{"title":"使用蒙特卡洛方法分析死层对平面和同轴 Ge 探测器响应函数的影响","authors":"R.A. El-Tayebany, N. Shaaban","doi":"10.15407/jnpae2023.04.388","DOIUrl":null,"url":null,"abstract":"Germanium crystals have a dead layer that causes a decrease in efficiency since the layer is not useful for detection but strongly attenuates photons. The thickness of this inactive layer is not well known due to the existence of a transition zone where photons are increasingly absorbed. Therefore, using data provided by manufacturers in the detector simulation model, some strong discrepancies appear by changing the dead layer. Investigations into the Ge detector response functions for gamma rays have been conducted using straightforward physical mechanisms implemented by Monte Carlo simulations. The detector response function feature's most probable interaction mechanisms are described. The Monte Carlo method is applied to simulate the calibration of a HPGe detector in order to determine the total inactive germanium layer thickness and the active volume that is needed in order to study the response function for both types of detectors. Results indicated a strong impact of dead layer variations on the response function of the simulated detectors.","PeriodicalId":42588,"journal":{"name":"Nuclear Physics and Atomic Energy","volume":null,"pages":null},"PeriodicalIF":0.4000,"publicationDate":"2023-12-28","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Influence of dead layer on the response function of planar and coaxial Ge detector using Monte Carlo method\",\"authors\":\"R.A. El-Tayebany, N. Shaaban\",\"doi\":\"10.15407/jnpae2023.04.388\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Germanium crystals have a dead layer that causes a decrease in efficiency since the layer is not useful for detection but strongly attenuates photons. The thickness of this inactive layer is not well known due to the existence of a transition zone where photons are increasingly absorbed. Therefore, using data provided by manufacturers in the detector simulation model, some strong discrepancies appear by changing the dead layer. Investigations into the Ge detector response functions for gamma rays have been conducted using straightforward physical mechanisms implemented by Monte Carlo simulations. The detector response function feature's most probable interaction mechanisms are described. The Monte Carlo method is applied to simulate the calibration of a HPGe detector in order to determine the total inactive germanium layer thickness and the active volume that is needed in order to study the response function for both types of detectors. Results indicated a strong impact of dead layer variations on the response function of the simulated detectors.\",\"PeriodicalId\":42588,\"journal\":{\"name\":\"Nuclear Physics and Atomic Energy\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.4000,\"publicationDate\":\"2023-12-28\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Nuclear Physics and Atomic Energy\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.15407/jnpae2023.04.388\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"PHYSICS, NUCLEAR\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Nuclear Physics and Atomic Energy","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.15407/jnpae2023.04.388","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"PHYSICS, NUCLEAR","Score":null,"Total":0}
引用次数: 0
摘要
锗晶体有一个死层,会导致效率降低,因为该层对探测没有作用,但会强烈衰减光子。由于存在一个过渡区,光子在该区被吸收的程度越来越高,因此这种非活性层的厚度并不十分清楚。因此,在探测器模拟模型中使用制造商提供的数据时,如果改变死层,就会出现一些很大的差异。对伽马射线的 Ge 探测器响应函数的研究是通过蒙特卡罗模拟实现的直接物理机制进行的。描述了探测器响应函数特征最可能的相互作用机制。蒙特卡洛方法被应用于模拟一个 HPGe 探测器的校准,以确定总的非活性锗层厚度和活性体积,这是研究两种类型探测器的响应函数所必需的。结果表明,死层变化对模拟探测器的响应函数有很大影响。
Influence of dead layer on the response function of planar and coaxial Ge detector using Monte Carlo method
Germanium crystals have a dead layer that causes a decrease in efficiency since the layer is not useful for detection but strongly attenuates photons. The thickness of this inactive layer is not well known due to the existence of a transition zone where photons are increasingly absorbed. Therefore, using data provided by manufacturers in the detector simulation model, some strong discrepancies appear by changing the dead layer. Investigations into the Ge detector response functions for gamma rays have been conducted using straightforward physical mechanisms implemented by Monte Carlo simulations. The detector response function feature's most probable interaction mechanisms are described. The Monte Carlo method is applied to simulate the calibration of a HPGe detector in order to determine the total inactive germanium layer thickness and the active volume that is needed in order to study the response function for both types of detectors. Results indicated a strong impact of dead layer variations on the response function of the simulated detectors.
期刊介绍:
The journal Nuclear Physics and Atomic Energy presents the publications on Nuclear Physics, Atomic Energy, Radiation Physics, Radioecology, Engineering and Methods of Experiment. The journal includes peer-reviewed articles which are completed works containing new results of theoretical and experimental researches and are of interest for the scientists, postgraduate students, engineers and for the senior students.