{"title":"从电诱发复合动作电位记录中去除伪影的新方法:单脉冲刺激","authors":"Jeffrey Skidmore, Yi Yuan, Shuman He","doi":"10.1101/2024.01.17.24301435","DOIUrl":null,"url":null,"abstract":"This report presents a new method for removing electrical artifact contamination from the electrically evoked compound action potential (eCAP) evoked by single cathodic-leading, biphasic-pulse stimulation. The development of the new method is motivated by results recorded in human cochlear implant (CI) users showing that the fundamental assumption of the classic forward masking artifact rejection technique is violated in up to 45% of cases tested at high stimulation levels when using default stimulation parameters. Subsequently, the new method developed based on the discovery that a hyperbola best characterizes the artifacts created during stimulation and recording is described. The eCAP waveforms obtained using the new method are compared to those recorded using the classic forward masking technique. The results show that eCAP waveforms obtained using both methods are comparable when the fundamental assumption of the classic forward masking technique is met. In contrast, eCAP amplitudes obtained using the two methods are significantly different when the fundamental assumption of the classic forward masking technique is violated, with greater differences in the eCAP amplitude for greater assumption violations. The new method also has excellent test-retest reliability (Intraclass correlation > 0.98). Overall, the new method is a viable alternative to the classic forward masking technique for obtaining artifact-free eCAPs evoked by single-pulse stimulation in CI users.","PeriodicalId":501185,"journal":{"name":"medRxiv - Otolaryngology","volume":"48 1","pages":""},"PeriodicalIF":0.0000,"publicationDate":"2024-01-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"A new method for removing artifacts from recordings of the electrically evoked compound action potential: Single-pulse stimulation\",\"authors\":\"Jeffrey Skidmore, Yi Yuan, Shuman He\",\"doi\":\"10.1101/2024.01.17.24301435\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"This report presents a new method for removing electrical artifact contamination from the electrically evoked compound action potential (eCAP) evoked by single cathodic-leading, biphasic-pulse stimulation. The development of the new method is motivated by results recorded in human cochlear implant (CI) users showing that the fundamental assumption of the classic forward masking artifact rejection technique is violated in up to 45% of cases tested at high stimulation levels when using default stimulation parameters. Subsequently, the new method developed based on the discovery that a hyperbola best characterizes the artifacts created during stimulation and recording is described. The eCAP waveforms obtained using the new method are compared to those recorded using the classic forward masking technique. The results show that eCAP waveforms obtained using both methods are comparable when the fundamental assumption of the classic forward masking technique is met. In contrast, eCAP amplitudes obtained using the two methods are significantly different when the fundamental assumption of the classic forward masking technique is violated, with greater differences in the eCAP amplitude for greater assumption violations. The new method also has excellent test-retest reliability (Intraclass correlation > 0.98). Overall, the new method is a viable alternative to the classic forward masking technique for obtaining artifact-free eCAPs evoked by single-pulse stimulation in CI users.\",\"PeriodicalId\":501185,\"journal\":{\"name\":\"medRxiv - Otolaryngology\",\"volume\":\"48 1\",\"pages\":\"\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2024-01-17\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"medRxiv - Otolaryngology\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1101/2024.01.17.24301435\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"medRxiv - Otolaryngology","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1101/2024.01.17.24301435","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
A new method for removing artifacts from recordings of the electrically evoked compound action potential: Single-pulse stimulation
This report presents a new method for removing electrical artifact contamination from the electrically evoked compound action potential (eCAP) evoked by single cathodic-leading, biphasic-pulse stimulation. The development of the new method is motivated by results recorded in human cochlear implant (CI) users showing that the fundamental assumption of the classic forward masking artifact rejection technique is violated in up to 45% of cases tested at high stimulation levels when using default stimulation parameters. Subsequently, the new method developed based on the discovery that a hyperbola best characterizes the artifacts created during stimulation and recording is described. The eCAP waveforms obtained using the new method are compared to those recorded using the classic forward masking technique. The results show that eCAP waveforms obtained using both methods are comparable when the fundamental assumption of the classic forward masking technique is met. In contrast, eCAP amplitudes obtained using the two methods are significantly different when the fundamental assumption of the classic forward masking technique is violated, with greater differences in the eCAP amplitude for greater assumption violations. The new method also has excellent test-retest reliability (Intraclass correlation > 0.98). Overall, the new method is a viable alternative to the classic forward masking technique for obtaining artifact-free eCAPs evoked by single-pulse stimulation in CI users.