研究有或无颞下颌疾病受试者咀嚼肌超声测量的可靠性:一项内部可靠性研究

Q4 Medicine
Leila Alizadeh, Khaled Rezaie, Vahid Sobhani, Ali Amiri, Mansoureh Samadi, Hossein Shakeri, Saman Salehi
{"title":"研究有或无颞下颌疾病受试者咀嚼肌超声测量的可靠性:一项内部可靠性研究","authors":"Leila Alizadeh, Khaled Rezaie, Vahid Sobhani, Ali Amiri, Mansoureh Samadi, Hossein Shakeri, Saman Salehi","doi":"10.18502/jmr.v17i4.13896","DOIUrl":null,"url":null,"abstract":"Introduction: Ultrasonography is a common tool for assessing muscle morphology. This study aims to investigate the intra-rater reliability of ultrasonography measurements of the masseter and lateral pterygoid muscles (LPM) in subjects with and without temporomandibular disorder (TMD) and to compare the measurements between the two groups.
 Materials and Methods: A total of 30 individuals (15 subjects with TMD and 15 subjects without TMD) participated in the study. Imaging of the masseter muscle at rest and contraction, in addition to LPM, was performed using sonography in 2 sessions with a 2-day interval and the thickness of these muscles was measured. To assess the reliability, the intraclass correlation coefficient (ICC(3, 3)), standard error of measurement, minimal detectable change, and limits of agreement were used. Additionally, for the between-group comparison of measurements, the independent sample t-test was used. The Cohen d was used to determine the size of the differences.
 Results: ICC(3, 3) for intra-rater reliability of masseter thickness in rest and clenched positions and LPM were 0.89, 0.80, and 0.86 in the asymptomatic group, and 0.88, 0.80, and 0.90 in the TMD group, respectively. The values of absolute reliability parameters, such as standard error of measurement and minimal detectable change were in a small range. Additionally, compared to the asymptomatic group, the TMD group had thicker masseter muscle in the rest position and the effect size revealed a large difference between the groups. No significant differences were found between the groups for the thickness of the masseter muscle in contraction position and LPM.
 Conclusion: The results of the intra-reliability study revealed that ultrasonography is a reliable method to measure the thickness of the masticatory muscles in TMD subjects and asymptomatic subjects. Additionally, TMDs can change the thickness of the masticatory muscles.","PeriodicalId":34281,"journal":{"name":"Journal of Modern Rehabilitation","volume":"56 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2023-10-17","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Investigating the Reliability of Ultrasonography Measurements of Masticatory Muscles in subjects with and without Temporomandibular Disorders: An Intra-rater Reliability Study\",\"authors\":\"Leila Alizadeh, Khaled Rezaie, Vahid Sobhani, Ali Amiri, Mansoureh Samadi, Hossein Shakeri, Saman Salehi\",\"doi\":\"10.18502/jmr.v17i4.13896\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Introduction: Ultrasonography is a common tool for assessing muscle morphology. This study aims to investigate the intra-rater reliability of ultrasonography measurements of the masseter and lateral pterygoid muscles (LPM) in subjects with and without temporomandibular disorder (TMD) and to compare the measurements between the two groups.
 Materials and Methods: A total of 30 individuals (15 subjects with TMD and 15 subjects without TMD) participated in the study. Imaging of the masseter muscle at rest and contraction, in addition to LPM, was performed using sonography in 2 sessions with a 2-day interval and the thickness of these muscles was measured. To assess the reliability, the intraclass correlation coefficient (ICC(3, 3)), standard error of measurement, minimal detectable change, and limits of agreement were used. Additionally, for the between-group comparison of measurements, the independent sample t-test was used. The Cohen d was used to determine the size of the differences.
 Results: ICC(3, 3) for intra-rater reliability of masseter thickness in rest and clenched positions and LPM were 0.89, 0.80, and 0.86 in the asymptomatic group, and 0.88, 0.80, and 0.90 in the TMD group, respectively. The values of absolute reliability parameters, such as standard error of measurement and minimal detectable change were in a small range. Additionally, compared to the asymptomatic group, the TMD group had thicker masseter muscle in the rest position and the effect size revealed a large difference between the groups. No significant differences were found between the groups for the thickness of the masseter muscle in contraction position and LPM.
 Conclusion: The results of the intra-reliability study revealed that ultrasonography is a reliable method to measure the thickness of the masticatory muscles in TMD subjects and asymptomatic subjects. Additionally, TMDs can change the thickness of the masticatory muscles.\",\"PeriodicalId\":34281,\"journal\":{\"name\":\"Journal of Modern Rehabilitation\",\"volume\":\"56 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2023-10-17\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Modern Rehabilitation\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.18502/jmr.v17i4.13896\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"Medicine\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Modern Rehabilitation","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.18502/jmr.v17i4.13896","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"Medicine","Score":null,"Total":0}
引用次数: 0

摘要

超声检查是评估肌肉形态的常用工具。本研究旨在探讨有和无颞下颌疾病(TMD)受试者咬肌和翼状外侧肌(LPM)超声测量的内部可靠性,并比较两组之间的测量结果。 材料与方法:共30例受试者(TMD患者15例,非TMD患者15例)参与研究。除LPM外,在休息和收缩时使用超声对咬肌进行成像,每隔2天进行2次,并测量这些肌肉的厚度。为了评估可靠性,使用了类内相关系数(ICC(3,3))、测量标准误差、最小可检测变化和一致性限。此外,对于测量值的组间比较,使用独立样本t检验。科恩d用于确定差异的大小。 结果:无症状组的咬肌休息位、握紧位厚度和LPM的内信度ICC(3,3)分别为0.89、0.80、0.86,TMD组的内信度ICC(3,3)分别为0.88、0.80、0.90。测量标准误差、最小可检测变化等绝对可靠性参数的取值范围较小。此外,与无症状组相比,TMD组在休息位置的咬肌更厚,效应量显示组间差异较大。两组间咬肌收缩位厚度及LPM均无显著差异。 结论:可靠性内研究结果表明,超声是测量TMD患者和无症状患者咀嚼肌厚度的可靠方法。此外,tmd可以改变咀嚼肌的厚度。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Investigating the Reliability of Ultrasonography Measurements of Masticatory Muscles in subjects with and without Temporomandibular Disorders: An Intra-rater Reliability Study
Introduction: Ultrasonography is a common tool for assessing muscle morphology. This study aims to investigate the intra-rater reliability of ultrasonography measurements of the masseter and lateral pterygoid muscles (LPM) in subjects with and without temporomandibular disorder (TMD) and to compare the measurements between the two groups. Materials and Methods: A total of 30 individuals (15 subjects with TMD and 15 subjects without TMD) participated in the study. Imaging of the masseter muscle at rest and contraction, in addition to LPM, was performed using sonography in 2 sessions with a 2-day interval and the thickness of these muscles was measured. To assess the reliability, the intraclass correlation coefficient (ICC(3, 3)), standard error of measurement, minimal detectable change, and limits of agreement were used. Additionally, for the between-group comparison of measurements, the independent sample t-test was used. The Cohen d was used to determine the size of the differences. Results: ICC(3, 3) for intra-rater reliability of masseter thickness in rest and clenched positions and LPM were 0.89, 0.80, and 0.86 in the asymptomatic group, and 0.88, 0.80, and 0.90 in the TMD group, respectively. The values of absolute reliability parameters, such as standard error of measurement and minimal detectable change were in a small range. Additionally, compared to the asymptomatic group, the TMD group had thicker masseter muscle in the rest position and the effect size revealed a large difference between the groups. No significant differences were found between the groups for the thickness of the masseter muscle in contraction position and LPM. Conclusion: The results of the intra-reliability study revealed that ultrasonography is a reliable method to measure the thickness of the masticatory muscles in TMD subjects and asymptomatic subjects. Additionally, TMDs can change the thickness of the masticatory muscles.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
Journal of Modern Rehabilitation
Journal of Modern Rehabilitation Medicine-Rehabilitation
CiteScore
0.50
自引率
0.00%
发文量
44
审稿时长
12 weeks
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信