血清中甲氨蝶呤的测定:aca MTHO免疫法与高效液相色谱法的比较。

W Martz, M Oellerich, G Schumann
{"title":"血清中甲氨蝶呤的测定:aca MTHO免疫法与高效液相色谱法的比较。","authors":"W Martz,&nbsp;M Oellerich,&nbsp;G Schumann","doi":"","DOIUrl":null,"url":null,"abstract":"<p><p>The determination of methotrexate on the aca clinical analyser (aca MTHO method) was evaluated and the method compared with a fluorescence polarisation immunoassay (FPIA) and a specific high-performance liquid chromatography (HPLC) procedure. The within series and the between days coefficients of variation of measurements in human pool serum spiked with methotrexate (0.07 mumol/l to 15 mumol/l) ranges from 11.7% to 2.5% and 14.0% to 5.8%, respectively. The calibration was found to be stable for 9 weeks. There was a good correlation between the results of the MTHO method when compared with FPIA and HPLC. At low methotrexate concentration, the results of HPLC were on average 11% and 14% lower than those obtained by the MTHO assay and FPIA respectively.</p>","PeriodicalId":15649,"journal":{"name":"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie","volume":"28 12","pages":"951-3"},"PeriodicalIF":0.0000,"publicationDate":"1990-12-01","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Determination of methotrexate in serum: the aca MTHO immunoassay and HPLC compared.\",\"authors\":\"W Martz,&nbsp;M Oellerich,&nbsp;G Schumann\",\"doi\":\"\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"<p><p>The determination of methotrexate on the aca clinical analyser (aca MTHO method) was evaluated and the method compared with a fluorescence polarisation immunoassay (FPIA) and a specific high-performance liquid chromatography (HPLC) procedure. The within series and the between days coefficients of variation of measurements in human pool serum spiked with methotrexate (0.07 mumol/l to 15 mumol/l) ranges from 11.7% to 2.5% and 14.0% to 5.8%, respectively. The calibration was found to be stable for 9 weeks. There was a good correlation between the results of the MTHO method when compared with FPIA and HPLC. At low methotrexate concentration, the results of HPLC were on average 11% and 14% lower than those obtained by the MTHO assay and FPIA respectively.</p>\",\"PeriodicalId\":15649,\"journal\":{\"name\":\"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie\",\"volume\":\"28 12\",\"pages\":\"951-3\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"1990-12-01\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie","FirstCategoryId":"1085","ListUrlMain":"","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

对aca临床分析仪(aca MTHO法)测定甲氨蝶呤进行了评价,并与荧光偏振免疫分析法(FPIA)和高效液相色谱法(HPLC)进行了比较。甲氨蝶呤(0.07 μ mol/l ~ 15 μ mol/l)加药人血清的序列内变异系数为11.7% ~ 2.5%,日间变异系数为14.0% ~ 5.8%。校正结果在9周内保持稳定。MTHO法测定结果与fia法和HPLC法测定结果具有良好的相关性。在低甲氨蝶呤浓度下,高效液相色谱法的测定结果比MTHO法和FPIA法分别平均低11%和14%。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Determination of methotrexate in serum: the aca MTHO immunoassay and HPLC compared.

The determination of methotrexate on the aca clinical analyser (aca MTHO method) was evaluated and the method compared with a fluorescence polarisation immunoassay (FPIA) and a specific high-performance liquid chromatography (HPLC) procedure. The within series and the between days coefficients of variation of measurements in human pool serum spiked with methotrexate (0.07 mumol/l to 15 mumol/l) ranges from 11.7% to 2.5% and 14.0% to 5.8%, respectively. The calibration was found to be stable for 9 weeks. There was a good correlation between the results of the MTHO method when compared with FPIA and HPLC. At low methotrexate concentration, the results of HPLC were on average 11% and 14% lower than those obtained by the MTHO assay and FPIA respectively.

求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信