同时测定生物液体和剂型中替诺福韦和阿莫西林的绿色超高效液相色谱法

Q3 Pharmacology, Toxicology and Pharmaceutics
Shaimaa Elsayed, A. El-Gindy, yasser el-shabrawy, ElSayed A. Ibrahim
{"title":"同时测定生物液体和剂型中替诺福韦和阿莫西林的绿色超高效液相色谱法","authors":"Shaimaa Elsayed, A. El-Gindy, yasser el-shabrawy, ElSayed A. Ibrahim","doi":"10.52711/0974-360x.2024.00012","DOIUrl":null,"url":null,"abstract":"This paper elucidates the process of developing and validating the UPLC technique, aiming to enable the concurrent measurement of tenofovir and amoxicillin. This UPLC method is accurate and sensitive. The experimental setup involved utilizing a Kinetex 1.7μ C18 100A (2.1-mm × 50-mm) column maintained at a temperature of 30°C. The mobile phase employed a combination of methanol and phosphate buffer (0.1% orthophosphoric acid in water, pH 3.5) in a ratio of 30:70 (v/v). The flow rate was set at 0.2mL/min, and the detection wavelength for analysis was 230nm. All these conditions resulted in optimum chromatographic separation. This developed method of UPLC took less than 4 minutes to separate this mixture. The detection limits for two medications were determined to be 0.45µg/mL for tenofovir and 0.26µg/mL for amoxicillin. The calibration plots of both substances exhibited satisfactory linearity within the concentration range of 1.0 - 2.0 µg/mL. Remarkably, the method demonstrated excellent performance with high percentage recoveries ranging from 96.8% to 102.92% and low percentage relative standard deviation (%RSD) values below 2%. Consequently, the proposed methodology proved to be highly effective for detecting the targeted drugs in their respective dosage forms.","PeriodicalId":21141,"journal":{"name":"Research Journal of Pharmacy and Technology","volume":"14 5","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-01-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A Green UPLC Method for the Simultaneous determination of Tenofovir and Amoxicillin in Biological Fluids and Dosage Forms\",\"authors\":\"Shaimaa Elsayed, A. El-Gindy, yasser el-shabrawy, ElSayed A. Ibrahim\",\"doi\":\"10.52711/0974-360x.2024.00012\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This paper elucidates the process of developing and validating the UPLC technique, aiming to enable the concurrent measurement of tenofovir and amoxicillin. This UPLC method is accurate and sensitive. The experimental setup involved utilizing a Kinetex 1.7μ C18 100A (2.1-mm × 50-mm) column maintained at a temperature of 30°C. The mobile phase employed a combination of methanol and phosphate buffer (0.1% orthophosphoric acid in water, pH 3.5) in a ratio of 30:70 (v/v). The flow rate was set at 0.2mL/min, and the detection wavelength for analysis was 230nm. All these conditions resulted in optimum chromatographic separation. This developed method of UPLC took less than 4 minutes to separate this mixture. The detection limits for two medications were determined to be 0.45µg/mL for tenofovir and 0.26µg/mL for amoxicillin. The calibration plots of both substances exhibited satisfactory linearity within the concentration range of 1.0 - 2.0 µg/mL. Remarkably, the method demonstrated excellent performance with high percentage recoveries ranging from 96.8% to 102.92% and low percentage relative standard deviation (%RSD) values below 2%. Consequently, the proposed methodology proved to be highly effective for detecting the targeted drugs in their respective dosage forms.\",\"PeriodicalId\":21141,\"journal\":{\"name\":\"Research Journal of Pharmacy and Technology\",\"volume\":\"14 5\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-01-19\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Research Journal of Pharmacy and Technology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.52711/0974-360x.2024.00012\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"Pharmacology, Toxicology and Pharmaceutics\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Research Journal of Pharmacy and Technology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.52711/0974-360x.2024.00012","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"Pharmacology, Toxicology and Pharmaceutics","Score":null,"Total":0}
引用次数: 0

摘要

本文阐述了开发和验证超高效液相色谱技术的过程,旨在实现替诺福韦和阿莫西林的同时测定。这种超高效液相色谱法准确、灵敏。实验装置采用 Kinetex 1.7μ C18 100A(2.1-mm × 50-mm)色谱柱,温度保持在 30°C。流动相采用甲醇和磷酸盐缓冲液(0.1% 正磷酸水溶液,pH 3.5)的组合,比例为 30:70(v/v)。流速设定为 0.2 毫升/分钟,检测波长为 230 纳米。所有这些条件都使色谱分离达到最佳效果。所开发的 UPLC 方法只用了不到 4 分钟就分离了这种混合物。两种药物的检测限分别为替诺福韦 0.45µg/mL 和阿莫西林 0.26µg/mL。在 1.0 - 2.0 微克/毫升的浓度范围内,两种物质的校准图均呈现出令人满意的线性关系。值得注意的是,该方法性能卓越,回收率高达 96.8%至 102.92%,相对标准偏差(%RSD)值低于 2%。因此,所提出的方法被证明对检测目标药物在其各自剂型中的含量非常有效。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
A Green UPLC Method for the Simultaneous determination of Tenofovir and Amoxicillin in Biological Fluids and Dosage Forms
This paper elucidates the process of developing and validating the UPLC technique, aiming to enable the concurrent measurement of tenofovir and amoxicillin. This UPLC method is accurate and sensitive. The experimental setup involved utilizing a Kinetex 1.7μ C18 100A (2.1-mm × 50-mm) column maintained at a temperature of 30°C. The mobile phase employed a combination of methanol and phosphate buffer (0.1% orthophosphoric acid in water, pH 3.5) in a ratio of 30:70 (v/v). The flow rate was set at 0.2mL/min, and the detection wavelength for analysis was 230nm. All these conditions resulted in optimum chromatographic separation. This developed method of UPLC took less than 4 minutes to separate this mixture. The detection limits for two medications were determined to be 0.45µg/mL for tenofovir and 0.26µg/mL for amoxicillin. The calibration plots of both substances exhibited satisfactory linearity within the concentration range of 1.0 - 2.0 µg/mL. Remarkably, the method demonstrated excellent performance with high percentage recoveries ranging from 96.8% to 102.92% and low percentage relative standard deviation (%RSD) values below 2%. Consequently, the proposed methodology proved to be highly effective for detecting the targeted drugs in their respective dosage forms.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Research Journal of Pharmacy and Technology
Research Journal of Pharmacy and Technology Pharmacology, Toxicology and Pharmaceutics-Pharmacology, Toxicology and Pharmaceutics (miscellaneous)
CiteScore
1.40
自引率
0.00%
发文量
0
期刊介绍: Research Journal of Pharmacy and Technology (RJPT) is an international, peer-reviewed, multidisciplinary journal, devoted to pharmaceutical sciences. The aim of RJPT is to increase the impact of pharmaceutical research both in academia and industry, with strong emphasis on quality and originality. RJPT publishes Original Research Articles, Short Communications, Review Articles in all areas of pharmaceutical sciences from the discovery of a drug up to clinical evaluation. Topics covered are: Pharmaceutics and Pharmacokinetics; Pharmaceutical chemistry including medicinal and analytical chemistry; Pharmacognosy including herbal products standardization and Phytochemistry; Pharmacology: Allied sciences including drug regulatory affairs, Pharmaceutical Marketing, Pharmaceutical Microbiology, Pharmaceutical biochemistry, Pharmaceutical Education and Hospital Pharmacy.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信