四家主要制造商血管造影系统腹部采集模式的图像质量比较

L. Sukupova, J. Rydlo, Ondřej Hlaváček, D. Vedlich, J. Peregrin
{"title":"四家主要制造商血管造影系统腹部采集模式的图像质量比较","authors":"L. Sukupova, J. Rydlo, Ondřej Hlaváček, D. Vedlich, J. Peregrin","doi":"10.5430/IJDI.V5N2P6","DOIUrl":null,"url":null,"abstract":"Objective: The aim of this study was to compare image quality of different abdominal acquisition modes under conditions simulating obese patients whose images suffer more from noise and scatter radiation. Images were acquired in clinically used acquisition modes on the static and dynamic phantom for four angiography systems.Methods: A LEGO cart with 34 cm of PMMA and Pro-RTG Fluo18 phantom were used to simulate obese patients. The low-contrast resolution was assessed subjectively by two readers and objectively using signal-difference-to-noise ratio (SDNR) and using SDNR to air kerma rate. The line-pair resolution was assessed using the transmitted contrast value for line-pair groups.Results: Systems use different exposure parameters and dose but they differ in postprocessing too. Qualitative and quantitative assessments of noise produced similar results, images produced by systems A and C were noisier than by systems B and D. Highest SDNR was provided by System B, whilst System A produced the lowest values, which were almost the same for objects with different contrast. The image quality was affected mainly by frame lengths and postprocessing, but also by the dose. The images of the static phantom were better compared to the images of the dynamic phantom, which was an expected result.Conclusions: It was possible to identify image quality differences and to characterize features of postprocessing from measurements on standardized objects. A potential for optimization on some systems was identified, although further work, including assessment of clinical images, would be needed as part of the optimization process.","PeriodicalId":135923,"journal":{"name":"International Journal of Diagnostic Imaging","volume":"101 1","pages":"0"},"PeriodicalIF":0.0000,"publicationDate":"2018-08-13","publicationTypes":"Journal Article","fieldsOfStudy":null,"isOpenAccess":false,"openAccessPdf":"","citationCount":"0","resultStr":"{\"title\":\"Comparison of image quality of an abdominal acquisition mode of angiography systems from four major manufacturers\",\"authors\":\"L. Sukupova, J. Rydlo, Ondřej Hlaváček, D. Vedlich, J. Peregrin\",\"doi\":\"10.5430/IJDI.V5N2P6\",\"DOIUrl\":null,\"url\":null,\"abstract\":\"Objective: The aim of this study was to compare image quality of different abdominal acquisition modes under conditions simulating obese patients whose images suffer more from noise and scatter radiation. Images were acquired in clinically used acquisition modes on the static and dynamic phantom for four angiography systems.Methods: A LEGO cart with 34 cm of PMMA and Pro-RTG Fluo18 phantom were used to simulate obese patients. The low-contrast resolution was assessed subjectively by two readers and objectively using signal-difference-to-noise ratio (SDNR) and using SDNR to air kerma rate. The line-pair resolution was assessed using the transmitted contrast value for line-pair groups.Results: Systems use different exposure parameters and dose but they differ in postprocessing too. Qualitative and quantitative assessments of noise produced similar results, images produced by systems A and C were noisier than by systems B and D. Highest SDNR was provided by System B, whilst System A produced the lowest values, which were almost the same for objects with different contrast. The image quality was affected mainly by frame lengths and postprocessing, but also by the dose. The images of the static phantom were better compared to the images of the dynamic phantom, which was an expected result.Conclusions: It was possible to identify image quality differences and to characterize features of postprocessing from measurements on standardized objects. A potential for optimization on some systems was identified, although further work, including assessment of clinical images, would be needed as part of the optimization process.\",\"PeriodicalId\":135923,\"journal\":{\"name\":\"International Journal of Diagnostic Imaging\",\"volume\":\"101 1\",\"pages\":\"0\"},\"PeriodicalIF\":0.0000,\"publicationDate\":\"2018-08-13\",\"publicationTypes\":\"Journal Article\",\"fieldsOfStudy\":null,\"isOpenAccess\":false,\"openAccessPdf\":\"\",\"citationCount\":\"0\",\"resultStr\":null,\"platform\":\"Semanticscholar\",\"paperid\":null,\"PeriodicalName\":\"International Journal of Diagnostic Imaging\",\"FirstCategoryId\":\"1085\",\"ListUrlMain\":\"https://doi.org/10.5430/IJDI.V5N2P6\",\"RegionNum\":0,\"RegionCategory\":null,\"ArticlePicture\":[],\"TitleCN\":null,\"AbstractTextCN\":null,\"PMCID\":null,\"EPubDate\":\"\",\"PubModel\":\"\",\"JCR\":\"\",\"JCRName\":\"\",\"Score\":null,\"Total\":0}","platform":"Semanticscholar","paperid":null,"PeriodicalName":"International Journal of Diagnostic Imaging","FirstCategoryId":"1085","ListUrlMain":"https://doi.org/10.5430/IJDI.V5N2P6","RegionNum":0,"RegionCategory":null,"ArticlePicture":[],"TitleCN":null,"AbstractTextCN":null,"PMCID":null,"EPubDate":"","PubModel":"","JCR":"","JCRName":"","Score":null,"Total":0}
引用次数: 0

摘要

目的:本研究的目的是比较不同腹部采集方式在模拟肥胖患者图像受噪声和散射辐射影响较大的情况下的图像质量。在临床上使用的采集模式下,对四种血管造影系统的静态和动态幻影进行图像采集。方法:采用装34 cm PMMA的LEGO推车和Pro-RTG Fluo18假体模拟肥胖患者。低对比度分辨率主观上由两名阅读者评估,客观上使用信噪比(SDNR)和SDNR对空气分辨率进行评估。使用线对组的传输对比度值评估线对分辨率。结果:系统采用不同的暴露参数和剂量,但后处理也不同。对噪声的定性和定量评估产生了类似的结果,系统A和C产生的图像比系统B和d产生的噪声更大。系统B提供的SDNR最高,而系统A产生的SDNR最低,对于不同对比度的物体几乎相同。图像质量主要受帧长和后处理的影响,但也受剂量的影响。静态模体的成像效果优于动态模体,这是预期的结果。结论:通过对标准化物体的测量,可以识别图像质量差异并表征后处理特征。虽然需要进一步的工作,包括临床图像评估,作为优化过程的一部分,但仍确定了某些系统的优化潜力。
本文章由计算机程序翻译,如有差异,请以英文原文为准。
Comparison of image quality of an abdominal acquisition mode of angiography systems from four major manufacturers
Objective: The aim of this study was to compare image quality of different abdominal acquisition modes under conditions simulating obese patients whose images suffer more from noise and scatter radiation. Images were acquired in clinically used acquisition modes on the static and dynamic phantom for four angiography systems.Methods: A LEGO cart with 34 cm of PMMA and Pro-RTG Fluo18 phantom were used to simulate obese patients. The low-contrast resolution was assessed subjectively by two readers and objectively using signal-difference-to-noise ratio (SDNR) and using SDNR to air kerma rate. The line-pair resolution was assessed using the transmitted contrast value for line-pair groups.Results: Systems use different exposure parameters and dose but they differ in postprocessing too. Qualitative and quantitative assessments of noise produced similar results, images produced by systems A and C were noisier than by systems B and D. Highest SDNR was provided by System B, whilst System A produced the lowest values, which were almost the same for objects with different contrast. The image quality was affected mainly by frame lengths and postprocessing, but also by the dose. The images of the static phantom were better compared to the images of the dynamic phantom, which was an expected result.Conclusions: It was possible to identify image quality differences and to characterize features of postprocessing from measurements on standardized objects. A potential for optimization on some systems was identified, although further work, including assessment of clinical images, would be needed as part of the optimization process.
求助全文
通过发布文献求助,成功后即可免费获取论文全文。 去求助
来源期刊
自引率
0.00%
发文量
0
×
引用
GB/T 7714-2015
复制
MLA
复制
APA
复制
导出至
BibTeX EndNote RefMan NoteFirst NoteExpress
×
提示
您的信息不完整,为了账户安全,请先补充。
现在去补充
×
提示
您因"违规操作"
具体请查看互助需知
我知道了
×
提示
确定
请完成安全验证×
copy
已复制链接
快去分享给好友吧!
我知道了
右上角分享
点击右上角分享
0
联系我们:info@booksci.cn Book学术提供免费学术资源搜索服务,方便国内外学者检索中英文文献。致力于提供最便捷和优质的服务体验。 Copyright © 2023 布克学术 All rights reserved.
京ICP备2023020795号-1
ghs 京公网安备 11010802042870号
Book学术文献互助
Book学术文献互助群
群 号:481959085
Book学术官方微信