{"title":"经导管二尖瓣主动脉瓣置换术中预扩张球囊破裂:分类、治疗策略和预防","authors":"Xinlei Wu, Tianbo Wu, Rutao Wang, Ahmed Elkoumy, Daozhu Wu, Osama Soliman, Xinmin Zhang, Lianpin Wu","doi":"10.15212/cvia.2023.0063","DOIUrl":null,"url":null,"abstract":"Balloon rupture during transcatheter aortic valve replacement (TAVR) is a rare but serious complication. Here, we present two cases of balloon rupture in patients with severe aortic stenosis and type 0 bicuspid aortic valves. Three-dimensional models based on pre-procedure cardiac CT angiography were used to investigate these cases post hoc. The models revealed asymmetrical distribution of calcifications with sharply spiked features in the bicuspid aortic valves. The narrow calcified orifices resulted in uneven force distribution on the expanded balloon, thus leading to balloon rupture. We additionally review the classification and causes of balloon rupture, summarize methods for avoiding complications, and describe treatment options. Accurate pre-procedural anatomy evaluation and computer modeling are crucial for planning and managing TAVR procedures. Further investigation through computer simulation is necessary to determine the appropriate balloon size and inflation locations, to provide a reference for pre-procedural preparation.","PeriodicalId":41559,"journal":{"name":"Cardiovascular Innovations and Applications","volume":"58 1","pages":"0"},"PeriodicalIF":0.9000,"publicationDate":"2023-01-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Balloon Rupture during Pre-Dilation for Transcatheter Aortic Valve Replacement in Patients with a Bicuspid Aortic Valve: Classification, Treatment Strategies, and Prevention\",\"authors\":\"Xinlei Wu, Tianbo Wu, Rutao Wang, Ahmed Elkoumy, Daozhu Wu, Osama Soliman, Xinmin Zhang, Lianpin Wu\",\"doi\":\"10.15212/cvia.2023.0063\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Balloon rupture during transcatheter aortic valve replacement (TAVR) is a rare but serious complication. Here, we present two cases of balloon rupture in patients with severe aortic stenosis and type 0 bicuspid aortic valves. Three-dimensional models based on pre-procedure cardiac CT angiography were used to investigate these cases post hoc. The models revealed asymmetrical distribution of calcifications with sharply spiked features in the bicuspid aortic valves. The narrow calcified orifices resulted in uneven force distribution on the expanded balloon, thus leading to balloon rupture. We additionally review the classification and causes of balloon rupture, summarize methods for avoiding complications, and describe treatment options. Accurate pre-procedural anatomy evaluation and computer modeling are crucial for planning and managing TAVR procedures. Further investigation through computer simulation is necessary to determine the appropriate balloon size and inflation locations, to provide a reference for pre-procedural preparation.\",\"PeriodicalId\":41559,\"journal\":{\"name\":\"Cardiovascular Innovations and Applications\",\"volume\":\"58 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.9000,\"publicationDate\":\"2023-01-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Cardiovascular Innovations and Applications\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.15212/cvia.2023.0063\",\"RegionNum\":4,\"RegionCategory\":\"医学\",\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"Q4\",\"JCRName\":\"CARDIAC & CARDIOVASCULAR SYSTEMS\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cardiovascular Innovations and Applications","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.15212/cvia.2023.0063","RegionNum":4,"RegionCategory":"医学","ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"Q4","JCRName":"CARDIAC & CARDIOVASCULAR SYSTEMS","Score":null,"Total":0}
Balloon Rupture during Pre-Dilation for Transcatheter Aortic Valve Replacement in Patients with a Bicuspid Aortic Valve: Classification, Treatment Strategies, and Prevention
Balloon rupture during transcatheter aortic valve replacement (TAVR) is a rare but serious complication. Here, we present two cases of balloon rupture in patients with severe aortic stenosis and type 0 bicuspid aortic valves. Three-dimensional models based on pre-procedure cardiac CT angiography were used to investigate these cases post hoc. The models revealed asymmetrical distribution of calcifications with sharply spiked features in the bicuspid aortic valves. The narrow calcified orifices resulted in uneven force distribution on the expanded balloon, thus leading to balloon rupture. We additionally review the classification and causes of balloon rupture, summarize methods for avoiding complications, and describe treatment options. Accurate pre-procedural anatomy evaluation and computer modeling are crucial for planning and managing TAVR procedures. Further investigation through computer simulation is necessary to determine the appropriate balloon size and inflation locations, to provide a reference for pre-procedural preparation.