{"title":"伽马通过率的摇摆特性是病人特定质量保证中校准场大小的函数","authors":"Sathiya Raj, Venugopal Sundaram, H. Godson, Retna John","doi":"10.1017/S1460396923000444","DOIUrl":null,"url":null,"abstract":"Abstract Purpose: This study aimed to investigate the influence of calibration field size on the gamma passing rate (GPR) in patient-specific quality assurance (PSQA). Methods: Two independent detectors, PTW OCTAVIUS 4D (4DOCT) and Arc Check, were utilised in volumetric modulated arc therapy plans for 26 patients (14 with Arc Check and 12 with 4DOCT). Plans were administered using Varian Unique machine (with 4DOCT) and Varian TrueBeam (with Arc Check), each employing different calibration factors (CFs): 4 × 4, 6 × 6, 8 × 8, 10 × 10, 12 × 12 and 15 × 15 cm2 field sizes. Gamma analysis was conducted with 2%2mm, 2%3mm and 3%3mm gamma criteria. Results: GPR exhibited variations across different CFs. GPR demonstrated an increasing trend below 10 × 10 cm² CFs, while it displayed a decreasing trend above 10 × 10 cm². Both detectors exhibited similar GPR patterns. The correlation between 4DOCT and Arc Check was strong in tighter criteria (2%2mm) with an R² value of 0·9957, moderate criteria (2%3mm) with an R² value of 0·9868, but reduced in liberal criteria (3%3mm) with an R² value of 0·4226. Conclusion: This study demonstrates that calibration field sizes significantly influence GPR in PSQA. This study recommends the plan specific calibration field must obtain to calibrate the QA devices for modulated plans.","PeriodicalId":44597,"journal":{"name":"Journal of Radiotherapy in Practice","volume":"12 3","pages":""},"PeriodicalIF":0.3000,"publicationDate":"2024-01-11","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Wobbling nature of gamma passing rate as a function of calibration field sizes in patient-specific quality assurance\",\"authors\":\"Sathiya Raj, Venugopal Sundaram, H. Godson, Retna John\",\"doi\":\"10.1017/S1460396923000444\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Abstract Purpose: This study aimed to investigate the influence of calibration field size on the gamma passing rate (GPR) in patient-specific quality assurance (PSQA). Methods: Two independent detectors, PTW OCTAVIUS 4D (4DOCT) and Arc Check, were utilised in volumetric modulated arc therapy plans for 26 patients (14 with Arc Check and 12 with 4DOCT). Plans were administered using Varian Unique machine (with 4DOCT) and Varian TrueBeam (with Arc Check), each employing different calibration factors (CFs): 4 × 4, 6 × 6, 8 × 8, 10 × 10, 12 × 12 and 15 × 15 cm2 field sizes. Gamma analysis was conducted with 2%2mm, 2%3mm and 3%3mm gamma criteria. Results: GPR exhibited variations across different CFs. GPR demonstrated an increasing trend below 10 × 10 cm² CFs, while it displayed a decreasing trend above 10 × 10 cm². Both detectors exhibited similar GPR patterns. The correlation between 4DOCT and Arc Check was strong in tighter criteria (2%2mm) with an R² value of 0·9957, moderate criteria (2%3mm) with an R² value of 0·9868, but reduced in liberal criteria (3%3mm) with an R² value of 0·4226. Conclusion: This study demonstrates that calibration field sizes significantly influence GPR in PSQA. This study recommends the plan specific calibration field must obtain to calibrate the QA devices for modulated plans.\",\"PeriodicalId\":44597,\"journal\":{\"name\":\"Journal of Radiotherapy in Practice\",\"volume\":\"12 3\",\"pages\":\"\"},\"PeriodicalIF\":0.3000,\"publicationDate\":\"2024-01-11\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Radiotherapy in Practice\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1017/S1460396923000444\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Radiotherapy in Practice","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1017/S1460396923000444","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"RADIOLOGY, NUCLEAR MEDICINE & MEDICAL IMAGING","Score":null,"Total":0}
Wobbling nature of gamma passing rate as a function of calibration field sizes in patient-specific quality assurance
Abstract Purpose: This study aimed to investigate the influence of calibration field size on the gamma passing rate (GPR) in patient-specific quality assurance (PSQA). Methods: Two independent detectors, PTW OCTAVIUS 4D (4DOCT) and Arc Check, were utilised in volumetric modulated arc therapy plans for 26 patients (14 with Arc Check and 12 with 4DOCT). Plans were administered using Varian Unique machine (with 4DOCT) and Varian TrueBeam (with Arc Check), each employing different calibration factors (CFs): 4 × 4, 6 × 6, 8 × 8, 10 × 10, 12 × 12 and 15 × 15 cm2 field sizes. Gamma analysis was conducted with 2%2mm, 2%3mm and 3%3mm gamma criteria. Results: GPR exhibited variations across different CFs. GPR demonstrated an increasing trend below 10 × 10 cm² CFs, while it displayed a decreasing trend above 10 × 10 cm². Both detectors exhibited similar GPR patterns. The correlation between 4DOCT and Arc Check was strong in tighter criteria (2%2mm) with an R² value of 0·9957, moderate criteria (2%3mm) with an R² value of 0·9868, but reduced in liberal criteria (3%3mm) with an R² value of 0·4226. Conclusion: This study demonstrates that calibration field sizes significantly influence GPR in PSQA. This study recommends the plan specific calibration field must obtain to calibrate the QA devices for modulated plans.
期刊介绍:
Journal of Radiotherapy in Practice is a peer-reviewed journal covering all of the current modalities specific to clinical oncology and radiotherapy. The journal aims to publish research from a wide range of styles and encourage debate and the exchange of information and opinion from within the field of radiotherapy practice and clinical oncology. The journal also aims to encourage technical evaluations and case studies as well as equipment reviews that will be of interest to an international radiotherapy audience.