{"title":"血清中甲氨蝶呤的测定:aca MTHO免疫法与高效液相色谱法的比较。","authors":"W Martz, M Oellerich, G Schumann","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The determination of methotrexate on the aca clinical analyser (aca MTHO method) was evaluated and the method compared with a fluorescence polarisation immunoassay (FPIA) and a specific high-performance liquid chromatography (HPLC) procedure. The within series and the between days coefficients of variation of measurements in human pool serum spiked with methotrexate (0.07 mumol/l to 15 mumol/l) ranges from 11.7% to 2.5% and 14.0% to 5.8%, respectively. The calibration was found to be stable for 9 weeks. There was a good correlation between the results of the MTHO method when compared with FPIA and HPLC. At low methotrexate concentration, the results of HPLC were on average 11% and 14% lower than those obtained by the MTHO assay and FPIA respectively.</p>","PeriodicalId":15649,"journal":{"name":"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie","volume":"28 12","pages":"951-3"},"PeriodicalIF":0.0000,"publicationDate":"1990-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Determination of methotrexate in serum: the aca MTHO immunoassay and HPLC compared.\",\"authors\":\"W Martz, M Oellerich, G Schumann\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The determination of methotrexate on the aca clinical analyser (aca MTHO method) was evaluated and the method compared with a fluorescence polarisation immunoassay (FPIA) and a specific high-performance liquid chromatography (HPLC) procedure. The within series and the between days coefficients of variation of measurements in human pool serum spiked with methotrexate (0.07 mumol/l to 15 mumol/l) ranges from 11.7% to 2.5% and 14.0% to 5.8%, respectively. The calibration was found to be stable for 9 weeks. There was a good correlation between the results of the MTHO method when compared with FPIA and HPLC. At low methotrexate concentration, the results of HPLC were on average 11% and 14% lower than those obtained by the MTHO assay and FPIA respectively.</p>\",\"PeriodicalId\":15649,\"journal\":{\"name\":\"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie\",\"volume\":\"28 12\",\"pages\":\"951-3\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1990-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
Determination of methotrexate in serum: the aca MTHO immunoassay and HPLC compared.
The determination of methotrexate on the aca clinical analyser (aca MTHO method) was evaluated and the method compared with a fluorescence polarisation immunoassay (FPIA) and a specific high-performance liquid chromatography (HPLC) procedure. The within series and the between days coefficients of variation of measurements in human pool serum spiked with methotrexate (0.07 mumol/l to 15 mumol/l) ranges from 11.7% to 2.5% and 14.0% to 5.8%, respectively. The calibration was found to be stable for 9 weeks. There was a good correlation between the results of the MTHO method when compared with FPIA and HPLC. At low methotrexate concentration, the results of HPLC were on average 11% and 14% lower than those obtained by the MTHO assay and FPIA respectively.