases of trunnion corrosion (1.4%), one failure of late infection (0.5%), and one inappropriate revision of components for trochanteric nonunion without instability (0.5%). Excluding failed cases, all components were radiographically stable with no radiolucencies. Except for the one wear failure, ion testing revealed that 97.2% of cases were within optimal whole blood metal ion levels with the remaining ion test results within acceptable levels.

Conclusions: With the uncemented Magnum metal-on-metal total hip, we achieved 97.4% 18-year implant survivorship, exceeding the NICE criteria and registry benchmarks for implant survivorship. We observed a trunnion corrosion rate of 1.4% and no cases of instability. The single case of adverse wear-related failure was caused by acetabular component malposition.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
MUSCULOSKELETAL SURGERY
MUSCULOSKELETAL SURGERY Medicine-Surgery
CiteScore
4.50
自引率
0.00%
发文量
35
期刊介绍: Musculoskeletal Surgery – Formerly La Chirurgia degli Organi di Movimento, founded in 1917 at the Istituto Ortopedico Rizzoli, is a peer-reviewed journal published three times a year. The journal provides up-to-date information to clinicians and scientists through the publication of original papers, reviews, case reports, and brief communications dealing with the pathogenesis and treatment of orthopaedic conditions.An electronic version is also available at http://www.springerlink.com.The journal is open for publication of supplements and for publishing abstracts of scientific meetings; conditions can be obtained from the Editors-in-Chief or the Publisher.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信