使用 NanoScope 的腕关节镜瓣比传统关节镜手术更安全。

IF 1.8 Q2 ORTHOPEDICS
HAND Pub Date : 2025-06-01 Epub Date: 2024-01-18 DOI:10.1177/15589447231221168
Lauren E Dittman, Nicholas Munaretto, Katharine Hinchcliff, Lauren Dutton, Sanjeev Kakar
{"title":"使用 NanoScope 的腕关节镜瓣比传统关节镜手术更安全。","authors":"Lauren E Dittman, Nicholas Munaretto, Katharine Hinchcliff, Lauren Dutton, Sanjeev Kakar","doi":"10.1177/15589447231221168","DOIUrl":null,"url":null,"abstract":"<p><strong>Background: </strong>The NanoScope, given its smaller size, may be further from critical structures when establishing volar wrist arthroscopy portals compared to the traditional 2.7-mm arthroscope.</p><p><strong>Methods: </strong>Ten fresh-frozen latex-injected cadaver specimens were utilized. The volar radial (VR) and volar ulnar (VU) portals were created using an inside-out approach. The volar radial midcarpal (VR-MC) and volar ulnar midcarpal (VU-MC) portals were created using an inside-out approach. The arm was then dissected under 3.5-mm loupe magnification. Digital calipers were used to measure the distance between the portals and the surrounding anatomic structures by two fellowship-trained hand surgeons.</p><p><strong>Results: </strong>The median nerve was on average 1.6, 7.2, 1.8, and 5.6 mm away from the trochar for the VR, VU, VR-MC, and VU-MC, respectively. The median nerve fascia was pierced by the VR portal in one specimen. The radial artery, ulnar artery, and ulnar nerve were not pierced in any specimen. Compared to historical controls, for the VR-MC portal, the NanoScope was further from all critical structures, aside from the radial artery, and did not pierce any neurovascular structures. For the VU-MC portal, the NanoScope was further from the ulnar artery and median nerve and did not pierce any neurovascular structures.</p><p><strong>Conclusion: </strong>In developing volar portals, after placing the cannula through the volar capsule from an inside-out approach, we recommend making a volar incision and dissecting the local anatomic structures to guard against injury. Compared to historical controls, the NanoScope was typically further from neurovascular structures.</p>","PeriodicalId":12902,"journal":{"name":"HAND","volume":" ","pages":"602-606"},"PeriodicalIF":1.8000,"publicationDate":"2025-06-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11571708/pdf/","citationCount":"0","resultStr":"{\"title\":\"Volar Wrist Arthroscopy Portals Using the NanoScope Are Safer Than Traditional Arthroscopy.\",\"authors\":\"Lauren E Dittman, Nicholas Munaretto, Katharine Hinchcliff, Lauren Dutton, Sanjeev Kakar\",\"doi\":\"10.1177/15589447231221168\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Background: </strong>The NanoScope, given its smaller size, may be further from critical structures when establishing volar wrist arthroscopy portals compared to the traditional 2.7-mm arthroscope.</p><p><strong>Methods: </strong>Ten fresh-frozen latex-injected cadaver specimens were utilized. The volar radial (VR) and volar ulnar (VU) portals were created using an inside-out approach. The volar radial midcarpal (VR-MC) and volar ulnar midcarpal (VU-MC) portals were created using an inside-out approach. The arm was then dissected under 3.5-mm loupe magnification. Digital calipers were used to measure the distance between the portals and the surrounding anatomic structures by two fellowship-trained hand surgeons.</p><p><strong>Results: </strong>The median nerve was on average 1.6, 7.2, 1.8, and 5.6 mm away from the trochar for the VR, VU, VR-MC, and VU-MC, respectively. The median nerve fascia was pierced by the VR portal in one specimen. The radial artery, ulnar artery, and ulnar nerve were not pierced in any specimen. Compared to historical controls, for the VR-MC portal, the NanoScope was further from all critical structures, aside from the radial artery, and did not pierce any neurovascular structures. For the VU-MC portal, the NanoScope was further from the ulnar artery and median nerve and did not pierce any neurovascular structures.</p><p><strong>Conclusion: </strong>In developing volar portals, after placing the cannula through the volar capsule from an inside-out approach, we recommend making a volar incision and dissecting the local anatomic structures to guard against injury. Compared to historical controls, the NanoScope was typically further from neurovascular structures.</p>\",\"PeriodicalId\":12902,\"journal\":{\"name\":\"HAND\",\"volume\":\" \",\"pages\":\"602-606\"},\"PeriodicalIF\":1.8000,\"publicationDate\":\"2025-06-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"https://www.ncbi.nlm.nih.gov/pmc/articles/PMC11571708/pdf/\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"HAND\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1177/15589447231221168\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/1/18 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q2\",\"JCRName\":\"ORTHOPEDICS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"HAND","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1177/15589447231221168","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/18 0:00:00","PubModel":"Epub","JCR":"Q2","JCRName":"ORTHOPEDICS","Score":null,"Total":0}
引用次数: 0

摘要

背景:由于 NanoScope 的尺寸较小,因此在建立腕关节镜下的腕关节切口时,可能会离关键结构更远:与传统的 2.7 毫米关节镜相比,NanoScope 由于尺寸较小,在建立腕关节镜下的腕关节侧切口时可能离关键结构较远:方法:使用 10 个新鲜冷冻的乳胶注射尸体标本。采用内向外的方法创建桡侧(VR)和尺侧(VU)门户。外侧桡骨中段(VR-MC)和外侧尺骨中段(VU-MC)切口采用由内向外的方法创建。然后在 3.5 毫米放大镜下解剖手臂。两位受过手外科医生研究培训的医生使用数字卡尺测量门户与周围解剖结构之间的距离:结果:VR、VU、VR-MC 和 VU-MC 正中神经与套管的平均距离分别为 1.6、7.2、1.8 和 5.6 毫米。一个标本的正中神经筋膜被 VR 门刺穿。没有任何标本刺穿桡动脉、尺动脉和尺神经。与历史对照组相比,对于 VR-MC 切口,除了桡动脉外,NanoScope 离所有关键结构都更远,没有刺穿任何神经血管结构。对于 VU-MC 门户,NanoScope 距离尺动脉和正中神经更远,没有刺穿任何神经血管结构:结论:在开发侧门时,从内向外将套管穿过侧囊后,我们建议做一个侧切口并解剖局部解剖结构,以防止损伤。与历史对照组相比,NanoScope 通常离神经血管结构更远。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Volar Wrist Arthroscopy Portals Using the NanoScope Are Safer Than Traditional Arthroscopy.

Background: The NanoScope, given its smaller size, may be further from critical structures when establishing volar wrist arthroscopy portals compared to the traditional 2.7-mm arthroscope.

Methods: Ten fresh-frozen latex-injected cadaver specimens were utilized. The volar radial (VR) and volar ulnar (VU) portals were created using an inside-out approach. The volar radial midcarpal (VR-MC) and volar ulnar midcarpal (VU-MC) portals were created using an inside-out approach. The arm was then dissected under 3.5-mm loupe magnification. Digital calipers were used to measure the distance between the portals and the surrounding anatomic structures by two fellowship-trained hand surgeons.

Results: The median nerve was on average 1.6, 7.2, 1.8, and 5.6 mm away from the trochar for the VR, VU, VR-MC, and VU-MC, respectively. The median nerve fascia was pierced by the VR portal in one specimen. The radial artery, ulnar artery, and ulnar nerve were not pierced in any specimen. Compared to historical controls, for the VR-MC portal, the NanoScope was further from all critical structures, aside from the radial artery, and did not pierce any neurovascular structures. For the VU-MC portal, the NanoScope was further from the ulnar artery and median nerve and did not pierce any neurovascular structures.

Conclusion: In developing volar portals, after placing the cannula through the volar capsule from an inside-out approach, we recommend making a volar incision and dissecting the local anatomic structures to guard against injury. Compared to historical controls, the NanoScope was typically further from neurovascular structures.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
HAND
HAND Medicine-Surgery
CiteScore
3.30
自引率
0.00%
发文量
209
期刊介绍: HAND is the official journal of the American Association for Hand Surgery and is a peer-reviewed journal featuring articles written by clinicians worldwide presenting current research and clinical work in the field of hand surgery. It features articles related to all aspects of hand and upper extremity surgery and the post operative care and rehabilitation of the hand.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:604180095
Book学术官方微信