{"title":"双TL系统x射线正电压治疗机的质量控制。","authors":"P Zaránd,&nbsp;I Polgár,&nbsp;E Katona","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>More than 40 orthovoltage X-ray therapy units are monitored by calibrated thermoluminescent dosimeters (TLD). Three pairs of LiF:Mg,Ti and CaSO4:Dy pellets are irradiated in air in a polypropane capsule. The range between 0.2 to 3.0 mm Cu the half-value layer (HVL) is determined from the CaSO4-LiF ratio. In this range, the response of the LiF dosimeters is corrected for energy dependence. We are using a fixed sensitivity for LiF in the MVL range of 2 mm Al to 0.2 mm Cu. In the first two years 173 monitorings were performed by using 485 badges. Usually not more then two different radiation qualities (in the HVL ranges 2 mm Al to 0.2 mm Cu and 0.9-2.5 mm Cu) or two beams used on a therapy unit are measured by using 2-3 badges per monitoring. In 134 cases (77.5%) the dose rates were within acceptable limits (+/- 10%). In 23 cases (13.3%) a failure of the X-ray unit was detected. We think that our system is simple enough and may be used for the dosimetric monitoring of orthovoltage X-ray therapy units. In some cases even the type of malfunction can be determined.</p>","PeriodicalId":77479,"journal":{"name":"Acta biochimica et biophysica Hungarica","volume":null,"pages":null},"PeriodicalIF":0.0000,"publicationDate":"1989-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Quality control of X-ray orthovoltage therapy units with a dual TL system.\",\"authors\":\"P Zaránd,&nbsp;I Polgár,&nbsp;E Katona\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>More than 40 orthovoltage X-ray therapy units are monitored by calibrated thermoluminescent dosimeters (TLD). Three pairs of LiF:Mg,Ti and CaSO4:Dy pellets are irradiated in air in a polypropane capsule. The range between 0.2 to 3.0 mm Cu the half-value layer (HVL) is determined from the CaSO4-LiF ratio. In this range, the response of the LiF dosimeters is corrected for energy dependence. We are using a fixed sensitivity for LiF in the MVL range of 2 mm Al to 0.2 mm Cu. In the first two years 173 monitorings were performed by using 485 badges. Usually not more then two different radiation qualities (in the HVL ranges 2 mm Al to 0.2 mm Cu and 0.9-2.5 mm Cu) or two beams used on a therapy unit are measured by using 2-3 badges per monitoring. In 134 cases (77.5%) the dose rates were within acceptable limits (+/- 10%). In 23 cases (13.3%) a failure of the X-ray unit was detected. We think that our system is simple enough and may be used for the dosimetric monitoring of orthovoltage X-ray therapy units. In some cases even the type of malfunction can be determined.</p>\",\"PeriodicalId\":77479,\"journal\":{\"name\":\"Acta biochimica et biophysica Hungarica\",\"volume\":null,\"pages\":null},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1989-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Acta biochimica et biophysica Hungarica\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Acta biochimica et biophysica Hungarica","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

通过校准的热释光剂量计(TLD)监测40多个正电压x射线治疗单元。三对LiF:Mg,Ti和CaSO4:Dy颗粒在丙烷胶囊中在空气中辐照。0.2 ~ 3.0 mm Cu半值层(HVL)的范围由CaSO4-LiF比值决定。在这个范围内,LiF剂量计的响应根据能量依赖进行了校正。我们在MVL范围内使用固定的LiF灵敏度为2mm Al到0.2 mm Cu。在头两年,使用485个徽章进行了173次监测。通常不超过两种不同的辐射质量(在HVL范围内为2mm Al至0.2 mm Cu和0.9-2.5 mm Cu)或治疗单元上使用的两束射线,每次监测使用2-3个胸卡。134例(77.5%)剂量率在可接受范围内(+/- 10%)。在23例(13.3%)病例中,检测到x射线单元故障。我们认为我们的系统足够简单,可以用于正电压x射线治疗单元的剂量学监测。在某些情况下,甚至可以确定故障的类型。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Quality control of X-ray orthovoltage therapy units with a dual TL system.

More than 40 orthovoltage X-ray therapy units are monitored by calibrated thermoluminescent dosimeters (TLD). Three pairs of LiF:Mg,Ti and CaSO4:Dy pellets are irradiated in air in a polypropane capsule. The range between 0.2 to 3.0 mm Cu the half-value layer (HVL) is determined from the CaSO4-LiF ratio. In this range, the response of the LiF dosimeters is corrected for energy dependence. We are using a fixed sensitivity for LiF in the MVL range of 2 mm Al to 0.2 mm Cu. In the first two years 173 monitorings were performed by using 485 badges. Usually not more then two different radiation qualities (in the HVL ranges 2 mm Al to 0.2 mm Cu and 0.9-2.5 mm Cu) or two beams used on a therapy unit are measured by using 2-3 badges per monitoring. In 134 cases (77.5%) the dose rates were within acceptable limits (+/- 10%). In 23 cases (13.3%) a failure of the X-ray unit was detected. We think that our system is simple enough and may be used for the dosimetric monitoring of orthovoltage X-ray therapy units. In some cases even the type of malfunction can be determined.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信