对使用 Leksell Gamma Knife Perfexion/Icon 的孕妇进行胎儿剂量评估。

IF 16.4 1区 化学 Q1 CHEMISTRY, MULTIDISCIPLINARY
Accounts of Chemical Research Pub Date : 2024-01-01 Epub Date: 2024-01-10 DOI:10.1159/000535366
Lucie Hamáčková, Josef Novotný, Gabriela Šimonová, Roman Liščák, Tomáš Chytka
{"title":"对使用 Leksell Gamma Knife Perfexion/Icon 的孕妇进行胎儿剂量评估。","authors":"Lucie Hamáčková, Josef Novotný, Gabriela Šimonová, Roman Liščák, Tomáš Chytka","doi":"10.1159/000535366","DOIUrl":null,"url":null,"abstract":"<p><strong>Introduction: </strong>It is a normal procedure to avoid the application of ionizing radiation during pregnancy. In very rare occasions, treatment can be performed, but doses to the fetus must be evaluated and reported, and the patient must sign informed consent. There can occur two types of damage caused by ionizing radiation - deterministic and stochastic effects. Deterministic effects may occur after reaching a certain threshold (100 mGy for this study); meanwhile, stochastic effects have no limit and their probability rises with dose. This study focuses on deterministic effects.</p><p><strong>Case presentations: </strong>This study compares the dose measured on phantom for the area of the pelvis and the dose measured on 3 patients with dosimeters positioned on the pelvis irradiated on Leksell Gamma Knife Perfexion/Icon. The mean dose for measurement on phantom for the pelvis was 0.73 ± 0.76 mGy, and for the patients, it was 1.28 mGy, 0.493 mGy, and 0.549 mGy which is 80 times lower, 200 times lower, and 180 times lower than the threshold for deterministic effects, respectively.</p><p><strong>Conclusion: </strong>The measurement carried on phantom served as the base for drafting informed consent and provided initial proof that treatment can be safely delivered. Measurements performed on patients only confirmed that irradiation of pregnant patients on Leksell Gamma Knife Perfexion/Icon is safe relative to the deterministic effects. Nevertheless, pregnant patients should be treated with ionizing radiation only in very extraordinary situations.</p>","PeriodicalId":1,"journal":{"name":"Accounts of Chemical Research","volume":null,"pages":null},"PeriodicalIF":16.4000,"publicationDate":"2024-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Fetal Dose Evaluation for Pregnant Patients on Leksell Gamma Knife Perfexion/Icon.\",\"authors\":\"Lucie Hamáčková, Josef Novotný, Gabriela Šimonová, Roman Liščák, Tomáš Chytka\",\"doi\":\"10.1159/000535366\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><strong>Introduction: </strong>It is a normal procedure to avoid the application of ionizing radiation during pregnancy. In very rare occasions, treatment can be performed, but doses to the fetus must be evaluated and reported, and the patient must sign informed consent. There can occur two types of damage caused by ionizing radiation - deterministic and stochastic effects. Deterministic effects may occur after reaching a certain threshold (100 mGy for this study); meanwhile, stochastic effects have no limit and their probability rises with dose. This study focuses on deterministic effects.</p><p><strong>Case presentations: </strong>This study compares the dose measured on phantom for the area of the pelvis and the dose measured on 3 patients with dosimeters positioned on the pelvis irradiated on Leksell Gamma Knife Perfexion/Icon. The mean dose for measurement on phantom for the pelvis was 0.73 ± 0.76 mGy, and for the patients, it was 1.28 mGy, 0.493 mGy, and 0.549 mGy which is 80 times lower, 200 times lower, and 180 times lower than the threshold for deterministic effects, respectively.</p><p><strong>Conclusion: </strong>The measurement carried on phantom served as the base for drafting informed consent and provided initial proof that treatment can be safely delivered. Measurements performed on patients only confirmed that irradiation of pregnant patients on Leksell Gamma Knife Perfexion/Icon is safe relative to the deterministic effects. Nevertheless, pregnant patients should be treated with ionizing radiation only in very extraordinary situations.</p>\",\"PeriodicalId\":1,\"journal\":{\"name\":\"Accounts of Chemical Research\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":16.4000,\"publicationDate\":\"2024-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Accounts of Chemical Research\",\"FirstCategoryId\":\"3\",\"ListUrlMain\":\"https://doi.org/10.1159/000535366\",\"RegionNum\":1,\"RegionCategory\":\"化学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"2024/1/10 0:00:00\",\"PubModel\":\"Epub\",\"JCR\":\"Q1\",\"JCRName\":\"CHEMISTRY, MULTIDISCIPLINARY\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Accounts of Chemical Research","FirstCategoryId":"3","ListUrlMain":"https://doi.org/10.1159/000535366","RegionNum":1,"RegionCategory":"化学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"2024/1/10 0:00:00","PubModel":"Epub","JCR":"Q1","JCRName":"CHEMISTRY, MULTIDISCIPLINARY","Score":null,"Total":0}
引用次数: 0

摘要

导言正常情况下,怀孕期间应避免使用电离辐射。在极少数情况下,可以进行治疗,但必须对胎儿所受剂量进行评估和报告,患者必须签署知情同意书。电离辐射可造成两种损害--确定性效应和随机效应。确定性效应可能在达到一定阈值(本研究为 100 mGy)后发生;而随机效应则没有阈值限制,其概率随剂量增加而上升。本研究侧重于确定性效应:本研究比较了在模型上测量的骨盆区域的剂量,以及使用 Leksell Gamma Knife Perfexion/Icon 照射骨盆的 3 位患者的剂量。在模型上测量骨盆的平均剂量为 0.73 ± 0.76 mGy,而在患者身上测量的剂量分别为 1.28 mGy、0.493 mGy 和 0.549 mGy,分别比确定性效应阈值低 80 倍、200 倍和 180 倍:在人体模型上进行的测量是起草知情同意书的基础,并初步证明了治疗是可以安全进行的。在患者身上进行的测量仅证实,相对于确定性效应,用 Leksell Gamma Knife Perfexion/Icon 对怀孕患者进行照射是安全的。不过,只有在非常特殊的情况下,孕妇才能接受电离辐射治疗。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Fetal Dose Evaluation for Pregnant Patients on Leksell Gamma Knife Perfexion/Icon.

Introduction: It is a normal procedure to avoid the application of ionizing radiation during pregnancy. In very rare occasions, treatment can be performed, but doses to the fetus must be evaluated and reported, and the patient must sign informed consent. There can occur two types of damage caused by ionizing radiation - deterministic and stochastic effects. Deterministic effects may occur after reaching a certain threshold (100 mGy for this study); meanwhile, stochastic effects have no limit and their probability rises with dose. This study focuses on deterministic effects.

Case presentations: This study compares the dose measured on phantom for the area of the pelvis and the dose measured on 3 patients with dosimeters positioned on the pelvis irradiated on Leksell Gamma Knife Perfexion/Icon. The mean dose for measurement on phantom for the pelvis was 0.73 ± 0.76 mGy, and for the patients, it was 1.28 mGy, 0.493 mGy, and 0.549 mGy which is 80 times lower, 200 times lower, and 180 times lower than the threshold for deterministic effects, respectively.

Conclusion: The measurement carried on phantom served as the base for drafting informed consent and provided initial proof that treatment can be safely delivered. Measurements performed on patients only confirmed that irradiation of pregnant patients on Leksell Gamma Knife Perfexion/Icon is safe relative to the deterministic effects. Nevertheless, pregnant patients should be treated with ionizing radiation only in very extraordinary situations.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Accounts of Chemical Research
Accounts of Chemical Research 化学-化学综合
CiteScore
31.40
自引率
1.10%
发文量
312
审稿时长
2 months
期刊介绍: Accounts of Chemical Research presents short, concise and critical articles offering easy-to-read overviews of basic research and applications in all areas of chemistry and biochemistry. These short reviews focus on research from the author’s own laboratory and are designed to teach the reader about a research project. In addition, Accounts of Chemical Research publishes commentaries that give an informed opinion on a current research problem. Special Issues online are devoted to a single topic of unusual activity and significance. Accounts of Chemical Research replaces the traditional article abstract with an article "Conspectus." These entries synopsize the research affording the reader a closer look at the content and significance of an article. Through this provision of a more detailed description of the article contents, the Conspectus enhances the article's discoverability by search engines and the exposure for the research.
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信