{"title":"三维USCT数据采集过程中患者运动分析","authors":"N. Ruiter, T. Hopp, M. Zapf, E. Kretzek, H. Gemmeke","doi":"10.1117/12.2216680","DOIUrl":null,"url":null,"abstract":"In our first clinical study with a full 3D Ultrasound Computer Tomography (USCT) system patient data was acquired in eight minutes for one breast. In this paper the patient movement during the acquisition was analyzed quantitatively and as far as possible corrected in the resulting images. The movement was tracked in ten successive reflectivity reconstructions of full breast volumes acquired during 10 s intervals at different aperture positions, which were separated by 41 s intervals. The mean distance between initial and final position was 2.2 mm (standard deviation (STD) ± 0.9 mm, max. 4.1 mm, min. 0.8 mm) and the average sum of all moved distances was 4.9 mm (STD ± 1.9 mm, max. 8.8 mm, min. 2.7 mm). The tracked movement was corrected by summing successive images, which were transformed according to the detected movement. The contrast of these images increased and additional image content became visible.","PeriodicalId":228011,"journal":{"name":"SPIE Medical Imaging","volume":"23 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2016-04-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"10","resultStr":"{\"title\":\"Analysis of patient movement during 3D USCT data acquisition\",\"authors\":\"N. Ruiter, T. Hopp, M. Zapf, E. Kretzek, H. Gemmeke\",\"doi\":\"10.1117/12.2216680\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"In our first clinical study with a full 3D Ultrasound Computer Tomography (USCT) system patient data was acquired in eight minutes for one breast. In this paper the patient movement during the acquisition was analyzed quantitatively and as far as possible corrected in the resulting images. The movement was tracked in ten successive reflectivity reconstructions of full breast volumes acquired during 10 s intervals at different aperture positions, which were separated by 41 s intervals. The mean distance between initial and final position was 2.2 mm (standard deviation (STD) ± 0.9 mm, max. 4.1 mm, min. 0.8 mm) and the average sum of all moved distances was 4.9 mm (STD ± 1.9 mm, max. 8.8 mm, min. 2.7 mm). The tracked movement was corrected by summing successive images, which were transformed according to the detected movement. The contrast of these images increased and additional image content became visible.\",\"PeriodicalId\":228011,\"journal\":{\"name\":\"SPIE Medical Imaging\",\"volume\":\"23 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2016-04-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"10\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"SPIE Medical Imaging\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.1117/12.2216680\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"SPIE Medical Imaging","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.1117/12.2216680","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 10
摘要
在我们的第一个临床研究中,使用完整的3D超声计算机断层扫描(USCT)系统在8分钟内获得了一个乳房的患者数据。本文对采集过程中患者的运动进行了定量分析,并在得到的图像中尽可能地进行了校正。以41 s为间隔,以10 s为间隔,在不同的孔径位置连续10次获得全乳房体积的反射率重建。初始位置与最终位置之间的平均距离为2.2 mm(标准偏差(STD)±0.9 mm,最大;4.1 mm,最小0.8 mm),所有移动距离的平均总和为4.9 mm (STD±1.9 mm,最大。8.8毫米,最小2.7毫米)。根据检测到的运动进行变换,将连续图像相加,对跟踪到的运动进行校正。这些图像的对比度增加了,并且可以看到额外的图像内容。
Analysis of patient movement during 3D USCT data acquisition
In our first clinical study with a full 3D Ultrasound Computer Tomography (USCT) system patient data was acquired in eight minutes for one breast. In this paper the patient movement during the acquisition was analyzed quantitatively and as far as possible corrected in the resulting images. The movement was tracked in ten successive reflectivity reconstructions of full breast volumes acquired during 10 s intervals at different aperture positions, which were separated by 41 s intervals. The mean distance between initial and final position was 2.2 mm (standard deviation (STD) ± 0.9 mm, max. 4.1 mm, min. 0.8 mm) and the average sum of all moved distances was 4.9 mm (STD ± 1.9 mm, max. 8.8 mm, min. 2.7 mm). The tracked movement was corrected by summing successive images, which were transformed according to the detected movement. The contrast of these images increased and additional image content became visible.