{"title":"新一代螺旋酮透析器的功效和安全性比较","authors":"Francisco Maduell ,&nbsp;José Jesús Broseta ,&nbsp;Diana Rodríguez-Espinosa ,&nbsp;Lida María Rodas ,&nbsp;Miquel Gómez ,&nbsp;Marta Arias-Guillén ,&nbsp;Néstor Fontseré ,&nbsp;Manel Vera ,&nbsp;Maria del Carmen Salgado ,&nbsp;Nayra Rico ,&nbsp;Rosa Ramos","doi":"10.1016/j.nefro.2024.02.005","DOIUrl":null,"url":null,"abstract":"<div><h3>Introduction</h3><p>New generation helixone dialyzers has recently been developed as part of the ongoing effort to improve dialyzer hemocompatibility and avoid adverse reactions to synthetic dialyzers. This study aimed to assess the performance and albumin loss of this new dialyzer series in hemodiafiltration and compare it with the previous generation helixone series.</p></div><div><h3>Material and methods</h3><p>A prospective study was conducted in 19 patients. Each patient underwent eight dialysis sessions with the same routine dialysis parameters; only the dialyzer varied: FX60 CorDiax, FX CorAL 60, FX600 CorDiax, FX CorAL 600, FX80 CorDiax, FX CorAL 80, FX800 CorDiax, and FX CorAL 800. The reduction ratios (RR) of urea, creatinine, β<sub>2</sub>-microglobulin, myoglobin, kappa-free immunoglobulin light chains (κFLC), prolactin, α<sub>1</sub>-microglobulin, α<sub>1</sub>-acid glycoprotein, lambda immunoglobulin light chains (λFLC), and albumin were compared intra-individually. Dialysate albumin loss was also measured.</p></div><div><h3>Results</h3><p>All treatments were well tolerated. The mean amount of replacement fluid ranged from 31 to 34<!--> <!-->litres. Comparison of dialysis treatments showed no differences between small molecules and even up to those the size of β<sub>2</sub>-microglobulins. Little differences were found between myoglobin, κFLC, prolactin, α<sub>1</sub>-microglobulin, and λFLC RRs, and only FX80 CorDiax was slightly superior to the others. Mean dialysate albumin losses were similar, with less than 2.5<!--> <!-->grams lost in each dialyzer. The FX80 CorDiax showed slightly higher global removal scores than the other dialyzers evaluated, except for FX CorAL 800.</p></div><div><h3>Conclusion</h3><p>The new generation helixone dialyzers series has been updated to minimise the risk of adverse reactions, while maintaining the effectiveness and albumin loss achieved by the previous most advanced helixone generation.</p></div>","PeriodicalId":18997,"journal":{"name":"Nefrologia","volume":"44 3","pages":"Pages 354-361"},"PeriodicalIF":2.0000,"publicationDate":"2024-05-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.sciencedirect.com/science/article/pii/S021169952400016X/pdfft?md5=2e453e9cfb8175b9515cc67280765dbe&pid=1-s2.0-S021169952400016X-main.pdf","citationCount":"0","resultStr":"{\"title\":\"Comparación de la eficacia y seguridad de la nueva generación dializadores de helixona\",\"authors\":\"Francisco Maduell ,&nbsp;José Jesús Broseta ,&nbsp;Diana Rodríguez-Espinosa ,&nbsp;Lida María Rodas ,&nbsp;Miquel Gómez ,&nbsp;Marta Arias-Guillén ,&nbsp;Néstor Fontseré ,&nbsp;Manel Vera ,&nbsp;Maria del Carmen Salgado ,&nbsp;Nayra Rico ,&nbsp;Rosa Ramos\",\"doi\":\"10.1016/j.nefro.2024.02.005\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<div><h3>Introduction</h3><p>New generation helixone dialyzers has recently been developed as part of the ongoing effort to improve dialyzer hemocompatibility and avoid adverse reactions to synthetic dialyzers. This study aimed to assess the performance and albumin loss of this new dialyzer series in hemodiafiltration and compare it with the previous generation helixone series.</p></div><div><h3>Material and methods</h3><p>A prospective study was conducted in 19 patients. Each patient underwent eight dialysis sessions with the same routine dialysis parameters; only the dialyzer varied: FX60 CorDiax, FX CorAL 60, FX600 CorDiax, FX CorAL 600, FX80 CorDiax, FX CorAL 80, FX800 CorDiax, and FX CorAL 800. The reduction ratios (RR) of urea, creatinine, β<sub>2</sub>-microglobulin, myoglobin, kappa-free immunoglobulin light chains (κFLC), prolactin, α<sub>1</sub>-microglobulin, α<sub>1</sub>-acid glycoprotein, lambda immunoglobulin light chains (λFLC), and albumin were compared intra-individually. Dialysate albumin loss was also measured.</p></div><div><h3>Results</h3><p>All treatments were well tolerated. The mean amount of replacement fluid ranged from 31 to 34<!--> <!-->litres. Comparison of dialysis treatments showed no differences between small molecules and even up to those the size of β<sub>2</sub>-microglobulins. Little differences were found between myoglobin, κFLC, prolactin, α<sub>1</sub>-microglobulin, and λFLC RRs, and only FX80 CorDiax was slightly superior to the others. Mean dialysate albumin losses were similar, with less than 2.5<!--> <!-->grams lost in each dialyzer. The FX80 CorDiax showed slightly higher global removal scores than the other dialyzers evaluated, except for FX CorAL 800.</p></div><div><h3>Conclusion</h3><p>The new generation helixone dialyzers series has been updated to minimise the risk of adverse reactions, while maintaining the effectiveness and albumin loss achieved by the previous most advanced helixone generation.</p></div>\",\"PeriodicalId\":18997,\"journal\":{\"name\":\"Nefrologia\",\"volume\":\"44 3\",\"pages\":\"Pages 354-361\"},\"PeriodicalIF\":2.0000,\"publicationDate\":\"2024-05-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.sciencedirect.com/science/article/pii/S021169952400016X/pdfft?md5=2e453e9cfb8175b9515cc67280765dbe&pid=1-s2.0-S021169952400016X-main.pdf\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nefrologia\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://www.sciencedirect.com/science/article/pii/S021169952400016X\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q2\",\"JCRName\":\"UROLOGY & NEPHROLOGY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nefrologia","FirstCategoryId":"3","ListUrlMain":"https://www.sciencedirect.com/science/article/pii/S021169952400016X","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q2","JCRName":"UROLOGY & NEPHROLOGY","Score":null,"Total":0}
引用次数: 0

摘要

导言:为了改善透析器的血液相容性,避免合成透析器的不良反应,最近开发出了新一代螺旋酮透析器。本研究旨在评估这种新型透析器系列在血液透析过滤中的性能和白蛋白流失情况,并将其与上一代螺旋酮系列透析器进行比较。每位患者均接受了八次透析,常规透析参数相同,只是透析器不同:FX60 CorDiax、FX CorAL 60、FX600 CorDiax、FX CorAL 600、FX80 CorDiax、FX CorAL 80、FX800 CorDiax 和 FX CorAL 800。对尿素、肌酐、β2-微球蛋白、肌红蛋白、无卡帕免疫球蛋白轻链 (κFLC)、催乳素、α1-微球蛋白、α1-酸性糖蛋白、λ免疫球蛋白轻链 (λFLC) 和白蛋白的还原率 (RR) 进行了单独比较。结果 所有治疗的耐受性都很好。平均补充液量为 31 至 34 升。透析治疗的比较结果显示,小分子透析剂和β2微球蛋白透析剂之间没有差异。肌红蛋白、κFLC、催乳素、α1-微球蛋白和λFLC RR之间的差异很小,只有FX80 CorDiax略优于其他产品。平均透析液白蛋白损失量相似,每种透析器的损失量均小于 2.5 克。除 FX CorAL 800 外,FX80 CorDiax 的总体去除率评分略高于其他接受评估的透析器。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Comparación de la eficacia y seguridad de la nueva generación dializadores de helixona

Introduction

New generation helixone dialyzers has recently been developed as part of the ongoing effort to improve dialyzer hemocompatibility and avoid adverse reactions to synthetic dialyzers. This study aimed to assess the performance and albumin loss of this new dialyzer series in hemodiafiltration and compare it with the previous generation helixone series.

Material and methods

A prospective study was conducted in 19 patients. Each patient underwent eight dialysis sessions with the same routine dialysis parameters; only the dialyzer varied: FX60 CorDiax, FX CorAL 60, FX600 CorDiax, FX CorAL 600, FX80 CorDiax, FX CorAL 80, FX800 CorDiax, and FX CorAL 800. The reduction ratios (RR) of urea, creatinine, β2-microglobulin, myoglobin, kappa-free immunoglobulin light chains (κFLC), prolactin, α1-microglobulin, α1-acid glycoprotein, lambda immunoglobulin light chains (λFLC), and albumin were compared intra-individually. Dialysate albumin loss was also measured.

Results

All treatments were well tolerated. The mean amount of replacement fluid ranged from 31 to 34 litres. Comparison of dialysis treatments showed no differences between small molecules and even up to those the size of β2-microglobulins. Little differences were found between myoglobin, κFLC, prolactin, α1-microglobulin, and λFLC RRs, and only FX80 CorDiax was slightly superior to the others. Mean dialysate albumin losses were similar, with less than 2.5 grams lost in each dialyzer. The FX80 CorDiax showed slightly higher global removal scores than the other dialyzers evaluated, except for FX CorAL 800.

Conclusion

The new generation helixone dialyzers series has been updated to minimise the risk of adverse reactions, while maintaining the effectiveness and albumin loss achieved by the previous most advanced helixone generation.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Nefrologia
Nefrologia 医学-泌尿学与肾脏学
CiteScore
3.40
自引率
7.70%
发文量
148
审稿时长
47 days
期刊介绍: Nefrología is the official publication of the Spanish Society of Nephrology. The Journal publishes articles on basic or clinical research relating to nephrology, arterial hypertension, dialysis and kidney transplants. It is governed by the peer review system and all original papers are subject to internal assessment and external reviews. The journal accepts submissions of articles in English and in Spanish languages.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信