{"title":"化学发光免疫测定法、放射性测定法和液相色谱串联质谱法检测血清总睾酮:比较研究","authors":"Fenghua Chen","doi":"10.31031/cjmi.2021.04.000598","DOIUrl":null,"url":null,"abstract":"Aim: LC-MS/MS is still considered the gold standard for detecting steroid hormones in clinic. The aim of this study was to compare the analytical sensitivity of chemiluminescence immunoassays (CLIA) platforms, radioimmunoassay (RIA) platform, and high-performance liquid-chromatography tandem mass spectrometry (LC-MS/MS) platform for detecting total testosterone (tTES) in human samples (males and females). Method: Sixty-nine healthy male serum samples and 114 female serum samples were collected. The serum testosterone was detected by automatic immunoassay systems of Beckman and SIEMENS, radioimmunoassay, and LC-MS/MS. Pearson’s correlation coefficients were calculated to assess the agreement of other platforms with LC-MS/MS. Result: When evaluating TES in males, the results detected by SIEMENS2000 had a good correlation with LC-MS/MS (r=0.986), followed by Beckman DXI800, while RIA showed a weak correlation. There was no statistical difference between the results detected by SIEMENS2000 compared to those detected by DXI800 (p=0.87). When evaluating TES in females, DXI800 showed a good correlation with LC-MS/ MS (r=0.8784). SIEMENS2000 had most of the detected values below its low limit, and the SIEMENS2000 results with definite values were not statistically different compared to DXI800 results. Conclusion: The results detection platforms (SIEMENS2000 and DXI800) both had a better correlation with those of LC-MS/MS than RIA. Yet, Siemens 2000 had better consistency with LC-MS/MS in detecting the serums with low tTES value than DXI800. the samples with definite values; were above the lowest limit of Simense (0.69nmom/L). The samples’ values the RIA obviously The correlation coefficient between the two immunoassay platforms and the LC-MS/MS was higher than that between RIA and LC-MS/MS in males. We further analyzed the samples’ results below 0.69nmol/L when using two chemiluminescence platforms.","PeriodicalId":406162,"journal":{"name":"Cohesive Journal of Microbiology & Infectious Disease","volume":"78 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2021-02-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Detecting Total Serum Testosterone by Chemiluminescence Immunoassays, Radioactive Assays, and Liquid-Chromatography Tandem Mass Spectrometry: A Comparative Study\",\"authors\":\"Fenghua Chen\",\"doi\":\"10.31031/cjmi.2021.04.000598\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Aim: LC-MS/MS is still considered the gold standard for detecting steroid hormones in clinic. The aim of this study was to compare the analytical sensitivity of chemiluminescence immunoassays (CLIA) platforms, radioimmunoassay (RIA) platform, and high-performance liquid-chromatography tandem mass spectrometry (LC-MS/MS) platform for detecting total testosterone (tTES) in human samples (males and females). Method: Sixty-nine healthy male serum samples and 114 female serum samples were collected. The serum testosterone was detected by automatic immunoassay systems of Beckman and SIEMENS, radioimmunoassay, and LC-MS/MS. Pearson’s correlation coefficients were calculated to assess the agreement of other platforms with LC-MS/MS. Result: When evaluating TES in males, the results detected by SIEMENS2000 had a good correlation with LC-MS/MS (r=0.986), followed by Beckman DXI800, while RIA showed a weak correlation. There was no statistical difference between the results detected by SIEMENS2000 compared to those detected by DXI800 (p=0.87). When evaluating TES in females, DXI800 showed a good correlation with LC-MS/ MS (r=0.8784). SIEMENS2000 had most of the detected values below its low limit, and the SIEMENS2000 results with definite values were not statistically different compared to DXI800 results. Conclusion: The results detection platforms (SIEMENS2000 and DXI800) both had a better correlation with those of LC-MS/MS than RIA. Yet, Siemens 2000 had better consistency with LC-MS/MS in detecting the serums with low tTES value than DXI800. the samples with definite values; were above the lowest limit of Simense (0.69nmom/L). The samples’ values the RIA obviously The correlation coefficient between the two immunoassay platforms and the LC-MS/MS was higher than that between RIA and LC-MS/MS in males. We further analyzed the samples’ results below 0.69nmol/L when using two chemiluminescence platforms.\",\"PeriodicalId\":406162,\"journal\":{\"name\":\"Cohesive Journal of Microbiology & Infectious Disease\",\"volume\":\"78 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2021-02-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Cohesive Journal of Microbiology & Infectious Disease\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.31031/cjmi.2021.04.000598\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Cohesive Journal of Microbiology & Infectious Disease","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.31031/cjmi.2021.04.000598","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:LC-MS/MS在临床上仍被认为是检测类固醇激素的金标准。本研究的目的是比较化学发光免疫测定(CLIA)平台、放射免疫测定(RIA)平台和高效液相色谱串联质谱(LC-MS/MS)平台检测人(男性和女性)样品中总睾酮(tTES)的分析灵敏度。方法:采集健康男性血清69份,女性血清114份。采用Beckman和SIEMENS自动免疫分析系统、放射免疫分析系统和LC-MS/MS检测血清睾酮。计算Pearson相关系数以评估其他平台与LC-MS/MS的一致性。结果:在评价男性TES时,SIEMENS2000检测结果与LC-MS/MS具有较好的相关性(r=0.986),其次是Beckman DXI800检测结果,RIA检测结果相关性较弱。SIEMENS2000检测结果与DXI800检测结果无统计学差异(p=0.87)。在评价女性TES时,DXI800与LC-MS/ MS具有良好的相关性(r=0.8784)。SIEMENS2000的大部分检测值都低于其下限,具有确定值的SIEMENS2000的结果与DXI800的结果没有统计学差异。结论:检测平台SIEMENS2000和DXI800与LC-MS/MS的相关性均优于RIA。而Siemens 2000与LC-MS/MS检测低tTES值血清的一致性优于DXI800。具有确定值的样本;均高于siense最低限量(0.69nmom/L)。两种免疫分析平台与LC-MS/MS的相关系数在男性中高于RIA与LC-MS/MS的相关系数。我们进一步分析了使用两种化学发光平台时样品在0.69nmol/L以下的结果。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Detecting Total Serum Testosterone by Chemiluminescence Immunoassays, Radioactive Assays, and Liquid-Chromatography Tandem Mass Spectrometry: A Comparative Study
Aim: LC-MS/MS is still considered the gold standard for detecting steroid hormones in clinic. The aim of this study was to compare the analytical sensitivity of chemiluminescence immunoassays (CLIA) platforms, radioimmunoassay (RIA) platform, and high-performance liquid-chromatography tandem mass spectrometry (LC-MS/MS) platform for detecting total testosterone (tTES) in human samples (males and females). Method: Sixty-nine healthy male serum samples and 114 female serum samples were collected. The serum testosterone was detected by automatic immunoassay systems of Beckman and SIEMENS, radioimmunoassay, and LC-MS/MS. Pearson’s correlation coefficients were calculated to assess the agreement of other platforms with LC-MS/MS. Result: When evaluating TES in males, the results detected by SIEMENS2000 had a good correlation with LC-MS/MS (r=0.986), followed by Beckman DXI800, while RIA showed a weak correlation. There was no statistical difference between the results detected by SIEMENS2000 compared to those detected by DXI800 (p=0.87). When evaluating TES in females, DXI800 showed a good correlation with LC-MS/ MS (r=0.8784). SIEMENS2000 had most of the detected values below its low limit, and the SIEMENS2000 results with definite values were not statistically different compared to DXI800 results. Conclusion: The results detection platforms (SIEMENS2000 and DXI800) both had a better correlation with those of LC-MS/MS than RIA. Yet, Siemens 2000 had better consistency with LC-MS/MS in detecting the serums with low tTES value than DXI800. the samples with definite values; were above the lowest limit of Simense (0.69nmom/L). The samples’ values the RIA obviously The correlation coefficient between the two immunoassay platforms and the LC-MS/MS was higher than that between RIA and LC-MS/MS in males. We further analyzed the samples’ results below 0.69nmol/L when using two chemiluminescence platforms.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信