使用 SSDE 和 K-Means 分段法提高 Samarinda A.W. Sjahranie RSUD 放射科装置中脑卒中患者 CT 图像的质量

E. R. Putri, Fransiska Lipa Payon, Ronel Arida Missinychrista, Devina Rayzy Perwitasari Sutaji Putri, Pratiwi Sri Wardani, Kadek Subagiada
{"title":"使用 SSDE 和 K-Means 分段法提高 Samarinda A.W. Sjahranie RSUD 放射科装置中脑卒中患者 CT 图像的质量","authors":"E. R. Putri, Fransiska Lipa Payon, Ronel Arida Missinychrista, Devina Rayzy Perwitasari Sutaji Putri, Pratiwi Sri Wardani, Kadek Subagiada","doi":"10.20527/flux.v21i1.17100","DOIUrl":null,"url":null,"abstract":"Stroke examination using a CT scan is a technique that produces 3D brain images without having to do surgery. The purpose of this study was to determine the dose received by each patient who underwent a CT-scan procedure and to improve the image quality of the results. Optimization of the dose for each patient is based on a comparison between the calculated SSDE value and the DRL in the form of CTDIvol and DLP, which are implemented in Indonesia. Improving the quality of CT-Scan images through digital processing using the segmentation method to determine the stroke area. The method used is K-Means segmentation. The results obtained for the SSDE value are the dose received by the patient which is higher than the dose issued by the device or the  value. However, the SSDE, CTDIvol, and DLP values are still below the DRL standard range set by BAPETEN, so they are still within the safe inspection threshold. The average CTDIvol value, which is the output dose value of the device from 22 patients, was 41.84 mGy and the 75th percentile value was 42.7 mGy. The average DLP value is 1054.08 mGy.cm and the 75th percentile value of the DLP value is 964.13 mGy.cm. The average value of the SSDE value which is the value of the dose received by the patient is 43.39 mGy and the 75th percentile value is 42.440 mGy. Based on these results, the value of the dose received by the patient is higher than the value of the dose issued by the device. Segmentation using the K-Means method with the number of clusters k = 3 produces an accuracy of 86.3%. This means that the patient receives a dose of CT-Scan radiation within safe limits and his image can be processed properly, as evidenced by the high success of the segmentation process.","PeriodicalId":511747,"journal":{"name":"Jurnal Fisika Flux: Jurnal Ilmiah Fisika FMIPA Universitas Lambung Mangkurat","volume":"29 26","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-02-19","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Improving the Quality of CT Images of Stroke Patients Using the SSDE and K-Means Segmentation in the Radiology Installation of RSUD A.W. Sjahranie Samarinda\",\"authors\":\"E. R. Putri, Fransiska Lipa Payon, Ronel Arida Missinychrista, Devina Rayzy Perwitasari Sutaji Putri, Pratiwi Sri Wardani, Kadek Subagiada\",\"doi\":\"10.20527/flux.v21i1.17100\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Stroke examination using a CT scan is a technique that produces 3D brain images without having to do surgery. The purpose of this study was to determine the dose received by each patient who underwent a CT-scan procedure and to improve the image quality of the results. Optimization of the dose for each patient is based on a comparison between the calculated SSDE value and the DRL in the form of CTDIvol and DLP, which are implemented in Indonesia. Improving the quality of CT-Scan images through digital processing using the segmentation method to determine the stroke area. The method used is K-Means segmentation. The results obtained for the SSDE value are the dose received by the patient which is higher than the dose issued by the device or the  value. However, the SSDE, CTDIvol, and DLP values are still below the DRL standard range set by BAPETEN, so they are still within the safe inspection threshold. The average CTDIvol value, which is the output dose value of the device from 22 patients, was 41.84 mGy and the 75th percentile value was 42.7 mGy. The average DLP value is 1054.08 mGy.cm and the 75th percentile value of the DLP value is 964.13 mGy.cm. The average value of the SSDE value which is the value of the dose received by the patient is 43.39 mGy and the 75th percentile value is 42.440 mGy. Based on these results, the value of the dose received by the patient is higher than the value of the dose issued by the device. Segmentation using the K-Means method with the number of clusters k = 3 produces an accuracy of 86.3%. This means that the patient receives a dose of CT-Scan radiation within safe limits and his image can be processed properly, as evidenced by the high success of the segmentation process.\",\"PeriodicalId\":511747,\"journal\":{\"name\":\"Jurnal Fisika Flux: Jurnal Ilmiah Fisika FMIPA Universitas Lambung Mangkurat\",\"volume\":\"29 26\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-02-19\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Jurnal Fisika Flux: Jurnal Ilmiah Fisika FMIPA Universitas Lambung Mangkurat\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.20527/flux.v21i1.17100\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Jurnal Fisika Flux: Jurnal Ilmiah Fisika FMIPA Universitas Lambung Mangkurat","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.20527/flux.v21i1.17100","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

使用 CT 扫描进行脑卒中检查是一种无需手术即可生成三维脑部图像的技术。这项研究的目的是确定每位接受 CT 扫描的患者所接受的剂量,并提高扫描结果的图像质量。每位患者的剂量优化都是基于计算出的 SSDE 值与以 CTDIvol 和 DLP 形式表示的 DRL 之间的比较,这些都在印度尼西亚实施。通过数字处理提高 CT 扫描图像的质量,使用分割方法确定中风区域。使用的方法是 K-Means 分割法。获得的 SSDE 值结果是患者接受的剂量高于设备发出的剂量或值。不过,SSDE、CTDIvol 和 DLP 值仍低于 BAPETEN 规定的 DRL 标准范围,因此仍在安全检查阈值之内。CTDIvol 平均值(即 22 位患者的设备输出剂量值)为 41.84 mGy,第 75 百分位值为 42.7 mGy。DLP 平均值为 1054.08 mGy.cm,第 75 百分位值为 964.13 mGy.cm。SSDE 值(即患者接受的剂量值)的平均值为 43.39 mGy,第 75 百分位数值为 42.440 mGy。根据这些结果,患者接受的剂量值高于设备发出的剂量值。使用 K-Means 方法(簇数 k = 3)进行分段,准确率为 86.3%。这意味着患者接受的 CT 扫描辐射剂量在安全范围内,其图像也能得到正确处理,分割过程的高度成功就是明证。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Improving the Quality of CT Images of Stroke Patients Using the SSDE and K-Means Segmentation in the Radiology Installation of RSUD A.W. Sjahranie Samarinda
Stroke examination using a CT scan is a technique that produces 3D brain images without having to do surgery. The purpose of this study was to determine the dose received by each patient who underwent a CT-scan procedure and to improve the image quality of the results. Optimization of the dose for each patient is based on a comparison between the calculated SSDE value and the DRL in the form of CTDIvol and DLP, which are implemented in Indonesia. Improving the quality of CT-Scan images through digital processing using the segmentation method to determine the stroke area. The method used is K-Means segmentation. The results obtained for the SSDE value are the dose received by the patient which is higher than the dose issued by the device or the  value. However, the SSDE, CTDIvol, and DLP values are still below the DRL standard range set by BAPETEN, so they are still within the safe inspection threshold. The average CTDIvol value, which is the output dose value of the device from 22 patients, was 41.84 mGy and the 75th percentile value was 42.7 mGy. The average DLP value is 1054.08 mGy.cm and the 75th percentile value of the DLP value is 964.13 mGy.cm. The average value of the SSDE value which is the value of the dose received by the patient is 43.39 mGy and the 75th percentile value is 42.440 mGy. Based on these results, the value of the dose received by the patient is higher than the value of the dose issued by the device. Segmentation using the K-Means method with the number of clusters k = 3 produces an accuracy of 86.3%. This means that the patient receives a dose of CT-Scan radiation within safe limits and his image can be processed properly, as evidenced by the high success of the segmentation process.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信