用近红外光谱测定冻干制剂中的水分含量

IF 1.6 4区 化学 Q3 CHEMISTRY, APPLIED
Aruna Khanolkar, Pranita Pawale, Viraj Thorat, Bhaskar Patil, Gautam Samanta
{"title":"用近红外光谱测定冻干制剂中的水分含量","authors":"Aruna Khanolkar, Pranita Pawale, Viraj Thorat, Bhaskar Patil, Gautam Samanta","doi":"10.1177/09670335241240309","DOIUrl":null,"url":null,"abstract":"A non-invasive near infrared (NIR) spectroscopic method was developed for the quantitative moisture determination in a lyophilized injection formulation. The calibration samples were prepared by exposing lyophilized samples at different temperatures and relative humidity. The samples from different scales and different process parameters were considered for adding robustness to the model. The NIR spectra were collected using a Fourier- transform (FT) NIR with a diffuse reflectance probe and the same samples were further analyzed by the Karl Fisher (KF) method for moisture content. The pre-treated NIR spectra were used for quantitative method development for moisture content. Partial least squares (PLS) regression was used to develop calibrations in the 5600-4950 cm<jats:sup>−1</jats:sup> region with calibration coefficient of determination (R<jats:sup>2</jats:sup>) of 0.96 and root mean square error of calibration (RMSEC) of 0.149. The model was cross-validated internally using the Kernel algorithm with r<jats:sup>2</jats:sup> = 0.96 and RMSECV = 0.15. The accuracy of the NIR method against the KF method, precision, and reproducibility were good and the model was robust in predicting different external validation samples. This work allowed NIR as an alternative measurement for moisture analysis as well as facilitate 100% monitoring before packaging and save the cost of sample and time of KF analysis.","PeriodicalId":16551,"journal":{"name":"Journal of Near Infrared Spectroscopy","volume":null,"pages":null},"PeriodicalIF":1.6000,"publicationDate":"2024-03-27","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Near infrared spectroscopy for determination of moisture content in lyophilized formulation\",\"authors\":\"Aruna Khanolkar, Pranita Pawale, Viraj Thorat, Bhaskar Patil, Gautam Samanta\",\"doi\":\"10.1177/09670335241240309\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"A non-invasive near infrared (NIR) spectroscopic method was developed for the quantitative moisture determination in a lyophilized injection formulation. The calibration samples were prepared by exposing lyophilized samples at different temperatures and relative humidity. The samples from different scales and different process parameters were considered for adding robustness to the model. The NIR spectra were collected using a Fourier- transform (FT) NIR with a diffuse reflectance probe and the same samples were further analyzed by the Karl Fisher (KF) method for moisture content. The pre-treated NIR spectra were used for quantitative method development for moisture content. Partial least squares (PLS) regression was used to develop calibrations in the 5600-4950 cm<jats:sup>−1</jats:sup> region with calibration coefficient of determination (R<jats:sup>2</jats:sup>) of 0.96 and root mean square error of calibration (RMSEC) of 0.149. The model was cross-validated internally using the Kernel algorithm with r<jats:sup>2</jats:sup> = 0.96 and RMSECV = 0.15. The accuracy of the NIR method against the KF method, precision, and reproducibility were good and the model was robust in predicting different external validation samples. This work allowed NIR as an alternative measurement for moisture analysis as well as facilitate 100% monitoring before packaging and save the cost of sample and time of KF analysis.\",\"PeriodicalId\":16551,\"journal\":{\"name\":\"Journal of Near Infrared Spectroscopy\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":1.6000,\"publicationDate\":\"2024-03-27\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Near Infrared Spectroscopy\",\"FirstCategoryId\":\"92\",\"ListUrlMain\":\"https://doi.org/10.1177/09670335241240309\",\"RegionNum\":4,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q3\",\"JCRName\":\"CHEMISTRY, APPLIED\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Near Infrared Spectroscopy","FirstCategoryId":"92","ListUrlMain":"https://doi.org/10.1177/09670335241240309","RegionNum":4,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q3","JCRName":"CHEMISTRY, APPLIED","Score":null,"Total":0}
引用次数: 0

摘要

本研究开发了一种非侵入式近红外光谱法,用于定量测定冻干注射剂配方中的水分。通过在不同温度和相对湿度下暴露冻干样品来制备校准样品。为了增加模型的稳健性,考虑了不同规模和不同工艺参数的样品。使用带有漫反射探头的傅立叶变换(FT)近红外光谱采集近红外光谱,并采用卡尔-费雪(KF)方法进一步分析相同样品的水分含量。预处理后的近红外光谱用于水分含量定量方法的开发。使用偏最小二乘法(PLS)回归法在 5600-4950 cm-1 区域进行定标,定标系数(R2)为 0.96,定标均方根误差(RMSEC)为 0.149。使用核算法对模型进行了内部交叉验证,r2 = 0.96,RMSECV = 0.15。与 KF 方法相比,近红外方法的准确度、精确度和重现性都很好,而且该模型在预测不同的外部验证样本时也很稳健。这项工作使近红外法成为水分分析的替代测量方法,并促进了包装前的 100% 监测,节省了样品成本和 KF 分析时间。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Near infrared spectroscopy for determination of moisture content in lyophilized formulation
A non-invasive near infrared (NIR) spectroscopic method was developed for the quantitative moisture determination in a lyophilized injection formulation. The calibration samples were prepared by exposing lyophilized samples at different temperatures and relative humidity. The samples from different scales and different process parameters were considered for adding robustness to the model. The NIR spectra were collected using a Fourier- transform (FT) NIR with a diffuse reflectance probe and the same samples were further analyzed by the Karl Fisher (KF) method for moisture content. The pre-treated NIR spectra were used for quantitative method development for moisture content. Partial least squares (PLS) regression was used to develop calibrations in the 5600-4950 cm−1 region with calibration coefficient of determination (R2) of 0.96 and root mean square error of calibration (RMSEC) of 0.149. The model was cross-validated internally using the Kernel algorithm with r2 = 0.96 and RMSECV = 0.15. The accuracy of the NIR method against the KF method, precision, and reproducibility were good and the model was robust in predicting different external validation samples. This work allowed NIR as an alternative measurement for moisture analysis as well as facilitate 100% monitoring before packaging and save the cost of sample and time of KF analysis.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
CiteScore
3.30
自引率
5.60%
发文量
35
审稿时长
6 months
期刊介绍: JNIRS — Journal of Near Infrared Spectroscopy is a peer reviewed journal, publishing original research papers, short communications, review articles and letters concerned with near infrared spectroscopy and technology, its application, new instrumentation and the use of chemometric and data handling techniques within NIR.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信