{"title":"PC无线显示中央蠕动输液液滴传感器精度分析(TCRT5000液滴传感器)","authors":"Hanna Firdaus, B. G. Irianto, Sumber, Jing Lu","doi":"10.35882/jeeemi.v4i1.5","DOIUrl":null,"url":null,"abstract":"In some hospitals the infusion is still done manually, medical staff observes fluid drip directly and then controls its rate using a mechanical resistor (clamp), this method is certainly far from the level of accuracy. Infusion pump is a medical aid that has functions to control and ensure the correct dose of infusion fluid that is given to patients under treatment. The purpose of this study is to analyze the accuracy of the TCRT5000 as a drop sensor, based on readings of the infusion pump monitoring system. This module consists of a TCRT5000 drop sensor module, comparator circuit, monostable circuit, stepper motor, L298N motor driver, and ATmega328 microcontroller. The droplets are detected by the TCRT 5000 sensor, then amplified by a comparator and monostable circuit, then the flow rate and remaining volume readings are generated by the ATmega328 microcontroller. Furthermore, this data is sent to the Personal Computer (PC) via wireless HC-11. The results of the flow rate module measurement show that the highest error value is 4% at the 30 ml/hour setting, and the lowest error value is 1% at the 60 ml/hour setting. While the results of the flow rate measurement using an Infuse Device Analyzer, the highest error value is 2,2% at the 30 ml/hour setting, and the lowest error value is 0,58% at the 100 ml/hour setting. This infusion pump monitoring is designed centrally to facilitate the nurse's task in monitoring the infusion dose accurately that is given to the patient.","PeriodicalId":369032,"journal":{"name":"Journal of Electronics, Electromedical Engineering, and Medical Informatics","volume":"105 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2022-01-16","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"8","resultStr":"{\"title\":\"Analysis of the Drop Sensors Accuracy in Central Peristaltic Infusion Monitoring Displayed on PC Based Wireless (TCRT5000 Drop Sensor)\",\"authors\":\"Hanna Firdaus, B. G. Irianto, Sumber, Jing Lu\",\"doi\":\"10.35882/jeeemi.v4i1.5\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"In some hospitals the infusion is still done manually, medical staff observes fluid drip directly and then controls its rate using a mechanical resistor (clamp), this method is certainly far from the level of accuracy. Infusion pump is a medical aid that has functions to control and ensure the correct dose of infusion fluid that is given to patients under treatment. The purpose of this study is to analyze the accuracy of the TCRT5000 as a drop sensor, based on readings of the infusion pump monitoring system. This module consists of a TCRT5000 drop sensor module, comparator circuit, monostable circuit, stepper motor, L298N motor driver, and ATmega328 microcontroller. The droplets are detected by the TCRT 5000 sensor, then amplified by a comparator and monostable circuit, then the flow rate and remaining volume readings are generated by the ATmega328 microcontroller. Furthermore, this data is sent to the Personal Computer (PC) via wireless HC-11. The results of the flow rate module measurement show that the highest error value is 4% at the 30 ml/hour setting, and the lowest error value is 1% at the 60 ml/hour setting. While the results of the flow rate measurement using an Infuse Device Analyzer, the highest error value is 2,2% at the 30 ml/hour setting, and the lowest error value is 0,58% at the 100 ml/hour setting. This infusion pump monitoring is designed centrally to facilitate the nurse's task in monitoring the infusion dose accurately that is given to the patient.\",\"PeriodicalId\":369032,\"journal\":{\"name\":\"Journal of Electronics, Electromedical Engineering, and Medical Informatics\",\"volume\":\"105 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2022-01-16\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"8\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of Electronics, Electromedical Engineering, and Medical Informatics\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.35882/jeeemi.v4i1.5\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of Electronics, Electromedical Engineering, and Medical Informatics","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.35882/jeeemi.v4i1.5","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 8

摘要

有些医院仍采用人工输液的方式,医务人员直接观察液体滴下,然后用机械电阻(钳)控制滴下速率,这种方法的准确性显然还差得远。输液泵是一种医疗辅助设备,具有控制和保证给病人正确的输液剂量的功能。本研究的目的是根据输液泵监测系统的读数,分析TCRT5000作为液滴传感器的准确性。该模块由TCRT5000跌落传感器模块、比较器电路、单稳电路、步进电机、L298N电机驱动器和ATmega328微控制器组成。液滴由TCRT 5000传感器检测,然后由比较器和单稳电路放大,然后由ATmega328微控制器生成流速和剩余体积读数。此外,这些数据通过无线HC-11发送到个人电脑(PC)。流速模块测量结果表明,在30 ml/小时设置时,误差值最高为4%,在60 ml/小时设置时,误差值最低为1%。而使用inuse Device Analyzer测量流速的结果,在30 ml/小时设置时,最大误差值为2.2%,在100 ml/小时设置时,最小误差值为0.58%。这种输液泵监测是集中设计的,以方便护士准确监测给病人的输液剂量。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Analysis of the Drop Sensors Accuracy in Central Peristaltic Infusion Monitoring Displayed on PC Based Wireless (TCRT5000 Drop Sensor)
In some hospitals the infusion is still done manually, medical staff observes fluid drip directly and then controls its rate using a mechanical resistor (clamp), this method is certainly far from the level of accuracy. Infusion pump is a medical aid that has functions to control and ensure the correct dose of infusion fluid that is given to patients under treatment. The purpose of this study is to analyze the accuracy of the TCRT5000 as a drop sensor, based on readings of the infusion pump monitoring system. This module consists of a TCRT5000 drop sensor module, comparator circuit, monostable circuit, stepper motor, L298N motor driver, and ATmega328 microcontroller. The droplets are detected by the TCRT 5000 sensor, then amplified by a comparator and monostable circuit, then the flow rate and remaining volume readings are generated by the ATmega328 microcontroller. Furthermore, this data is sent to the Personal Computer (PC) via wireless HC-11. The results of the flow rate module measurement show that the highest error value is 4% at the 30 ml/hour setting, and the lowest error value is 1% at the 60 ml/hour setting. While the results of the flow rate measurement using an Infuse Device Analyzer, the highest error value is 2,2% at the 30 ml/hour setting, and the lowest error value is 0,58% at the 100 ml/hour setting. This infusion pump monitoring is designed centrally to facilitate the nurse's task in monitoring the infusion dose accurately that is given to the patient.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信